Home > Adam Equipment > Indicator > Adam Equipment AE Indicators AE402REVB1 User Manual

Adam Equipment AE Indicators AE402REVB1 User Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Adam Equipment AE Indicators AE402REVB1 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 35 Adam Equipment manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							29 | P a g e              © Adam Equipment Company 2010  
     Description  ENGLISH  FRENCH  GERMAN   SPANISH 
    Net weight    Net Wt.  Pds Net  Net-Gew  Pso Net 
    Weight  per  unit  
    counted   Unit Wt.  Pds unit  Gew/Einh  Pso/Unid 
    Number  of  items  
    counted   Pcs  Pcs  Stck.  Piezas 
    Number  of  
    weighing    added 
    to subtotals  No.  Nb.  Anzhl  Num. 
    Total  weight  and  
    count printed   
    Total  Total  Gesamt  Total 
    Print date   
    Date  Date  Datum  Fecha 
    Print time   Time  Heure  Zeit  Hora 
    Scale ID number   Scale ID  Bal ID  Waagen ID  Bal ID 
    User ID Number   User ID  Util ID  Nutzer ID  Usuario ID 
     
      
    9.1  INPUT
     COMMANDS FORMAT 
     
    The  scale  can  be  controlled  with  the  following  comm ands.  Press  the [Enter]  key  of 
    the PC after each command.     
    T   Tares  the  scale  to  display  the  net  weight.  This  is 
    the  same  as 
    pressing  [Z/T]. 
     
    Z     Sets the zero point for all subsequent weighing. Th
    e display shows 
    zero.  
    P      Prints  the  results  to  a  PC  or  printer  using  the  RS-
    232  interface.  It 
    also  adds  the  value  to  the  accumulation  memory  if  t he 
    accumulation function is not set to automatic. 
     
      
    						
    							30 | P a g e              © Adam Equipment Company 2010  
     
    10.0  RELAY INTERFACE 
       
    The AE 402 indicator is supplied with drivers to control external relays.  The drivers 
    could be used to control a number of different rela ys depending upon the users 
    need.    The relay drivers are isolated outputs req uiring the use of an external power 
    supply for the relays.    Also see section 3.4.3.   
    Contact Adam Equipment or your supplier for a relay  interface that is compatible 
    with the relay drivers, however other interfaces ca n be used as long as the following 
    conditions apply.    
    Connections to the drivers:  
     
      
    Connections are made to the PCB, Connector P1.   Th e connector is a spring 
    activated type, simply press on the top of the conn ector and insert the wire. 
      
    Do not exceed the safe limits of voltage or current  of 24VDC or 15ma per output. 
      
    Depending upon the application it may be necessary  to use a small relay to drive 
    larger relays, or to provide additional protection  to prevent electromagnetic 
    interference (diodes as shown above) to this or oth er machinery.  
       
    						
    							31 | P a g e              © Adam Equipment Company 2010  
     
    11.0  CALIBRATION   
    The scale can be calibrated using the following procedure.     To enter this procedure 
    it is necessary to use Func 4 accessible using the  [Func/C] key as described in section 
    7.4   or using the password access as described in  section 12. 
     
    The AE 402 scales calibrate using either metric or  pound weights, depending on the 
    weighing unit in use before calibration.   The disp lay will show either "kg" or "lb" to 
    identify the weights expected. 
     
      
    PROCEDURE
     
     
       Enter the calibration section using Func 4, C8 CAL  or using the password 
    as described in section 12.    
     
       The display will show " unLoAd". 
     
      Remove  any  weight  from  platform,  when  the  stable  sy mbol  is  on  press 
    [Z/T] . 
     
      The  display  will  show  " Ld"  the  “ 0000XX ”  to  show  the  last  calibration 
    weight used.    Place this calibration weight on th e scale. Press the [Z/T] 
    key.      If  the  weight  you  put  on  the  scale  does  not  match  the  value 
    displayed,  press  the  [Func/C] key  to  clear  the  vale  then  use  the 
    [Unit/ 
    
    
    ]key  and  [Pcs/
    
    
    ]  key  to  set  the  correct    value.    When  it  is 
    correct press  [Z/T]. 
     
      If the calibration is acceptable the scale will run  a self-test during which 
    the  calibration  weight  should  be  removed.   If  an er ror  message  “FAiL L” 
    is  shown  try  calibration  again  as  a  disturbance  may   have  prevented  a 
    successful calibration.   
     
    After  calibration  the  scale  should  be  checked  to  ve rify  the  calibration  is  correct.  If 
    necessary repeat calibration, make sure that the sc ale is stable before accepting any 
    weight.  
    						
    							32 | P a g e              © Adam Equipment Company 2010  
     
    12.0  SERVICE SECTION      
      
    The  scales  will  allow  entry  to  the  parameters  if  the [Tare]   key  is  pressed  during  the 
    power on cycle.   The passwords work as above.  
    In  this  case  the  display  will  show  the  passcode  req uest  screen,  “  P  -  -  -  -  “ .    To 
    continue enter a passcode as described below.    
    Entering passcode 0000 will allow calibration as sh own in section 11. 
     
    Entering  1000  will  allow  access  to  a  limited  set  of   parameters  described  in  section 
    12.1.   
    Entering  the  passcode  2006  will  allow  access  to  all   parameters  as  shown  in  section 
    12.3.   
    12.1
       USING “1000” TO ENTER THE SERVICE PARAMETERS 
    Press the  [Z/T] key during the display counting when turned on,   
      
    When  “Pn    “  is  displayed,  enter  the  number  1000  using  the  [Unit/
    
    
    ] and   [Pcs/ 
    
    
    ] 
    keys and  then press  [Z/T]. 
      
    The displays will show the first parameters, called  “F4 Int”. 
        
    To  select  another  parameter  press  the  [Pcs/
    
    
    ]  key  to  advance  through  the 
    parameters.    
    Press the  [Z/T] key to enter a parameter.   
      
    To exit a parameter, press the  [Print/M+/Esc] key.   
     
    The display will show the parameter number and a na me.   
    When  a  parameter  is  entered  by  pressing  the   [Z/T]  key,  the  displays  will  guide  you 
    through the parameter selected and the options avai lable.  
    						
    							33 | P a g e              © Adam Equipment Company 2010  
    The parameters available are:  
     “F4 Int”        Initial Zero Range  
    “F5 rEZ”        Re-Zero range  
    “F6 SCS”        Successive Tare Enable 
    “F7  Cnt”        Display the A/D counts 
    “F8 Zem”       Zero Mode   
    “F9 Lvd”        Low voltage detection   
    “FA AZn”         Auto Zero Range 
    “Fb FPS”        User Function Password 
      
    The description of the parameters is shown in secti on 12.3 
     
    12.2
       USING “2006” TO ENTER THE SECURE PARAMETERS 
    Press the  [Z/T] key during the display counting when turned on,   
      
    When  “Pn    “  is  displayed,  enter  the  number  2006  using  the  [Unit/
    
    
    ] and   [Pcs/ 
    
    
    ] 
    keys  and   then press  [Z/T].        The displays will show  the first parameters,  c alled “F1”  
    “CAL” .   
    To  select  another  parameter  press  the  [Pcs/
    
    
    ]  key  to  advance  through  the 
    parameters.      
    Press the  [Z/T] key to enter a parameter.   
     
    To exit a parameter, press the  [Print/M+/Esc] key.   
      
    The display  will show the parameter number and a n ame.   
    When  a  parameter  is  entered  by  pressing  the   [Z/T]  key,  the  displays  will  guide  you 
    through the parameter selected and the options avai lable. 
    The parameters available are:  
    “F1 CAL”        To enter the Calibration 
    “F2 dEC”        Decimal Point Position 
    “F3 CAP”         Default Weighing Unit and capacity 
    “F4 Int”        Initial Zero Range  
    “F5 rEZ”         Re-Zero range  
    “F6 SCS”        Successive Tare Enable 
    “F7  Cnt”        Display the A/D counts 
    “F8 Zem”       Zero Mode   
    “F9 Lvd”        Low voltage detection   
    “FA AZn”         Auto Zero Range 
    “Fb FPS”        User Function Password  
    						
    							34 | P a g e              © Adam Equipment Company 2010  
      
    12.3
       PARAMETER DETAILS  
      
    12.3.1  F1 -CALIBRATION 
     
    To  enter  the  calibration  parameter,  press  the   [Z/T] key  when  “F1”  is  displayed.    The 
    scale  will  be  calibrated  using  2  masses  of  approxim ately  1/3Maximum  and 
    Maximum.  If  the  scale  has been calibrated once  the   values  will  be  stored.   If  this  is 
    the  first  time  the  scale  is  calibrated  the  user  mus t  enter  the  values  for  the 
    calibration masses.    
    NOTE:    It  is  necessary  to  set  the  decimal  point  loc ation  and  the  capacity  before 
    calibration is possible.     
    The  display  will  instruct  you  to  remove  any  weight  from  the  scale, “UnloAd”.  Press 
    [Z/T] . 
     
    The  display  will  tell  you  to  add  the  first    weight  to  the  scale:  “Ld  1”  “        10  kg”      If 
    necessary  change  the  value  shown  to  match  the  weigh t  to  be  used.  Press [CE] to 
    clear  the  old  value  and  then  enter  the  new  value.    All  values  entered  are  in  whole 
    numbers only.    
    Add the weight shown, wait for stability then press  the [Z/T]  key. 
      
    The display will tell you to add the second   weigh t to the scale: “Ld 2 “   “    30 kg” 
      
    Either add the weight shown, wait for stability the n press the [Z/T] key or press the 
    [Print/M+/Esc]  key to accept the single point calibration and ret urn to weighing.   
      
    The display will show  “PASS” if the calibration is OK.  
      
    Remove the weight.   
    12.3.2  F2–DECIMAL POINT POSITION 
     
    To  set  the  value  for  the  decimal  point  location.        The  options  are    0,  0.0,    0.00,  
    0.000,  0.0000    
    To enter this parameter, press the  [Z/T] key when  “F2 dEC” is shown. 
    The display will show the current setting.  
    Press the  [Pcs/
    
    
    ] key to change the value. Select from 0, 0.0,  0.00,   0.000,  0.0000   
     
    Press  [Z/T]  to accept the displayed value.  
    						
    							35 | P a g e              © Adam Equipment Company 2010  
      
    12.3.3   F3 – CAPACITY/ WEIGHING UNIT/INCREMENT 
     
    To enter this parameter, press the [Z/T] key when  “F3 CAP” is shown. 
    The display will show the current capacity and defa ult weighing unit.   
    Enter  the  numeric  values  using  the  [Unit/
    
    
    ] and   [Pcs/ 
    
    
    ] keys.        The  scale  will 
    check the number of division’s n = maximum/incremen t is less than 30,000 divisions.  
    (3000 divisions for EC approved versions).    
    To  select  the  weighing  unit  press  the  [Unit/
    
    
    ]until  the  default  weighing  unit  is 
    flashing, either kilogram (kg) or pounds (lb).     
    Press  [Z/T]  to accept the displayed value. 
      
    On non-approved versions the display then lets you  select the increment, “InC 2” 
    For example 100kg x 0.01kg  the  increment  is 10grams , but  the  last digit  increments 
    by 1.   
    The display will  show  the  current  increment  value  a s  used  with  the current  decimal 
    point position.    
    Press the  [Pcs/
    
    
    ] key to change the value. Select from 1, 2, 5, 10,  20 or 50 
     
    Not all increments may be available for the capacit y you selected. 
     
    For EC Approved versions the scale will determine th e increment that maintains the 
    number of divisions to be 3000 or less.  
    Press  [Z/T] to accept the displayed value. 
    Press  [Print/M+/Esc]  to return to weighing. 
        
    						
    							36 | P a g e              © Adam Equipment Company 2010  
    12.3.4   F4 –INITIAL ZERO RANGE    
    To enter this parameter, press the [Z/T] key when  “F4 int” is shown. 
    The display will show the current initial zero rang e.      
    Press the  [Pcs/] key to change the value and press  [Z/T] to accept the value. 
    Press  [Print/M+/esc]  to return to weighing. 
      
    12.3.5   F5 -RE-ZERO RANGE 
     
    To enter this parameter, press the  [Z/T] key when  “F5 rEZ” is shown. 
    The display will show the current re-zero range.       
    Press the  [Pcs/
    
    
    ] key to change the value. 
    Press  [Z/T] to accept the value. 
    Press  [Print/M+/esc]  to return to weighing. 
      
    12.3.6   F6 -SUCCESSIVE TARE  
     
    To enter this parameter, press the  [Z/T] key when “ F6 SCS” is shown. 
    The display will show if the successive tare is on  or off. 
    Press the  [Pcs/
    
    
    ] key to change the value. 
    Press  [Z/T]  to accept the displayed value. 
    Press  [Print/M+/Esc]  to return to weighing. 
     
    12.3.7   F7 –ADC COUNTS  
     
    To enter this parameter, press the  [Z/T] key when  “F7 Cnt” is shown. 
    This  parameter  allows  you  to  view  the  A/D  counts  fr om  the  internal  A/D  converter. 
    This can be an aid to service.  
    Press the  [Z/T] key to return to the PARAMETER menu. 
    Press the  [Print/M+/Esc]  key to return to weighing. 
    Typical value at zero is 30,000-90,000 (approx.).  
    Typical value at full capacity is 500,000 (approx.) . 
      
    12.3.8   F8 –ZERO MODE  
     
    To enter this parameter, press the  [Z/T] key when “ F8 ZEm” is shown. 
    Select  the  Zero  mode  desired.  In  all  but  special  ca ses  Zero  Mode  1  is  used.      The 
    other 2 zero modes are for unique locations in the  world and effect the +/- range of 
    the zero.      
    Press the  [Pcs/
    
    
    ] key to change the value. 
    Press  [Z/T]  to accept the displayed value. 
    Press  [Print/M+/Esc]  to return to weighing. 
      
    						
    							37 | P a g e              © Adam Equipment Company 2010  
    12.3.9    F9 –LOW VOLTAGE DETECTION     
    This parameter allows detection of low voltage when the battery wears down. 
    To enter this parameter, press the  [Z/T] key when  “F9 LVd” is shown. 
    The display will show if the LVD Mode is set to on  or oFF.     
    Press the  [Pcs/
    
    
    ] key to change the value. 
    Press  [Z/T] to accept the displayed value. 
    Press  [Print/M+/Esc]  to return to weighing. 
      
    12.3.10    FA –AZn Auto Zero Range   
    This parameter sets the auto zero range from 0 (Off ) to 4d. 
    Calibration  and  parameter  counting  function  to  be  a ctive.  To  enter  this  parameter, 
    press the  [Z/T] key when  “FA  AZn” is shown. 
      
    The display will show if the current Auto Zero rang e.  
               
    Press the  [Pcs/
    
    
    ] key to change the value. 
    Press  [Z/T] to accept the displayed value. 
    Press  [Print/M+/Esc]  to return to weighing. 
     
    12.3.11    Fb –FPS User Function Password   
    This  parameter  set  a  password  to  control  access  to  the  user  functions  as  described 
    in  Section  7.    If  the  password  is  set  to  0000  the  c ontrol  is  disable  and  the  user  has 
    free access to the functions.    
    To enter this parameter, press the  [Z/T] key when  “Fb  FPS” is shown. 
      
    The  display  will  show  “ P1 - - - -“  Enter  the  new  password  code  followed  by  [Z/T] to 
    accept.    
    The display will then show “ P2 - - - -“ Enter the new password code again followed by 
    [Z/T]  to  accept.    The  same  code  number  must  be  entered  bo th  times  for  the 
    password to be set.   
              
    Press the  [Unit/
    
    
    ] and  [Pcs/ 
    
    
    ] keys to enter the values. 
    Press  [Print/M+/Esc]  to return to weighing. 
     
      
    						
    							38 | P a g e              © Adam Equipment Company 2010  
     
    13.0  ERROR CODES 
     
    ERROR  
    CODES 
    DESCRIPTION SUGGESTIONS 
    - -oL - - Over-range Remove weight from the scale.   
     
    If  the  problem  persists  contact  your  dealer  or  Adam 
    Equipment for assistance. 
      
    Err 1 Time Setting Error Enter  time  using  correct  format  and  reasonable  values. 
    Format: hh:mm:ss 
     
    Err 2 Date Setting Error Enter  date  using  correct  format  and  reasonable  values. 
    Format: yy:mm:dd 
     
    Err 4 Zero Setting Error The  scale  was  outside  the  normal  zero  setting  range  
    either  when  it  was  turned  on  or  when  the [Zero]  key 
    was pressed.   
    Remove weight from the scale and try re-zeroing aga in. 
    Use the [Z/T] key to set the display to zero value. 
    If  the  problem  persists  contact  your  dealer  or  Adam  
    Equipment for assistance. 
     
    Err 6 A/D out of range The  values  from  the  A/D  converter  are  outside  the 
    normal range.     
    Remove the weight from the scale if overloaded.  
    Make sure the pan is attached.  
    Indicates the load cell or the electronics may be faulty.   
    If  the  problem  persists  contact  your  dealer  or  Adam  
    Equipment for assistance. 
     
    Err 9 Check weigh 
    limits error 
    Shown  if  the  low  limit  is  set  higher  than  the  current 
    high limit.  Reset High limit or different low limit. 
     
    FAIL  Calibration error. Improper  calibration  (should  be  within  +  10%  of  the 
    factory  calibration).  The  old  calibration  data  will   be 
    retained until the calibration process is complete. 
    If  the  problem  persists  contact  your  dealer  or  Adam  
    Equipment for assistance. 
     
      
    						