Page 11
19 18 READ FUNCTIONS Select read by pressing the DISPLAY/SOURCE/READ pushbut- ton until the word READ appears on the LCD display. The READ functions measure the desired signal. Multiple scales are avail- able for some functions. MIN/MAX To read the Maximum or Minimum INPUT since READ mode was entered, simply switch to MAX or MIN. The value will appear on the LCD along with the word MAX or MIN. The MAX/MIN values are automatically updated and may be viewed at any time without interrupting the other...
Page 12
21 20 READ DC and AC VOLTAGE OUTPUTS mV, V, Vhi and Vac Choose this function to measure from 0.00 to 110.00 millivolts (mV), 0.00 to 10.25 DC Volts (V). Use the high voltage connection to read from 0.0 to 200.0 VDC (Vhi). See the next page for Volts AC. 1) Turn the Selector Knob to V 3) Press the DISPLAY/SOURCE/READ push-button until READ appears on the display 4) Press the TYPE/ENG UNITS push-button to select mV, V, Vhi or VAC on the display 5) Connect the red READ (+) lead and the black READ (-) lead...
Page 13
23 22 MEASURE THERMOCOUPLE SENSORS Choose this function to read a thermocouple. The input of the TechChek 830 is automatically cold junction compensated. 1) Disconnect the thermocouple from any instrument. 2) Turn the Selector Knob to T/C 3) Press the DISPLAY/SOURCE/READ push-button until READ and any T/C TYPE appear on the display 4) Use the proper thermocouple wire and corresponding miniature T/C connector to connect TechChek 830 to the thermocouple. To Change the Thermocouple type 1) Press the...
Page 14
25 24 CHECK CONTINUITY Choose this function to check continuity. A tone will sound and a sound symbol will appear on the display when the resistance between the leads is less than 100 Ohms. 1) Plug the leads into the TechChek 830 as shown below. 2) Turn the Selector Knob to Ω 3) Press the DISPLAY/SOURCE/READ push-button until READ or CONTINUITY appears on the display 4) Press the TYPE/ENG UNITS pushbuuton once to switch between READ and CONTINUITY on the display OPERATING INSTRUCTIONS REDBLACK READ...
Page 15
27 26 REDBLACK BLACK BLACK BLACK RED 3 2 1RTD Element OPERATING INSTRUCTIONS MEASURE RTD SENSORS MEASURE RTD SENSORS Choose this function to read an RTD. Three wires must be use for both 2 and three wire RTDs. 1) Disconnect the RTD from the instrument being calibrated. 2) Turn the Selector Knob to RTD 3) Press the DISPLAY/SOURCE/READ push-button until SOURCE and any RTD TYPE appear on the display 4) Connect using 3 wires as in the diagrams on the opposite page. Use a 3 wire connection to read a 4 wire...
Page 16
29 28 REDBLACK Frequenc y Output Signal Flowmeter Flow Sensor Variable Speed Drive Controller Transmitter REDBLACK Frequency Output Signal Flow Meter Line Frequency Variable Speed Drives Connection for signals with amplitudes below 10.25V RMSConnection for signals with amplitudes to 250 V RMS COUNT FREQUENCIES OPERATING INSTRUCTIONS COUNT FREQUENCIES Choose this function to use the 830 as a frequecny counter. Available ranges are from 0.01 to 10.00 kHz, 1 to 1000 Hz and from 1 to 1000 CPM...
Page 17
31 30 OPERATING INSTRUCTIONS CALIBRATE TRANSMITTERS ANY SOURCE FUNCTION AND READ mA, READ %, READ DP% Choose this function to supply the input signal to the transmitter and displaying the 4-20 mA output of the transmitter (used with 4- Wire Transmitters). ANY SOURCE FUNCTION AND P-XMTR mA, P-XMTR %, P-XMTR DP% Choose this function to simultaneously supply power to a 2-Wire transmitter while supply the input signal to the transmitter and display the 4-20 mA output of the transmitter. 1) Disconnect both...
Page 18
33 32 Pressure Source PROCESS CALIBRATORSOURCE READQUIK-CHEK® HI SET LOMIN READMAX TYPE ENG UNITS °C / °F ZERO DISPLAY SOURCE READREAD/POWER TRANSMITTERPOWER STORE RESET mA / % % DP FLOW ΩRTD T/CVmAFREQPRESSURE READ PRESSURE Choose this function to measure pressure with a ModPak 91 module holder and one of the many QuikCal 90 pressure modules. 1) Select the proper module for the pressure range of the process 2) Place the QuikCal 90 Module in the ModPak 91 holder 3) Plug the ModPak cable into the...
Page 19
35 34 PROCESS CALIBRATOR TYPE ENG UNITS °C / °F ZERO DISPLAY SOURCE READREAD/POWER TRANSMITTERPOWER STORE RESET mA / % % DP FLOWSOURCE READ Ω QUIK-CHEK® RTD T/CVmA HI SET LOMIN READMAX FREQPRESSURE OPERATING INSTRUCTIONS CALIBRATE PRESSURE TRANSMITTERS READ PRESSURE AND P-XMTR mA, P-XMTR %, P-XMTR DP% Choose this function to simultaneously supply power to a 2-Wire pressure transmitter while supplying the input signal to the trans- mitter and displaying the 4-20 mA output of the transmitter. 1)...
Page 20
37 36 SPECIFICATIONS POWER & MEASURE 2-WIRE TRANSMITTERS RANGES & ACCURACY:Same as for MILLIAMP SOURCE OUTPUT CURRENT: up to 24.00 mA TYPICAL DRIVE CAPABILITY:1200 Ohms @ 20.00 mA COMPLIANCE VOLTAGE:nominal 25 VDC @ 20 mA COMMON MODE ERROR:0.01% Full Scale/Common Mode Volt 2-WIRE TRANSMITTER SIMULATOR RANGES: 1.00 to 24.00 mA; -18.8 to 125.0% of 4 to 20 mA; % DP Flow ACCURACY:Same as for MILLIAMP SOURCE LOOP VOLTAGE LIMITS:Mininum, 3 VDC; Maximum 45 VDC OVERLOAD PROTECTION:Current limited to 25 mA...