MSI 3460 Challenger 3 Manual
Here you can view all the pages of manual MSI 3460 Challenger 3 Manual. The MSI manuals for Scale are available online for free. You can easily download all the documents as PDF.
Page 11
Introduction 7 The MSI-3460 scale has a safe mech anical overload of 200%, and an ultimate overload of 500%. Overloads great er than 500% may result in structural failure and dropped loads. Dropped loads may cause serious personal injury or death. OPTIONS Wireless Remote Contr oler 50 (15 m) typical range Light- of-Sight. Uses 418 MHz (USA) handheld transmitter. 802.15.4 RF Mode m for connectivity with MSI-8000 RF Remote Display Integral 802.15.4 RF Modem for co nnectivity to...
Page 12
8 MSI-3460 Operator’s Manual 1.5 Features • Designed to meet or exceed all U.S. and international standards. • Typically 50 hours of weighing time utilizing automatic sleep mode. • Automatic power off conserves battery life by sensing no activity after 15, 30, 45 or 60 minutes, determined by operator, and turns off power. • Automatic Sleep Mode preserves battery life by dimming the LED display after 5, 15, or 30 minutes of no activity. • Rugged construction throughout. Buttons are sealed and rated...
Page 13
Introduction 9 1.8 Assembly Battery Cotter Load Hook Clevis Clevis Figure 1-2. MSI-3460 Assembly 1. Slide hook clevis over load cell with open end of hook toward front of scale. 2. Align holes of clevis and load cell. 3. Slide the clevis pin through the clevis and load cell holes. 4. Lock clevis pin in place with cotter pin. Bend cotter pin. WARNINGScale will be unsafe for use if clevis pin is not properly secured with the cotter pin. 5. Slide battery pack into battery comp artment. The...
Page 14
10 MSI-3460 Operator’s Manual 1. 9 B a t t e r y P a c k The MSI-3460 is powered by a six volt rechargeable battery. The battery door is part of the battery pack. Access Door FastenerHandle Figure 1-3. Battery Pack Removal To Remove the Battery Pack 1. Turn the two fasteners on the access door counte r clockwise 1/4 turn, 2. Grasp handle and pull the battery pack straight out. The battery will auto matically disengage from its connectors. The battery will operate for up to 100 hours...
Page 15
Introduction 11 1.10 Battery Charger 1. Remove the battery from the scale. 2. Plug battery charger into an AC power r eceptacle. The input voltage is universal from 115/230 VAC, 50/60 Hz. If the power input plug doesn’t match, contact MSI for information on international plugs. 3. Slide the battery charger connector plat e over the top of the battery until the battery terminals mate with the charge r connectors, as seen through the two observation holes. 4. Approximately six to eight hours...
Page 16
12 MSI-3460 Operator’s Manual 2.0 Operation 2.1 Power To turn on the power. 1. Press POWER, the following will display in order: • LED will light all segments at full brightness as a display test. • Display brightness will change to the setting determined in the display menu. • Software version number will display. • The M SI-3460 is ready for use. 2.2 Zero Sets the zero reading of the Scale. Use the ZERO key to take out small deviations in zero when the scale is unloaded. See “TAR E” for...
Page 17
Operation 13 2.3 Tare Tare is typically used to zero out a known weight such as a packing container or pallet and display the load in NET weight. A tare value is entered by pressing the TARE key. The TARE function in the MSI-3460 is defined as a Tare-In, Tare-Out operation. The first press of the TA R E key stores the current weight as a tare value and then the scale subtracts the tare value from the gr oss weight and changes the display to NET mode. The next press of the TARE key will clear...
Page 18
14 MSI-3460 Operator’s Manual 3.0 User Defined Function Keys The following function descriptions are for optional user defined functions that are programmed on the front panel USER key or the two function keys (F1 & F2) on the RF remote control. The functions TO TAL, VIEW TOTAL, and NET/GROSS are available full time on the RF remote co ntrol. To enable any of the USER key functions, you must set up the USER keys following the procedures in Section 4.0. TEST The TEST function provides an LED test...
Page 19
User Defined Function Keys 15 AUTO TOTAL The USER key under the AUTO TOTAL mode functions as Auto Total On / Auto To t a l O ff . The auto mode has three variations which are programmed in the SETUP menu: •AutoLoad – Any settled weight above the ‘Rise above’ threshold will be automatically totaled. Then the scale must fall below the ‘Drop below’ threshold before another total is allowed. •AutoNorm – This mode takes the last settled weight to auto total with. The total occurs only once the scale...
Page 20
16 MSI-3460 Operator’s Manual 4.0 Configuration 4.1 Menu Map )YQF IYQF DII 6/((3 GLVSO 6W3W 6W3W 6W3W WRWDO ILOWU 81L7 2)) WHVW WRWDO YWWO 1(W U /HDUQ 3+OG 8QLW 2II II /R +L +L DYWR /R /R +L +L kg lb 2)) UHDW /(66 Number Entry Off Test Display *Total *View Total *Net/Gross Remote Learn **Units ***Peak Hold USER KEY FUNCTIONS TOTAL MODE Total Off (default) Total On Manual Autototal on Load Autototal Last Autototal on High Load 2II WWO2Q $/R$ $/$6W...