Page 31

29 | P a g e              © Adam Equipment Company 2010  
 Description  ENGLISH  FRENCH  GERMAN   SPANISH 
Net weight    Net Wt.  Pds Net  Net-Gew  Pso Net 
Weight  per  unit  
counted   Unit Wt.  Pds unit  Gew/Einh  Pso/Unid 
Number  of  items  
counted   Pcs  Pcs  Stck.  Piezas 
Number  of  
weighing    added 
to subtotals  No.  Nb.  Anzhl  Num. 
Total  weight  and  
count printed   
Total  Total  Gesamt  Total 
Print date   
Date  Date  Datum  Fecha 
Print time   Time  Heure  Zeit  Hora 
Scale ID...

Page 32

30 | P a g e              © Adam Equipment Company 2010  
 
10.0  RELAY INTERFACE 
   
The AE 402 indicator is supplied with drivers to control external relays.  The drivers 
could be used to control a number of different rela ys depending upon the users 
need.    The relay drivers are isolated outputs req uiring the use of an external power 
supply for the relays.    Also see section 3.4.3.   
Contact Adam Equipment or your supplier for a relay  interface that is compatible 
with the relay drivers,...

Page 33

31 | P a g e              © Adam Equipment Company 2010  
 
11.0  CALIBRATION   
The scale can be calibrated using the following procedure.     To enter this procedure 
it is necessary to use Func 4 accessible using the  [Func/C] key as described in section 
7.4   or using the password access as described in  section 12. 
 
The AE 402 scales calibrate using either metric or  pound weights, depending on the 
weighing unit in use before calibration.   The disp lay will show either "kg" or...

Page 34

32 | P a g e              © Adam Equipment Company 2010  
 
12.0  SERVICE SECTION      
  
The  scales  will  allow  entry  to  the  parameters  if  the [Tare]   key  is  pressed  during  the 
power on cycle.   The passwords work as above.  
In  this  case  the  display  will  show  the  passcode  req uest  screen,  “  P  -  -  -  -  “ .    To 
continue enter a passcode as described below.    
Entering passcode 0000 will allow calibration as sh own in section 11. 
 
Entering  1000  will  allow  access...

Page 35

33 | P a g e              © Adam Equipment Company 2010  
The parameters available are:  
 “F4 Int”        Initial Zero Range  
“F5 rEZ”        Re-Zero range  
“F6 SCS”        Successive Tare Enable 
“F7  Cnt”        Display the A/D counts 
“F8 Zem”       Zero Mode   
“F9 Lvd”        Low voltage detection   
“FA AZn”         Auto Zero Range 
“Fb FPS”        User Function Password 
  
The description of the parameters is shown in secti on 12.3 
 
12.2
   USING “2006” TO ENTER THE SECURE PARAMETERS 
Press...

Page 36

34 | P a g e              © Adam Equipment Company 2010  
  
12.3
   PARAMETER DETAILS  
  
12.3.1  F1 -CALIBRATION 
 
To  enter  the  calibration  parameter,  press  the   [Z/T] key  when  “F1”  is  displayed.    The 
scale  will  be  calibrated  using  2  masses  of  approxim ately  1/3Maximum  and 
Maximum.  If  the  scale  has been calibrated once  the   values  will  be  stored.   If  this  is 
the  first  time  the  scale  is  calibrated  the  user  mus t  enter  the  values  for  the 
calibration...

Page 37

35 | P a g e              © Adam Equipment Company 2010  
  
12.3.3   F3 – CAPACITY/ WEIGHING UNIT/INCREMENT 
 
To enter this parameter, press the [Z/T] key when  “F3 CAP” is shown. 
The display will show the current capacity and defa ult weighing unit.   
Enter  the  numeric  values  using  the  [Unit/


] and   [Pcs/ 


] keys.        The  scale  will 
check the number of division’s n = maximum/incremen t is less than 30,000 divisions.  
(3000 divisions for EC approved versions).    
To  select...

Page 38

36 | P a g e              © Adam Equipment Company 2010  
12.3.4   F4 –INITIAL ZERO RANGE    
To enter this parameter, press the [Z/T] key when  “F4 int” is shown. 
The display will show the current initial zero rang e.      
Press the  [Pcs/] key to change the value and press  [Z/T] to accept the value. 
Press  [Print/M+/esc]  to return to weighing. 
  
12.3.5   F5 -RE-ZERO RANGE 
 
To enter this parameter, press the  [Z/T] key when  “F5 rEZ” is shown. 
The display will show the current re-zero range....

Page 39

37 | P a g e              © Adam Equipment Company 2010  
12.3.9    F9 –LOW VOLTAGE DETECTION     
This parameter allows detection of low voltage when the battery wears down. 
To enter this parameter, press the  [Z/T] key when  “F9 LVd” is shown. 
The display will show if the LVD Mode is set to on  or oFF.     
Press the  [Pcs/


] key to change the value. 
Press  [Z/T] to accept the displayed value. 
Press  [Print/M+/Esc]  to return to weighing. 
  
12.3.10    FA –AZn Auto Zero Range   
This...

Page 40

38 | P a g e              © Adam Equipment Company 2010  
 
13.0  ERROR CODES 
 
ERROR  
CODES 
DESCRIPTION SUGGESTIONS 
- -oL - - Over-range Remove weight from the scale.   
 
If  the  problem  persists  contact  your  dealer  or  Adam 
Equipment for assistance. 
  
Err 1 Time Setting Error Enter  time  using  correct  format  and  reasonable  values. 
Format: hh:mm:ss 
 
Err 2 Date Setting Error Enter  date  using  correct  format  and  reasonable  values. 
Format: yy:mm:dd 
 
Err 4 Zero Setting Error...