Page 31
© Adam Equipment Company 2011 Page 29 Sample of output: Using M+ function Using PRINT function DATE 01/27/2011 TIME 12:38:49 LOCAL SCALE SID: ABS123 UID: 123ABC NAME:Text 4 S. No. 1.234 kg TARE 12.456 kg NET 1.1234 g U.W. 11 PCS TOTAL - - - - - - - - - - - - 49.824 kg TW 44 TPC 4 No. DATE 01/27/2011 TIME 12:39:58 LOCAL SCALE SID: ABS123 UID: 123ABC NAME:Text 1.234 kg TARE 12.456 kg NET 1.1234 g U.W. 11 PCS 9.1...
Page 32
© Adam Equipment Company 2011 Page 30 Basic Commands: PLUxx Select PLU from scale memory to be used T Tare current weight value T123.456 preset tare value is 123.456 Z Zero the display P Print using selected format M+ Store current results into memory and print MR Recall memory values to scale display MC Clear memory U123.456 Store unit weight of 123.456 (grams if in kilograms or pounds if currently in po unds) S123 Enter sample size of 123 parts. Same as pressing [Smpl]...
Page 33
© Adam Equipment Company 2011 Page 31 9.3 PLU ENTRY USING RS-232 INTERFACE This method enables the user to send the data from a PC program as well as from the keypad. The most common PLUs can be stored and recalled from the scale memory. Other PLU data can be stored on a PC, then the Description (text data), Unit Weight and Tare Weigh t data can be sent from the PC for the PLUxx (where xx can be any desired n umber up to 99). This...
Page 34
© Adam Equipment Company 2011 Page 32 10.0 CALIBRATION ACTION DISPLAYS Press the [Tare/Zero] key during the self- test at power on. The scale will ask to enter the password. “ Pi n” Default Password is 0000. Enter “0” four times. Password can be changed in technical parameters. Press [Print] . “ Pi n” “ - - - -” Select the scale to be set up by using the [Local/Remote] key. Press the [Print] key to enter the technical section. “tECH” “ LocAL” “...
Page 35
© Adam Equipment Company 2011 Page 33 The Calibration is complete, the scale will run the self-test during which the weight should be removed. If calibration is not possible an error message will be shown “- - - -” “ - - - -” “ - - - - ” “SPAn” “ PASS” “ ” “SPAn” “ FAiL” “ LoW ” or “HiGH If The Calibration is being done while in the TECH parameters (section 12) the linearity is set at the same time as calibration by performing t he...
Page 36
© Adam Equipment Company 2011 Page 34 11.0 ERROR CODES During the initial power-on testing or during the o peration, the scale may show an error message. The meaning of the error me ssages are described below. If an error message is shown, repeat the procedure that caused the message, such as turning the balance on, calibratio n or any other functions. If the error message is still shown then contact yo ur dealer for further...
Page 37
© Adam Equipment Company 2011 Page 35 12.0 TECHNICAL PARAMETERS The technical parameters accessed via the “ tECH” prompt at the end of user parameters and are password controlled to prevent u nauthorised access. These parameters set the metrology for the scales. Each scale is set independently. The parameters will set capacity, di vision, decimal point position, initial zero range, auto and manual zero range as well as factory...
Page 38
© Adam Equipment Company 2011 Page 36 Press the [U.Wt/Unit] to go to next menu. Press [Print] to enter. Use [U.Wt/Unit] to select the decimal point position and press [Print]. To change the remote scale cap press CE to clear and then enter new value followed by [Print]. “ CAP ”. Scale capacity setting. “dESC” “ 0.00” “LOAd” “0060” Press [U.Wt/Unit] to go to next menu. Press [Print] to enter. Use [U.Wt/Unit] to select the division which you would...
Page 40
© Adam Equipment Company 2011 Page 38 13.0 REPLACEMENT PARTS AND ACCESSORIES If you need to order any spare parts and accessories, contact your supplier or Adam Equipment. A partial list of such items is as follows- ·Power Supply Module · Main Power cord · Replacement Battery · Stainless Steel Pan · In use cover · Printer, etc. 14.0 SERVICE INFORMATION This manual covers the details of operation. If you have a problem with the scale that is not directly addressed by this manual...