GE Logiq 7 Service Manual
Here you can view all the pages of manual GE Logiq 7 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 361
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 10-3 - Periodic Maintenance Schedule 10-3 10-3-1 How often should PMs be performed? (cont’d) Table 10-2 Periodic Maintenance Schedule Service at Indicated TimeDailyWeeklyMonthly Per Facilities QA Program Notes Clean Probes•** or before each use Clean Probe Holders • Clean Air Filter•more frequently depending on your environment Inspect AC Mains Cable•Mobile Unit Check Weekly Inspect Cables and Connectors• Clean Console• Clean...
Page 362
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 10-4Section 10-3 - Periodic Maintenance Schedule Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs. Table 10-3 Overview of GE-1 Tool Kit Contents Tool IDDescriptionTool IDDescription 9-45358 Pliers Retaining Ring 9-XL9971MM Xcelite-hex Blade 1.27mm 9-4078 Scribe 9-XL9972MM Xcelite-hex Blade 1.5mm 9-44572 Wrench Open End...
Page 363
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 10-3 - Periodic Maintenance Schedule 10-5 9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal 9-46696 Hex Keys 9-XL100X Wire Stripper/Cutter 5 inch - 100X 9-39829 Torpedo Level, Magnetic 9-XL87CG Pliers - very fine needle nose-87CG 9-38461Hammer, Ball Peen, 4 oz9-WEWDT-07Weller-Soldering-Replacement Tip(1) 9-4280Universal Joint 1/4 inch9-WS175-EWiss - Surgical Scissors 9-WEW60P3Weller - Soldering Iron, 3...
Page 364
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 10-6Section 10-3 - Periodic Maintenance Schedule 10-4-2 Special Tools, Supplies and Equipment 10-4-2-1 Specific Requirements for Periodic Maintenance See Chapter 7 Table 10-5 Overview of Requirements for Periodic Maintenance Tool Part NumberComments Digital Volt Meter (DVM) Electric Safety Analyzer DALE 600 46-285652G1 For 120V Unit Electric Safety Analyzer DALE 600E 46-328406G2 For 220V Units Leakage Current Ultrasound...
Page 365
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 10-3 - Periodic Maintenance Schedule 10-7 Section 10-5 System Periodic Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-6 System Preliminary Checks StepItemDescription 1 Ask & Listen Ask the customer if they have any problems or questions about the equipment. 2 Paperwork Fill in the top of the Periodic...
Page 366
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 10-8Section 10-3 - Periodic Maintenance Schedule 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary. 10-5-2-1 System Checks NOTE: * Some software may be considered standard depending upon system model configuration. Table 10-7 System Functional Checks ÷StepDescription B-Mode Verify basic B-Mode (2D) operation. Check the bas...
Page 367
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 10-3 - Periodic Maintenance Schedule 10-9 10-5-2-2 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. 10-5-3 Input Power 10-5-3-1 Mains Cable Inspection 10-5-4 Cleaning 10-5-4-1 General Cleaning Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks StepItemDescription...
Page 368
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 10-10Section 10-3 - Periodic Maintenance Schedule 10-5-4-2 Air Filter Cleaning \ NOTE: For your convenience or of the air filter is too dirty, replacement filters are available. Refer to Chapter 9 for the air filter replacement part number. 10-5-5 Physical Inspection Table 10-11 Air Filter Cleaning - frequency varies with your environment StepItemDescription 1Remove Filter Cover Refer to Chapter 8 for air filter location and...
Page 369
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 10-6 - Using a Phantom 10-11 10-5-6 Optional Diagnostic Checks To complete the PM checks, access the diagnostic software as described in Chapters 5 or 7. View the error logs and run desired diagnostics. 10-5-6-1 View the Logs 1.) Review the system error log for any problems. 2.) Check the temperature log to see if there are any trends that could cause problems in the future. 10-5-7 Probe Maintenance 10-5-7-1 Probe Related...
Page 370
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 10-12Section 10-7 - Electrical Safety Tests Section 10-7 Electrical Safety Tests 10-7-1 Safety Test Overview The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. They are intended for the electrical safety evaluation of cord-connected, electrically operated, patient care equipment. If additional information is needed, refer to the NFPA 99 (For...