Getting started Vivid7/Vivid7 PRO Users Manual17 FC092326-03 S Video OutNTSC/PAL S- VideoVCR or video monitorSignal level 1Vp-p SVGA OutRGB high resolution videoComputer monitor USBUniversal serial busPrinter Ethernet10/100 Base-TX Ethernet IEEE 8023Network device Socket Signal type Device type Note
Page 32
Getting started 18Vivid7/Vivid7 PRO Users Manual FC092326-03 Switching On/Off To switch on the unit: 1. Switch on the circuit breaker on the rear of the unit (see Figure 1-4). 2. Press (on/off button) on the top left of the control panel (see Figure 1-4). After initialization the default scanning screen (2D mode) is displayed, the active probe being the one connected to the left most connector socket. Switching off the unit When the Vivid 7/Vivid 7 PRO is switched off, the system performs an...
Page 33
Getting started Vivid7/Vivid7 PRO Users Manual19 FC092326-03 In case of total lock- up of the system, hold the on/off but- ton down a few sec- onds to turn the system off.2. Select Shutdown. The shutdown process takes a few seconds and is completed when the control panel illumination is turned off. To switch off before moving the unit, follow the additional steps below: 1. Set the circuit breaker to OFF. 2. Remove the plug from the mains power socket. 3. Secure the unit power cable around the cable...
Page 34
Getting started 20Vivid7/Vivid7 PRO Users Manual FC092326-03 At the new location: 1. Secure the power plug in the wall outlet. 2. Switch on the circuit breaker on the rear of the unit (Figure 1-4). 3. Press (on/off button) (Figure 1-4). When turning on a system in standby it takes a few seconds before it responds. Do not push the on/off button again during this period (A second push will initiate a full shutdown). If the power cable is removed from the wall outlet the system will remain in Standby...
Page 35
Getting started Vivid7/Vivid7 PRO Users Manual21 FC092326-03 Moving and transporting the unit Wheels The wheels of the unit are controlled by the pedals situated between the front wheels of the unit (see Figure 1-5). Examine the wheels frequently for defects to avoid breaking or jamming. Figure 1-5: The Vivid 7/Vivid 7 PRO ultrasound unit pedalsWheel Characteristics Front Swivel, swivel lock and full lock Rear Swivel, but do not lock 1. Swivel lock pedal (Free position) 2. Full lock pedal (Free...
Page 36
Getting started 22Vivid7/Vivid7 PRO Users Manual FC092326-03 Wheel operation To engage the pedal in full lock To release the brake To engage swivel lock Pedal Action Press on pedal no. 2 Pedal Action Press on pedal no.1 Pedal Action Press on pedal no.1 12 12
Page 37
Getting started Vivid7/Vivid7 PRO Users Manual23 FC092326-03 To release swivel lock Moving the unit To prepare the unit to be moved 1. If not locked, push the keyboard console to its park position. 2. Turn the system off, including the circuit breaker (see page 18), and remove the plug from the wall. 3. Disconnect all cables linking the unit to any off-board peripheral devices and network. Note the marks on each cable to recon- nect them later.4. Secure the unit’s power cable. 5. Place all probes in...
Page 38
Getting started 24Vivid7/Vivid7 PRO Users Manual FC092326-03 bar and push or pull. Do not attempt to move the unit using cables or probe connectors. Take extra care while moving the unit on inclines. 3. Ensure that the unit does not strike the walls or door frames. 4. Ensure the pathway is clear. 5. Move the unit slowly and carefully. 6. Use two or more persons to move the unit over long distances or on inclines. Transpor ting the unit Take extra care when transporting the unit by vehicle. In...
Page 39
Getting started Vivid7/Vivid7 PRO Users Manual25 FC092326-03 brakes (see page 22). 2. Follow the installation procedure described on page 12. Unit acclimation time Following transport the unit may be very cold or hot. Allow the unit to acclimate before being switched on. Acclimation will take one hour for each 2.5 ºC increment when the unit’s temperature is below 10 ºC or above 40 ºC. ºC02.557.510354042.5 ºF32 36.5 41 45.5 50 95 104 108.5 Hours43210023 ºC45 47.5 50 52.5 55 57.5 60 ºF113 117.5 122...
Page 40
Getting started 26Vivid7/Vivid7 PRO Users Manual FC092326-03 System description System over view Figure 1-6: the Vivid 7/Vivid 7 PRO ultrasound system 1 2 3 4 5 6 78 9 10 11 12 13