GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 10-7 - Electrical Safety Tests 10-23 10-7-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the PM Inspection Certificate. Table 10-20 Maximum Allowance Limit for ECG Leakage Current AC Power Source Maximum Allowance Limit GROUND OPEN GROUND CLOSED Patient Lead to Ground Leakage Current Test and Patient Lead to Lead Leakage Current...
Page 382
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 10-24Section 10-7 - Electrical Safety Tests 10-7-9 Probe Leakage Current Test 10-7-9-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface. NOTE: Some leakage current is expected on each probe, depending on its design. Small variations in probe leakage currents are normal from probe to probe. Other...
Page 383
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 10-7 - Electrical Safety Tests 10-25 10-7-9-2 Test Equipment (cont’d) . Table 10-24 Test Equipment and Accessory Description Dale Part numberAccessory NamePictureDescription Vendor MODEL 600/100 46-285647P2 CHASSIS CABLE Black coil cord with extended length and black grips. Used on DALE601/601E to measure earth resistance and enclosure leakage current. Also used as reference lead for external measurement. Vendor MODEL...
Page 384
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 10-26Section 10-7 - Electrical Safety Tests 10-7-9-3 Generic Procedure for Leakage Current Measurements should be made with the ground open and closed, with power line mains polarity normal and reversed, and with the unit Off and On. For each combination, the probe must be active to find the worst case condition. 10-7-9-4 Meter Procedure Using Dale Meter to Measure Leakage Current The ultrasound probe’s imaging area is immersed in a...
Page 385
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 10-7 - Electrical Safety Tests 10-27 10-7-9-4 Meter Procedure Using Dale Meter to Measure Leakage Current (cont’d) 11.) Keep a record of the results with other hard copies of Planned Maintenance data. CAUTION Power off Equipment damage possibility. Never switch the Po larity or the status of the Neutral when the Ultrasound unit is powered on. Power off the Ultrasound unit, allow the stored energy to bleed down, and turn the...
Page 386
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 10-28Section 10-7 - Electrical Safety Tests 10-7-9-5 Meter Procedure Using Probe Adapter to Measure Probe Isolation (Sink) Current The Dale 600/600E provides a method for testing probes independently from the system. The meter utilizes a probe adapter to apply a test potential commonly to all connector pins. The ultrasound probe’s imaging area is immersed in a solution along with a grounding probe from the test meter to complete the...
Page 387
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 10-7 - Electrical Safety Tests 10-29 10-7-9-5 Meter Procedure Using Probe Adapter to Measure Probe Isolation (Sink) Current (cont’d) 8.) Test every probe and record in Table 10-26 the test results for every probe. The test passes when all readings measure less than the values in Table 10-30. 9.) Keep a record of the results with other hard copies of Planned Maintenance data. . Table 10-26 Probes Tested for Isolation (Sink)...
Page 388
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 10-30Section 10-7 - Electrical Safety Tests GE Healthcare Leakage Current Limits for LOGIQ™ 7 The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. Measurement limits per IEC 60601-1 Medical Equipment Safety Standards, Table IV. NOTE: *Measurement limits per IEC 60601-1 Medical Equipment Safety Standards, Table 19.5DV.1 **Most meters (like Dale 600/601) only measure AC (rms)....
Page 389
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 10-7 - Electrical Safety Tests 10-31 Section 10-1 When Theres Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent. Replace any defective part. Tighten all grounds. Ensure star washers are under all ground studs. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded and is...
Page 390
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 10-32Section 10-7 - Electrical Safety Tests PM INSPECTION CERTIFICATE * Scan Format: Phased Array, Linear Array, Curved Array, Mechanical Array or Other FUNCTIONAL CHECKS PHYSICAL INSPECTION AND CLEANING COMMENTS: ______________________________________________________________________________ ______________________________________________________________________________...