Adam Equipment AELP Pallet Beams AELPREVA6 User Manual
Here you can view all the pages of manual Adam Equipment AELP Pallet Beams AELPREVA6 User Manual. The Adam Equipment manuals for Pallet Beam are available online for free. You can easily download all the documents as PDF.
Page 11
© Adam Equipment Company 2005 9 6.5 cont = CONTINUOUS PRINTING This parameter is used to configure to print contin uously whenever there is any weight on the balance or only when [Print] is pressed. To set the parameter press the [Func] key to show “ cont” Press [Print] to see the parameter. Press [Print] to select either “ cont= 0” or “cont= 1 ”. “ cont= 0 ” Continuous output is disabled. Prints only when t he [Print] key is pressed. “ cont= 1 ”...
Page 12
© Adam Equipment Company 2005 10 of parts. If more items are added or removed from t he balance, the display will show the new quantity. Note that the display is limited to 6 digits, i.e. 999,999 pieces. If the quantity on the scale exceeds 999,999 the left most digits and any leading zeros will be lost, for example 1,002,3 05 will display as 2,305 pieces. When in use the balance can be tared normally to e liminate the package weight from the gross weight. To return to...
Page 13
© Adam Equipment Company 2005 11 6.7 StEPS = CHECK-WEIGHING FUNCTION (LOW-OK-HIGH) The check-weighing function will show a symbol on the left side of the display to indicate if the current weight displ ayed is: - above the High setpoint - between the setpoints - below the Low setpoint To enable the check-weighing function and set the value of the setpoints press the [Func] key until “ StEPS” is displayed. Press [Print] to enter the...
Page 14
© Adam Equipment Company 2005 12 6.8 CALIbr = CALIBRATION ROUTINE To set the parameter press the [Func] key to show “ CALIbr”. Remove all weight from the platform. Press [Print] and the display will show “ noCAL” while the scale is measuring the initial mass (pan must be empty). Then the display will show “ LOAd xxx” where xxx is the required calibration weight. Place the weight on the platform then press [Print] – display...
Page 15
© Adam Equipment Company 2005 13 6.10 StAb = PRINT WHEN STABLE or INSTANTANEOUS When the scale is set to manual print ( rePL = 0), the balance can be set to print either immediately after the [Print] key is pressed or only after the balance is stable. The function c an also be enabled from the RS-232 interface. See Section 3: Communication with a Computer or Printer. To set the parameter press the [Func] key to show “ StAb”. ...
Page 16
© Adam Equipment Company 2005 14 6.12 t1 = AUTOMATIC POWER SWITCH OFF The scale includes an internal battery. The typical operation time using only the battery is 50 hours if one load cell is used an d about 12 hours if four are used. The scale includes the “ t1” function to turn the power off after 5 minutes if it is not being used. This function can be disab led if the balance is powered from the main power supply or if the disruption of power affect...
Page 17
© Adam Equipment Company 2005 15 6.15 SuPP = Power supply voltage to PCB To set the parameter press the [Func] key to show “ SuPP”. Press [Print] to see the parameter. The display will show the current voltage to the P CB. Press any key to return to normal weighing. 7.0 COMMUNICATION WITH A COMPUTER/PRINTER The following applies only to scales supplied with an RS-232 interface. Press the [Print] key to transmit weighing data (value and unit of...
Page 18
© Adam Equipment Company 2005 16 If the output connector is a 6 pin audio connector. The output pins are: It is necessary to jumper pin 1 to pin 6 to enable the RS-232 interface circuits. Do not connect these pins to the handshak ing pins on the interfaced device. 7.3 OUTPUT FORMAT The balance will output the weight with the units o f measure on one line. The output is initiated when the [Print] key is pressed or a command is received over the RS-232...
Page 19
© Adam Equipment Company 2005 Manufacturer’s Declaration of Conformity This product has been manufactured in accordance with the harmonised European standards, following the provisions of the below stated directives: Electro Magnetic Compatibility Directive 89/336/EEC Low Voltage Directive 73/23/EEC Adam Equipment Co. Ltd. Maidstone Road, Kingston, Milton Keynes, MK10 0BD United Kingdom FCC COMPLIANCE This equipment has been tested and found to comply with the limits for a...
Page 20
© Adam Equipment Company 2005 ADAM EQUIPMENT is an ISO 9001:2000 certified global organisation with more than 30 years experience in the production and sale of electronic weighing equipments. Products are sold through a world wide distributor network -supported from our company locations in the UK, USA and SOUTH AFRICA. The company and their distributors offer a full range of Technical Services such as on site and workshop repair, preventative maintenance and...