29 | P a g e © Adam Equipment Company 2010 Description ENGLISH FRENCH GERMAN SPANISH Net weight Net Wt. Pds Net Net-Gew Pso Net Weight per unit counted Unit Wt. Pds unit Gew/Einh Pso/Unid Number of items counted Pcs Pcs Stck. Piezas Number of weighing added to subtotals No. Nb. Anzhl Num. Total weight and count printed Total Total Gesamt Total Print date Date Date Datum Fecha Print time Time Heure Zeit Hora Scale ID number Scale ID Bal ID Waagen ID Bal ID User ID Number User ID Util ID Nutzer ID Usuario ID 9.1 INPUT COMMANDS FORMAT The scale can be controlled with the following comm ands. Press the [Enter] key of the PC after each command. T Tares the scale to display the net weight. This is the same as pressing [Z/T]. Z Sets the zero point for all subsequent weighing. Th e display shows zero. P Prints the results to a PC or printer using the RS- 232 interface. It also adds the value to the accumulation memory if t he accumulation function is not set to automatic.
30 | P a g e © Adam Equipment Company 2010 10.0 RELAY INTERFACE The AE 402 indicator is supplied with drivers to control external relays. The drivers could be used to control a number of different rela ys depending upon the users need. The relay drivers are isolated outputs req uiring the use of an external power supply for the relays. Also see section 3.4.3. Contact Adam Equipment or your supplier for a relay interface that is compatible with the relay drivers, however other interfaces ca n be used as long as the following conditions apply. Connections to the drivers: Connections are made to the PCB, Connector P1. Th e connector is a spring activated type, simply press on the top of the conn ector and insert the wire. Do not exceed the safe limits of voltage or current of 24VDC or 15ma per output. Depending upon the application it may be necessary to use a small relay to drive larger relays, or to provide additional protection to prevent electromagnetic interference (diodes as shown above) to this or oth er machinery.
31 | P a g e © Adam Equipment Company 2010 11.0 CALIBRATION The scale can be calibrated using the following procedure. To enter this procedure it is necessary to use Func 4 accessible using the [Func/C] key as described in section 7.4 or using the password access as described in section 12. The AE 402 scales calibrate using either metric or pound weights, depending on the weighing unit in use before calibration. The disp lay will show either "kg" or "lb" to identify the weights expected. PROCEDURE Enter the calibration section using Func 4, C8 CAL or using the password as described in section 12. The display will show " unLoAd". Remove any weight from platform, when the stable sy mbol is on press [Z/T] . The display will show " Ld" the “ 0000XX ” to show the last calibration weight used. Place this calibration weight on th e scale. Press the [Z/T] key. If the weight you put on the scale does not match the value displayed, press the [Func/C] key to clear the vale then use the [Unit/ ]key and [Pcs/ ] key to set the correct value. When it is correct press [Z/T]. If the calibration is acceptable the scale will run a self-test during which the calibration weight should be removed. If an er ror message “FAiL L” is shown try calibration again as a disturbance may have prevented a successful calibration. After calibration the scale should be checked to ve rify the calibration is correct. If necessary repeat calibration, make sure that the sc ale is stable before accepting any weight.
32 | P a g e © Adam Equipment Company 2010 12.0 SERVICE SECTION The scales will allow entry to the parameters if the [Tare] key is pressed during the power on cycle. The passwords work as above. In this case the display will show the passcode req uest screen, “ P - - - - “ . To continue enter a passcode as described below. Entering passcode 0000 will allow calibration as sh own in section 11. Entering 1000 will allow access to a limited set of parameters described in section 12.1. Entering the passcode 2006 will allow access to all parameters as shown in section 12.3. 12.1 USING “1000” TO ENTER THE SERVICE PARAMETERS Press the [Z/T] key during the display counting when turned on, When “Pn “ is displayed, enter the number 1000 using the [Unit/ ] and [Pcs/ ] keys and then press [Z/T]. The displays will show the first parameters, called “F4 Int”. To select another parameter press the [Pcs/ ] key to advance through the parameters. Press the [Z/T] key to enter a parameter. To exit a parameter, press the [Print/M+/Esc] key. The display will show the parameter number and a na me. When a parameter is entered by pressing the [Z/T] key, the displays will guide you through the parameter selected and the options avai lable.
33 | P a g e © Adam Equipment Company 2010 The parameters available are: “F4 Int” Initial Zero Range “F5 rEZ” Re-Zero range “F6 SCS” Successive Tare Enable “F7 Cnt” Display the A/D counts “F8 Zem” Zero Mode “F9 Lvd” Low voltage detection “FA AZn” Auto Zero Range “Fb FPS” User Function Password The description of the parameters is shown in secti on 12.3 12.2 USING “2006” TO ENTER THE SECURE PARAMETERS Press the [Z/T] key during the display counting when turned on, When “Pn “ is displayed, enter the number 2006 using the [Unit/ ] and [Pcs/ ] keys and then press [Z/T]. The displays will show the first parameters, c alled “F1” “CAL” . To select another parameter press the [Pcs/ ] key to advance through the parameters. Press the [Z/T] key to enter a parameter. To exit a parameter, press the [Print/M+/Esc] key. The display will show the parameter number and a n ame. When a parameter is entered by pressing the [Z/T] key, the displays will guide you through the parameter selected and the options avai lable. The parameters available are: “F1 CAL” To enter the Calibration “F2 dEC” Decimal Point Position “F3 CAP” Default Weighing Unit and capacity “F4 Int” Initial Zero Range “F5 rEZ” Re-Zero range “F6 SCS” Successive Tare Enable “F7 Cnt” Display the A/D counts “F8 Zem” Zero Mode “F9 Lvd” Low voltage detection “FA AZn” Auto Zero Range “Fb FPS” User Function Password
34 | P a g e © Adam Equipment Company 2010 12.3 PARAMETER DETAILS 12.3.1 F1 -CALIBRATION To enter the calibration parameter, press the [Z/T] key when “F1” is displayed. The scale will be calibrated using 2 masses of approxim ately 1/3Maximum and Maximum. If the scale has been calibrated once the values will be stored. If this is the first time the scale is calibrated the user mus t enter the values for the calibration masses. NOTE: It is necessary to set the decimal point loc ation and the capacity before calibration is possible. The display will instruct you to remove any weight from the scale, “UnloAd”. Press [Z/T] . The display will tell you to add the first weight to the scale: “Ld 1” “ 10 kg” If necessary change the value shown to match the weigh t to be used. Press [CE] to clear the old value and then enter the new value. All values entered are in whole numbers only. Add the weight shown, wait for stability then press the [Z/T] key. The display will tell you to add the second weigh t to the scale: “Ld 2 “ “ 30 kg” Either add the weight shown, wait for stability the n press the [Z/T] key or press the [Print/M+/Esc] key to accept the single point calibration and ret urn to weighing. The display will show “PASS” if the calibration is OK. Remove the weight. 12.3.2 F2–DECIMAL POINT POSITION To set the value for the decimal point location. The options are 0, 0.0, 0.00, 0.000, 0.0000 To enter this parameter, press the [Z/T] key when “F2 dEC” is shown. The display will show the current setting. Press the [Pcs/ ] key to change the value. Select from 0, 0.0, 0.00, 0.000, 0.0000 Press [Z/T] to accept the displayed value.
35 | P a g e © Adam Equipment Company 2010 12.3.3 F3 – CAPACITY/ WEIGHING UNIT/INCREMENT To enter this parameter, press the [Z/T] key when “F3 CAP” is shown. The display will show the current capacity and defa ult weighing unit. Enter the numeric values using the [Unit/ ] and [Pcs/ ] keys. The scale will check the number of division’s n = maximum/incremen t is less than 30,000 divisions. (3000 divisions for EC approved versions). To select the weighing unit press the [Unit/ ]until the default weighing unit is flashing, either kilogram (kg) or pounds (lb). Press [Z/T] to accept the displayed value. On non-approved versions the display then lets you select the increment, “InC 2” For example 100kg x 0.01kg the increment is 10grams , but the last digit increments by 1. The display will show the current increment value a s used with the current decimal point position. Press the [Pcs/ ] key to change the value. Select from 1, 2, 5, 10, 20 or 50 Not all increments may be available for the capacit y you selected. For EC Approved versions the scale will determine th e increment that maintains the number of divisions to be 3000 or less. Press [Z/T] to accept the displayed value. Press [Print/M+/Esc] to return to weighing.
36 | P a g e © Adam Equipment Company 2010 12.3.4 F4 –INITIAL ZERO RANGE To enter this parameter, press the [Z/T] key when “F4 int” is shown. The display will show the current initial zero rang e. Press the [Pcs/] key to change the value and press [Z/T] to accept the value. Press [Print/M+/esc] to return to weighing. 12.3.5 F5 -RE-ZERO RANGE To enter this parameter, press the [Z/T] key when “F5 rEZ” is shown. The display will show the current re-zero range. Press the [Pcs/ ] key to change the value. Press [Z/T] to accept the value. Press [Print/M+/esc] to return to weighing. 12.3.6 F6 -SUCCESSIVE TARE To enter this parameter, press the [Z/T] key when “ F6 SCS” is shown. The display will show if the successive tare is on or off. Press the [Pcs/ ] key to change the value. Press [Z/T] to accept the displayed value. Press [Print/M+/Esc] to return to weighing. 12.3.7 F7 –ADC COUNTS To enter this parameter, press the [Z/T] key when “F7 Cnt” is shown. This parameter allows you to view the A/D counts fr om the internal A/D converter. This can be an aid to service. Press the [Z/T] key to return to the PARAMETER menu. Press the [Print/M+/Esc] key to return to weighing. Typical value at zero is 30,000-90,000 (approx.). Typical value at full capacity is 500,000 (approx.) . 12.3.8 F8 –ZERO MODE To enter this parameter, press the [Z/T] key when “ F8 ZEm” is shown. Select the Zero mode desired. In all but special ca ses Zero Mode 1 is used. The other 2 zero modes are for unique locations in the world and effect the +/- range of the zero. Press the [Pcs/ ] key to change the value. Press [Z/T] to accept the displayed value. Press [Print/M+/Esc] to return to weighing.
37 | P a g e © Adam Equipment Company 2010 12.3.9 F9 –LOW VOLTAGE DETECTION This parameter allows detection of low voltage when the battery wears down. To enter this parameter, press the [Z/T] key when “F9 LVd” is shown. The display will show if the LVD Mode is set to on or oFF. Press the [Pcs/ ] key to change the value. Press [Z/T] to accept the displayed value. Press [Print/M+/Esc] to return to weighing. 12.3.10 FA –AZn Auto Zero Range This parameter sets the auto zero range from 0 (Off ) to 4d. Calibration and parameter counting function to be a ctive. To enter this parameter, press the [Z/T] key when “FA AZn” is shown. The display will show if the current Auto Zero rang e. Press the [Pcs/ ] key to change the value. Press [Z/T] to accept the displayed value. Press [Print/M+/Esc] to return to weighing. 12.3.11 Fb –FPS User Function Password This parameter set a password to control access to the user functions as described in Section 7. If the password is set to 0000 the c ontrol is disable and the user has free access to the functions. To enter this parameter, press the [Z/T] key when “Fb FPS” is shown. The display will show “ P1 - - - -“ Enter the new password code followed by [Z/T] to accept. The display will then show “ P2 - - - -“ Enter the new password code again followed by [Z/T] to accept. The same code number must be entered bo th times for the password to be set. Press the [Unit/ ] and [Pcs/ ] keys to enter the values. Press [Print/M+/Esc] to return to weighing.
38 | P a g e © Adam Equipment Company 2010 13.0 ERROR CODES ERROR CODES DESCRIPTION SUGGESTIONS - -oL - - Over-range Remove weight from the scale. If the problem persists contact your dealer or Adam Equipment for assistance. Err 1 Time Setting Error Enter time using correct format and reasonable values. Format: hh:mm:ss Err 2 Date Setting Error Enter date using correct format and reasonable values. Format: yy:mm:dd Err 4 Zero Setting Error The scale was outside the normal zero setting range either when it was turned on or when the [Zero] key was pressed. Remove weight from the scale and try re-zeroing aga in. Use the [Z/T] key to set the display to zero value. If the problem persists contact your dealer or Adam Equipment for assistance. Err 6 A/D out of range The values from the A/D converter are outside the normal range. Remove the weight from the scale if overloaded. Make sure the pan is attached. Indicates the load cell or the electronics may be faulty. If the problem persists contact your dealer or Adam Equipment for assistance. Err 9 Check weigh limits error Shown if the low limit is set higher than the current high limit. Reset High limit or different low limit. FAIL Calibration error. Improper calibration (should be within + 10% of the factory calibration). The old calibration data will be retained until the calibration process is complete. If the problem persists contact your dealer or Adam Equipment for assistance.