USER F8) To enter a calibrationweight other than capacity, press USER. T h e display far left digit will flash zero indicating that a number should be entered. blinking Error Correction: If you input a wrong value, press ZERO to step back one digit and reenter. USER FUSER F TARE USER F TARE TAREUSER F TAREUSER F 5 times USER F TARE fixed blinking blinking blinking blinking blinkingblinking fixed fixed fixed3DVVTARE 9) Press the USER key to scroll the number and...
Page 32
14) The display now readsLoAd3 or LoAd4. Repeat steps 12 and 13 for additional span points press TARE or press ZERO if you are finished. After LoAD4 has passed, the 3460 will automatically advance to the next step. /RDG /RDG or TARE To enter another span point 15) The display shows CAL’d to indicate that the calibration was successful. 16) Press TARE. The display flashes C-CAL followed by the C-CAL number. Make a note of this number. If the X1000 annunciator is lit, also make a note of that. This...
Page 33
Calibration 29 5.1 Initial Calibration Use this procedure only if the capacity and count by (d) needs to be modified. The initial steps of the initial calibration will totally erase user setups as well as any previous calibration. 1) Turn the 3460 off. 2) Remove the seal screw. With a small screwdriver press and hold the CAL switch behind the seal screw. 3) Still holding down the CAL switch, press the POWER switch. 4) The display blinks rESEt. 5) To reset all calibration constants and...
Page 34
30 MSI-3460 Operator’s Manual Figure 5-5. 14) The first digit blinks. Use the USER key to scroll through the numbers. When the desired number is shown, push TARE. In this example, we’ll enter 2500 as a capacity. USER FUSER F TAREUSER Ffixed blinking blinking blinking TARE TAREUSER F TAREUSER F 5 times USER F TAREblinking blinking blinkingfixed fixed fixed Continue inputing the desired capacity using the USER key for scrolling the number and the TARE key to...
Page 35
Calibration 31 Due to LFT requirements and general scale de sign criteria, the weight must be stable for certain features to work: ZERO – weight must be stable to be zeroed. TARE – weight must be stable to be tared. TOTAL – weight must be stable to be added to the total registers. One way to improve the stability is to increase the filtering, at the risk of increasing settling time. The other is to increase the ‘d’ (reduce resolution). The third way is to increase the motion window. The...
Page 36
TARE XQ/G &&$/ TARE USER F 1)Remove the seal screw. With a small screwdriver press the cal switch behind the seal screw. The CAL setup menu appears. 2)Press the USER key to scroll to the C-Cal menu selection. 3)Press TARE to start the C- Cal procedure. If this calibration is using a cleared 3460, the INITIAL CALIBRATION procedure will start here. You must set the cal unit, capacity, and count- before proceding to step 4, 4)The display reads UnLd (unload) indicating you should remove all weight...
Page 37
TARE TARE TARE USER F 10) Press the USER key to scroll the number and the TARE key to enter each digit of the C-Cal value. In this example, we’ll enter the C- Cal value of 2503. However, it should be noted that most C-CAL values are 5 digits. Do not push the TARE key two times in a row until the entire number is entered. C-Cal’s lessthan 4 digits will not be accepted. If the number is too small the display will read “LittL” and goback to the “C-CAL” message. Error Correction during number entry: If...
Page 38
34 MSI-3460 Operator’s Manual 5.4 Calibration Setup Menu The calibration setup menu contains two items: standard and auto zero maintenance “Auto0.” In addition, more menus will appear that are transferred from the main setup menu when Legal for Trade settings are used. &$/ &&D/ 6HWYS 6WDQG $YWR LQGYV +E U XQLW 2Q 2IICal Standard Auto Zero Maintenance Calibrate with Weights Calibrate with Constant Industrial Setting (default) Handbook 44, NTEP R-76, OIML Industrial, restricted to 1...
Page 39
TARE If you are already in the Cal menu skip to step 2 1) Remove the seal screw. With a small screwdriver press the cal switch behind the seal screw. The CAL setup menu appears. 2) Press the USER key to scroll to Setup. 3) Press TARE to enter the cal setup menu. 4) Press TARE to enter the standard menu. The current standard setting is displayed. 5) Scroll to the desired standard pressing the USER key. 6) Press TARE to set the standard. The display goes to the next item in the CAL setup...
Page 40
TARE If you are already in the Cal menu skip to step 2 1) Remove the seal screw. With a small screwdriver press the cal switch behind the seal screw. The CAL setup menu appears. 2) Press the USER key to scroll to Setup. 3) Press TARE to enter the cal setup menu. 4) Press USER to scroll to the Auto0 menu. 5) Press TARE to enter the auto- zero menu. The current setting is displayed (blinking) 6) Change the setting between On or OFF pressing the USER key. 7) Press TARE to set the autozero. The...