Home > Adam Equipment > Scale > Adam Equipment CBD Bench Counting Scales CBDREVBAE2 User Manual

Adam Equipment CBD Bench Counting Scales CBDREVBAE2 User Manual

Here you can view all the pages of manual Adam Equipment CBD Bench Counting Scales CBDREVBAE2 User Manual. The Adam Equipment manuals for Scale are available online for free. You can easily download all the documents as PDF.

Page 31

 
© Adam Equipment Company 2011  Page  
29 
 
Sample of output:   
Using M+ function Using PRINT function 
DATE   01/27/2011  
TIME   12:38:49 
LOCAL SCALE 
SID: ABS123  
UID: 123ABC  
NAME:Text  
  4    S. No.  
 1.234 kg TARE 
12.456 kg NET 
1.1234  g U.W.  
    11 PCS  
TOTAL 
- - - - - - - - - - - -  
 49.824 kg TW 
     44 TPC 
      4 No.    DATE   01/27/2011  
TIME   12:39:58 
LOCAL SCALE 
SID: ABS123  
UID: 123ABC   
NAME:Text  
 1.234 kg TARE 
12.456 kg NET 
1.1234  g U.W.  
    11 PCS    
 
9.1...

Page 32

 
© Adam Equipment Company 2011  Page  
30 
 
Basic Commands:  
PLUxx  Select PLU from scale memory to be used 
T  Tare current weight value 
T123.456 preset tare value is 123.456 
Z Zero the display 
P Print using selected format 
M+ Store current results into memory and print 
MR Recall memory values to scale display 
MC  Clear memory 
U123.456 Store unit weight of 123.456  
(grams if in kilograms or pounds if currently in po unds)
 
S123  Enter sample size of 123 parts.   
Same as pressing [Smpl]...

Page 33

 
© Adam Equipment Company 2011  Page  
31 
 
  
9.3  PLU ENTRY USING RS-232 INTERFACE    
This method enables the user to send the data from  a PC program as well as 
from  the  keypad.  The  most  common  PLUs  can  be  stored   and  recalled  from 
the  scale  memory.  Other  PLU  data  can  be  stored  on  a   PC,  then  the 
Description  (text  data),  Unit  Weight  and  Tare  Weigh t  data  can  be  sent  from 
the  PC  for  the  PLUxx  (where  xx  can  be  any  desired  n umber  up  to  99).  This...

Page 34

 
© Adam Equipment Company 2011  Page  
32 
 
   
10.0 CALIBRATION 
 
ACTION  DISPLAYS 
Press the  [Tare/Zero]  key during the self-
test at power on. The scale will ask to  
enter the password. 
 
  
“ Pi n” 
Default Password is 0000.  
Enter “0” four times. 
Password can be changed in technical 
parameters.  
Press  [Print] . 
    
“ Pi n”     “ - - - -” 
Select the scale to be set up by using the  
[Local/Remote]
 key. 
Press the  [Print] key to enter the 
technical section.  
 
“tECH”  “   LocAL”   “...

Page 35

 
© Adam Equipment Company 2011  Page  
33 
 
 
The Calibration is complete, the scale  
will run the self-test during which the 
weight should be removed.   
If calibration is not possible an error 
message will be shown  
“- - - -”    “  - - - -”    “ - - - - ”  
“SPAn”    “ PASS”    “        ”    
   
“SPAn”    “ FAiL”    “ LoW   ”  
or  
“HiGH 
  
If The Calibration is being done while in the TECH  parameters (section 12)  the linearity is 
set at the same time as calibration by performing t he...

Page 36

 
© Adam Equipment Company 2011  Page  
34 
 
 
11.0  ERROR CODES 
 
During  the  initial  power-on  testing  or  during  the  o peration,  the  scale  may 
show  an  error  message.    The  meaning  of  the  error  me ssages  are  described 
below. 
 
If  an  error  message  is  shown,  repeat  the  procedure  that  caused  the 
message,  such  as  turning  the  balance  on,  calibratio n  or  any  other  functions.  
If  the  error  message  is  still  shown  then  contact  yo ur  dealer  for  further...

Page 37

 
© Adam Equipment Company 2011  Page  
35 
 
 
12.0 TECHNICAL PARAMETERS 
 
The  technical  parameters  accessed  via  the  “ tECH”  prompt  at  the  end  of  user 
parameters  and  are  password  controlled  to  prevent  u nauthorised  access.  
These  parameters  set  the  metrology  for  the  scales.      Each  scale  is  set 
independently.  The  parameters  will  set  capacity,  di vision,  decimal  point 
position,  initial  zero  range,  auto  and  manual  zero  range  as  well  as  factory...

Page 38

 
© Adam Equipment Company 2011  Page  
36 
 
 
Press the 
[U.Wt/Unit]  to go to next 
menu. Press  [Print] to enter. 
Use  [U.Wt/Unit]  to select the 
decimal point position and press  
[Print].  
To change the remote scale cap 
press CE to clear and then enter  
new value followed by  [Print]. 
      “
CAP ”.  Scale capacity setting.  
 
“dESC” “  0.00”      
“LOAd” “0060”    
Press  [U.Wt/Unit]  to go to next 
menu. Press  [Print] to enter. 
Use  [U.Wt/Unit]  to select the 
division which you would...

Page 39

 
© Adam Equipment Company 2011  Page  
37 
  

Page 40

©

 Adam Equipment Company 2011  Page  
38 
13.0 REPLACEMENT PARTS AND ACCESSORIES 
If you need to order any spare parts and accessories, contact your supplier or Adam 
Equipment. A partial list of such items is as follows- 
·Power Supply Module
· Main Power cord
· Replacement Battery
· Stainless Steel Pan ·
In use cover
· Printer, etc.
14.0 SERVICE INFORMATION 
This  manual  covers  the  details  of  operation.  If  you  have  a  problem  with  the 
scale that is not directly addressed by this manual...