29 | P a g e © Adam Equipment Company 2010 Description ENGLISH FRENCH GERMAN SPANISH Net weight Net Wt. Pds Net Net-Gew Pso Net Weight per unit counted Unit Wt. Pds unit Gew/Einh Pso/Unid Number of items counted Pcs Pcs Stck. Piezas Number of weighing added to subtotals No. Nb. Anzhl Num. Total weight and count printed Total Total Gesamt Total Print date Date Date Datum Fecha Print time Time Heure Zeit Hora Scale ID...
Page 32
30 | P a g e © Adam Equipment Company 2010 10.0 RELAY INTERFACE The AE 402 indicator is supplied with drivers to control external relays. The drivers could be used to control a number of different rela ys depending upon the users need. The relay drivers are isolated outputs req uiring the use of an external power supply for the relays. Also see section 3.4.3. Contact Adam Equipment or your supplier for a relay interface that is compatible with the relay drivers,...
Page 33
31 | P a g e © Adam Equipment Company 2010 11.0 CALIBRATION The scale can be calibrated using the following procedure. To enter this procedure it is necessary to use Func 4 accessible using the [Func/C] key as described in section 7.4 or using the password access as described in section 12. The AE 402 scales calibrate using either metric or pound weights, depending on the weighing unit in use before calibration. The disp lay will show either "kg" or...
Page 34
32 | P a g e © Adam Equipment Company 2010 12.0 SERVICE SECTION The scales will allow entry to the parameters if the [Tare] key is pressed during the power on cycle. The passwords work as above. In this case the display will show the passcode req uest screen, “ P - - - - “ . To continue enter a passcode as described below. Entering passcode 0000 will allow calibration as sh own in section 11. Entering 1000 will allow access...
Page 35
33 | P a g e © Adam Equipment Company 2010 The parameters available are: “F4 Int” Initial Zero Range “F5 rEZ” Re-Zero range “F6 SCS” Successive Tare Enable “F7 Cnt” Display the A/D counts “F8 Zem” Zero Mode “F9 Lvd” Low voltage detection “FA AZn” Auto Zero Range “Fb FPS” User Function Password The description of the parameters is shown in secti on 12.3 12.2 USING “2006” TO ENTER THE SECURE PARAMETERS Press...
Page 36
34 | P a g e © Adam Equipment Company 2010 12.3 PARAMETER DETAILS 12.3.1 F1 -CALIBRATION To enter the calibration parameter, press the [Z/T] key when “F1” is displayed. The scale will be calibrated using 2 masses of approxim ately 1/3Maximum and Maximum. If the scale has been calibrated once the values will be stored. If this is the first time the scale is calibrated the user mus t enter the values for the calibration...
Page 37
35 | P a g e © Adam Equipment Company 2010 12.3.3 F3 – CAPACITY/ WEIGHING UNIT/INCREMENT To enter this parameter, press the [Z/T] key when “F3 CAP” is shown. The display will show the current capacity and defa ult weighing unit. Enter the numeric values using the [Unit/ ] and [Pcs/ ] keys. The scale will check the number of division’s n = maximum/incremen t is less than 30,000 divisions. (3000 divisions for EC approved versions). To select...
Page 38
36 | P a g e © Adam Equipment Company 2010 12.3.4 F4 –INITIAL ZERO RANGE To enter this parameter, press the [Z/T] key when “F4 int” is shown. The display will show the current initial zero rang e. Press the [Pcs/] key to change the value and press [Z/T] to accept the value. Press [Print/M+/esc] to return to weighing. 12.3.5 F5 -RE-ZERO RANGE To enter this parameter, press the [Z/T] key when “F5 rEZ” is shown. The display will show the current re-zero range....
Page 39
37 | P a g e © Adam Equipment Company 2010 12.3.9 F9 –LOW VOLTAGE DETECTION This parameter allows detection of low voltage when the battery wears down. To enter this parameter, press the [Z/T] key when “F9 LVd” is shown. The display will show if the LVD Mode is set to on or oFF. Press the [Pcs/ ] key to change the value. Press [Z/T] to accept the displayed value. Press [Print/M+/Esc] to return to weighing. 12.3.10 FA –AZn Auto Zero Range This...
Page 40
38 | P a g e © Adam Equipment Company 2010 13.0 ERROR CODES ERROR CODES DESCRIPTION SUGGESTIONS - -oL - - Over-range Remove weight from the scale. If the problem persists contact your dealer or Adam Equipment for assistance. Err 1 Time Setting Error Enter time using correct format and reasonable values. Format: hh:mm:ss Err 2 Date Setting Error Enter date using correct format and reasonable values. Format: yy:mm:dd Err 4 Zero Setting Error...