    							19 | P a g e              © Adam Equipment Company 2010  
    6.8.1 ANIMAL WEIGHING PROCEDURE     With  the platform empty  the display  will  show  zero w eight.   Place  containers 
    or blankets onto the platform and press the  [Z/T] key to zero the display.   The 
    scale  may  go  into  the  animal  (dynamic)  weighing  pro cedure  when  the  items 
    are  placed  on  the  scale  but  will  return  to  showing  zero  when  the  [Z/T] key  is 
    pressed.     
     
      Place the animal or sample to be weighed on the pla tform.    
       
      The display will show the Animal/Dynamic weighing s ymbol 
     until a stable 
    weight is determined.  The time it takes for the st able value will depend upon 
    the setting of the filter parameter in the first st ep.   
       
      When  a  stable  reading  is  found,  the  display  will  sho w  this  value,  and  the 
    display will be locked until the  [Z/T] key is pressed.  The display will show the 
    “Hold”  symbol while the display is locked. 
     
     
     
       To weigh a second subject press the  [Z/T] key if necessary to zero the display, 
    and  place  the  next  animal  on  the  scale.  The  scale  w ill  detect  the  new  weight 
    and hold it as before. 
     
      The  scale  will  remain  in  the  animal  weighing  mode  u ntil  the [Unit/
    
    
    ] key  is 
    pressed. Then it will return to normal weighing.   
            
    						
    							20 | P a g e              © Adam Equipment Company 2010  
     
    7.0  USER PARAMETERS 
    Pressing  the  [Func/C]  key during  normal  operation  allows  the  user  to  access  the 
    parameters for customizing the scale. The parameter s are split into 4 groups-  
      
    1. Check weighing parameters,   
    2. Percent and Animal Weighing Functions  
    3. RS-232 parameters and  
    4. Scale parameters   
      When  [Func/C]  is  pressed,  display  will  first  show  “FunC 1” for  Check  weighing 
    parameters.  
     
      Press  either  the  [Func/C]  key  or  the    [Pcs/
    
    
    ] to  advance  through  the  groups 
    “FunC  1” , “FunC  2”  , “FunC  3”  and “FunC  4” .  Press [Z/T] to  enter  the  desired 
    group of parameters.  
      
      When  in  one  of  the  sections  press  [Print/M+/Esc]  to  return  to  the  group 
    “FunC  1” .  If  you  press  [Print/M+/Esc]  again,  the  scale  will  exit  the  User 
    Parameter section and return to normal weighing. 
        
    7.1  CHECK
     WEIGHING PARAMETERS 
     
       Press  [Z/T]  to enter the group.  
     
      Use  the  [Unit/
    
    
    ] key  and  [Pcs/
    
    
    ]  key  to  set  the  values  or  just  the  [Pcs/
    
    
    ] 
    key to select the options. 
     
      Press  [Z/T]  to confirm the change and then advance to the next  parameter by 
    pressing the  [Pcs/
    
    
    ] key. 
     
    This group of parameters-  -   Set low and high limits for check-weighing 
    -   Enables or disables the check weighing alarm  
    -   Enables or disables the negative check weighing   
    						
    							21 | P a g e              © Adam Equipment Company 2010  
      
    Parameter  Description  Options  Default setting 
    F1  Lo Set Low limit.    Use  the 
    [Unit/
    
    
    ]  key 
    and  [Pcs/ 
    
    
    ]  key  to  set 
    the  values  of  the  lower 
    limit.  
    When  set  press  the  [Z/T] 
    key  to  store  the  value  
    and go to  F2 Hi 
      000000 
    F2  Hi 
    Set High limit.    Use  the 
    [Unit/
    
    
    ]  key 
    and  [Pcs/ 
    
    
    ]  key  to  set 
    the  values  of  the  high 
    limit.  
    When  set  press  the  [Z/T] 
    key  to  store  the  value 
    and go to  F3 bEP 
      0000000
     
    F3  bEP  This  parameter  sets  the  
    Beeper  to  off  or  on.  If  set  to 
    on,  the  beeper  can  further 
    be  set  to  sound  when  the 
    weighing  result  is  within  or 
    outside  the  check-weighing  
    limits.    bP oFF  - Off 
    bP inL   - Within limits 
    bP  otL  -  Outside  limits 
    (>20d)  
    bP inL
     
    F4  nCK  This  parameter  enables  
    negative  check  weighing 
    function  with  ability  to  do 
    negative tare.    on  
    oFF   
    on 
      
    7.2  PERCENT
     WEIGHING AND ANIMAL WEIGHING  
     
    See section 6.7 and 6.8   for details of these spec
    ial weighing modes. 
       
      Press  [Z/T]  to  enter  a  parameter.  Use  the  [Unit/
    
    
    ] key  and  [Pcs/
    
    
    ]  key  to 
    set the values or just the  [Pcs/
    
    
    ] key to select the options. 
     
      Press  [Z/T] to confirm the change and then advance to the next  parameter by 
    pressing the  [Pcs/
    
    
    ] key.  
    						
    							22 | P a g e              © Adam Equipment Company 2010  
        Press  [Print/M+/Esc]  to  return  to  the  group  “FUnC  2”.  If  you  press  
    [Print/M+/Esc]  again,  the  scale  will  exit  the  User  Parameter  secti on  and 
    return to weighing. 
        
    Parameter  Description  Options  Default  setting 
    P1   PCt   This  parameter  allows  the  
    user  to  enter  the  Percent 
    weighing  Function.  See 
    Section 6.7.    None  Enabled 
    always 
    P2 AnL Enter  the  Animal  Weighing  
    mode  of  operation,  See  
    section 6.8  Set the filter value.  Enabled 
    Always 
              
    7.3  RS-232 PARAMETERS 
     
      Press  [Z/T]  to  enter  a  parameter.  Use  the  [Unit/
    
    
    ] key  and  [Pcs/
    
    
    ]  key  to 
    set the values or just the  [Pcs/
    
    
    ] key to select the options. 
     
      Press  [Z/T] to confirm the change and then advance to the next  parameter by 
    pressing the  [Pcs/
    
    
    ] key. 
         Press  [Print/M+/Esc]  to  return  to  the  group  “FUnC  3”.  If  you  press  
    [Print/M+/Esc]  again,  the  scale  will  exit  the  User  Parameter  secti on  and 
    return to weighing. 
    This group of parameters can be set by the user for  setting the RS-232 active or not, 
    baud rate, printing mode, accumulation mode, RS-232  language, and user or scale ID 
    numbers.    
     Parameter  Description  Options  Default  Values  or setting 
    C1  on Enable  or  disable  the  
    RS-232 interface  Prt on   
    Prt oFF
      Prt on 
    C2  bd Baud Rate   
     
    600  
    1200  
    2400  
    4800 
    9600
      
    19200   9600   
    						
    							23 | P a g e              © Adam Equipment Company 2010  
    C3 PrM Printing  Mode-  
    Manual, Continuous or  
    Automatic  mAn
    ,  
    Cont 
    (not  on  EC 
    approved scales) 
    AUto     mAn
     
    C4 Aon  Enable  or  disable  the  
    Accumulation   AC on
      
    AC oFF   AC on
     
    C5 Ln  Select Language 
    EnGLi (English)   
    FrEnCH  (French)  
    GErmAn  (German)      
    SPAn  (Spanish)   EnGLi 
     
    C6  UId  Set User ID  To  be  entered 
    manually  000000 
    C7   Sid 
    Set Scale ID  To  be  entered 
    manually  000000 
      
    Scale will perform the following, depending on the 
    Accumulation and Print Settings: 
     
    ACCUMULATION 
    SETTINGS
     
     
    PRINT SETTINGS 
    AC on AC Off 
    AUto  
      Accumulate  and  print  
    automatically    Print automatically,  
    Do not accumulate 
    mAn 
      Accumulate  and  Print  only  
    when 
    [Print/M+/Esc]   key 
    pressed.  If  [Print/M+/Esc]  is 
    pressed  a  second  time  only 
    print the weight.  Print  when
      [Print/M+/Esc]  key  is 
    pressed,  
    Do not accumulate. 
    Cont 
    Not  available  on  
    approved scales  Print continuously.  
    Accumulate when  
    [Print/M+/Esc]
     key is pressed 
     
      Print continuously.  
    Do not accumulate. 
     
    7.4  SCALE PARAMETERS  
     
    
      Press  [Z/T] to view the list of parameters.  
       Press  [Z/T]  to  enter  a  parameter.  Use  the  [Unit/
    
    
    ] key  and  [Pcs/
    
    
    ]  key  to 
    set the values or just the  [Pcs/
    
    
    ] key to select the options. 
       Press  [Z/T] to confirm the change and then advance to the next  parameter by 
    pressing the  [Pcs/
    
    
    ] key. 
       Press  [Print/M+/Esc]  to  return  to  the  group  “FUnC  4”.  If  you  press 
    [Print/M+/Esc]  again,  the  scale  will  exit  the  User  Parameter  secti on  and 
    return to normal weighing.  
    						
    							24 | P a g e              © Adam Equipment Company 2010  
    This group of parameters is used to control the operation of the scale.   
      
    Parameter  Description  Options  Default   setting 
    S1  Un Enable or disable weighing  
    units,  will  not  allow  to 
    disable  all  units,  at  least 
    one  has  to  be  enabled. 
    Parts  counting  can  be  
    enabled/disabled    Kg  
    Grams   
    lb  
    oz 
    lb:oz 
    N (Newtons)  
    PCS Kg     
     
    S2  bL  Backlight set to always on,  
    always off or automatic on 
    whenever  a  weight  is 
    placed or a key is pressed  EL   oFF
      
    EL   on   
    EL   AU  EL   AU
     
    S3  AoF  Auto  Off-  Disable  or  set  
    time  increment  to  turn off  
    scale  SLP   0  
    SLP   1 
    SLP   5  
    SLP   10   SLP   0  
    S4   dt 
    Set  Time  and  Date  format  
    and settings.  
    Format  for  date  can  be  
    changed  when  the  display  
    shows  mmddyy,  ddmmyy 
    or  yymmdd  by  pressing 
    the
      [Pcs/ 
    
    
    ]  key,  then 
    enter the date.   Enter  the  time  
    manually.    
    Enter  the  date  format  
    and  then  the  numeric 
    value manually.     
    00:00:00    
    mm:dd:yy
     
    S5  diS  Display  all  weights  or  only  
    when stable    ALL 
    StAb   ALL 
     
    S6  Fi  Filter  setting  to  slow,  
    normal or fast  SLow  
    nor  
    FASt   nor   
    S7 SPS 
    Password   Enter  new  pass  word 
    twice,,  controls  access  
    to Func 1 to Func 4 Not active 
    S8 CAL 
    Calibration  Calibrate  the  scale. 
    See Section 11.0  - 
                
    						
    							25 | P a g e              © Adam Equipment Company 2010  
     
    8.0  BATTERY OPERATION 
     
      The scales can be operated from  the  battery  if  desi red.  The battery  life 
    can  be  up  to  70  hours  depending  on  the  load  cells  u sed  and  how  the 
    backlight is used.  
     
      When the battery needs charging a symbol on the dis play will show less 
    bars  in  the  battery  symbol.      The  battery  should  be   charged  when  only 
    the battery outline is on.  Once the bars have been  turned off the scale 
    will still operate for a short time after which it  will automatically switch 
    off to protect the battery. 
     
      To  charge  the  battery,  simply  plug  the  adaptor  into   the  mains  power. 
    The scale does not need to be turned on. 
     
      The battery should be charged for 12 hours for full  capacity. 
     
      Near  the  display  is  an  LED  to  indicate  the  status  o f  battery  charging.  
    When the scale is plugged into the mains power the  internal battery will 
    be charged.  If the LED is green the battery has a  full charge.  If it is Red 
    the  battery  is  nearly  discharged  and  yellow  indicat es  the  battery  is 
    being charged. 
      
    						
    							26 | P a g e              © Adam Equipment Company 2010  
    31
    2
    4
          
    9.0  RS-232 INTERFACE 
       
    The  AE  402  indicator  is  supplied  with  bi-directiona l  RS-232  interface  as  standard. 
    The  scale  when  connected  to  a  printer  or  computer  o utputs  the  weight  with  the 
    selected weighing unit through the RS-232 interface . 
      Specifications: 
                        
      RS-232 serial interface is a plug as figure 2 shows : 
     
      
        1: Pin GND,   Signal Ground 2: Pin RXD,    Received Data 
              3: Pin TXD,    Transmitted Data               
     
      As viewed from the back of the indicator 
        
    The scale can be set to print text in English, Fren ch, German or Spanish.  See the RS-
    232 parameters section for details. 
    RS-232 output of weighing data 
    ASCII code  
    9600 Baud (user selectable)  
    8 data bits  
    No Parity  
    						
    							27 | P a g e              © Adam Equipment Company 2010  
      
    Data Format-Normal Output:  
    Only  weight  value  along  with  the  weighing  unit  is  printed.    If  Percent  weighing  is 
    used then % is shown in place of weighing units.   
    Data Format-Parts Counting Output: 
    Weight, Unit weight and number of parts are printed.  
     
     
      
    Date           12/09/2006   
    Time           14:56:27     
     
    Scale ID    123456    
    User ID    234567    
      
    Net Wt.        1.234 Kg           Net Wt. (or Gross Wt.) 
    Unit Wt.         123 g            g for metric and lb for pounds 
    PCS               10 pcs    
                         
      
      
       
     
      
    Date           12/09/2006   
    Time           14:56:27     
      
    Scale ID    123456                If ID is zero, it is left blank 
    User ID    234567    
      
    Net Wt          1.234 Kg               Net Wt. (or Gross Wt.) 
                          
      
      
      
    						
    							28 | P a g e              © Adam Equipment Company 2010  
      
    Data Format- Memory Recall Output:  
     
    Data Format- Continuous Output- Normal weighing:   
     
      
    Data Format- Continuous Output- Parts Counting: 
     
     
    NOTE: 
    1. The  accumulated  total  will  not  be  sent  to  the  RS-23 2  when  the  continuous 
    print is turned on.   
      2. The continuous print will only be for the current w eight and the display data.  
      3. In other languages the format is the same but the t ext will be in the language 
    selected. 
      
    Date            12/09/2006   
    Time            14:56:27    
      
    Scale ID     123456    
    User ID     234567   
      
    ------------------                                       
    TOTAL  
    No.               5         
    Wt.           1.234 Kg      
       
    PCS              10 pcs     
    ------------------  
                     
    ST,GROSS  1.234 Kg       ST or US  for STable or UnStable,  
    US,NET    0.000 Kg                 NET or GROSS for Net Weight   
    or Gross wt.  and the weighing unit, kg, lb etc. 
    ST Net    1.234 Kg            Net Weight (or Gross wt.) 
    U.W.      123 g               Kg and g for metric and Lb for pounds 
    PCS       10 pcs   
                                  
      
    						