    READ DC VOLTAGE OUTPUTS
    V,  m V
    Choose this function to measure from 0.00 to 10.25 DC Volts. For
    checking low level sensor outputs and other low levels change the
    input range to display from 0.00 to 110.00 millivolts. Use the high
    voltage connection to read from 0.0 to 200.0 VDC
    1) Connect the red READ (+) lead and the black READ (-) lead of
    the calibrator across the voltage to be measured.
    Signals above or below those available for the currently selected
    range will be indicated by OVER and UNDER on the display.
    OPERATING INSTRUCTIONS
    !!
    1 4 52 3
    5V MAX
    INPUT 45V MAX
    INPUT
    !
    300V
    MAX
    INPUT
    45V
    MAX
    INPUT!T/CVoltage Output Signal
    Controller
    Transmitter
    Power SupplyRED
    BLACK
    Connection for millivolts and Volts below 10.25 VDC
    Connection for Volts to 200.0 VDC
    18
    READ MILLIAMP OUTPUTS
    mA, mA % (Percent of 4 to 20 mA)
    Choose this function to measure from 0.00 to +24.00 milliamps or
    -25.0 to 125.0%. 
    1) Open the current loop at any convenient point along the signal
    path
    2) Connect the red READ (+) lead of the calibrator to the more
    positive point of the break and the black READ lead (-) to the
    more negative
    Display the present reading, Maximum or Minimum by moving the
    toggle switch from READ to MAX or MIN. If TechChek 820 is
    connected in the wrong polarity, the word POLARITY will appear
    in the display. Simply reverse the leads for correct indication.
    OPERATING INSTRUCTIONS
    !!
    1 4 52 3
    5V MAX
    INPUT 45V MAX
    INPUT
    !
    300V
    MAX
    INPUT
    45V
    MAX
    INPUT!T/CMilliamp Output Signal
    Controller
    Transmitter
    P/I
    DCSRED
    BLACK 
    						
    							21
    READ RESISTANCE
    Ohms
    Choose this function to measure resistance from 0.0 to 1000.0
    Ohms. 
    1) Connect the red READ (+) lead and the black READ (-) lead of
    the calibrator across the resistance to be measured.
    Signals above or below those available for the currently selected
    range will be indicated by OVER and UNDER on the display.
    OPERATING INSTRUCTIONS
    !!
    1 4 52 3
    5V MAX
    INPUT 45V MAX
    INPUT
    !
    300V
    MAX
    INPUT
    45V
    MAX
    INPUT!T/CRED
    BLACK
    Receiver InputOhms or 2-Wire RTD
    Controller
    Transmitter
    Computer
    20
    MEASURE  THERMOCOUPLE SENSORS
    Choose this function to read a thermocouple. The input of the
    TechChek 820 is automatically cold junction compensated.
    1) Disconnect the thermocouple from any instrument.
    2) Press the RANGE/TYPE push-button until the desired T/C type
    and temperature scale appear.
    3) Use the proper thermocouple wire and corresponding minia-
    ture thermocouple connector to connect TechChek 820 to the
    thermocouple.
    OPERATING INSTRUCTIONS
    !!
    T/C
    1 4
    52 3
    5V MAX
    INPUT 45V MAX
    INPUT
    !
    300V
    MAX
    INPUT
    45V
    MAX
    INPUT!+- 
    						
    							23
    MEASURE RTD SENSORS
    Choose this function to read an RTD. Three wires must be use for
    both 2 and three wire RTDs.
    1) Disconnect the RTD from any instrument.
    2) Press the RANGE/TYPE push-button until the desired RTD
    type and temperature scale appear.
    3) Connect using 3 wires as in the diagrams below. Spade lugs
    are recommended to minimize any contact resistance. 
    OPERATING INSTRUCTIONS
    !!
    1 4 52 3
    5V MAX
    INPUT 45V MAX
    INPUT
    !
    300V
    MAX
    INPUT
    45V
    MAX
    INPUT!T/C
    RED
    BLACKBLACK
    BLACK
    BLACK
    RED3
    2
    1RTD Element
    CHECK CONTINUITY
    Choose this function to check continuity. A tone will sound and a
    sound symbol will appear on the display when the resistance
    between the leads is less than 100 Ohms.
    1) Plug the leads into the TechChek 820 as shown below.
    2) Turn the selector knob to OHMs
    3) Press the SOURCE/READ pushbutton until the word READ
    appears on the display.
    4) Press the RANGE/TYPE pushbutton until the word
    CONTINUITY appears on the display
    22
    OPERATING INSTRUCTIONS
    !!
    1 4 52 3
    5V MAX
    INPUT 45V MAX
    INPUT
    !
    300V
    MAX
    INPUT
    45V
    MAX
    INPUT!T/CRED
    BLACK 
    						
    							25
    READ AC VOLTAGES
    Choose this function to measure from 0.0 to 250.0 V True RMS.
    CAUTION: Care should be used when measuring AC voltage.
    The included safety test probes or safety alligator clips  should be
    used. Do not exceed voltage limits shown on calibrator.
    AC Voltage Signal
    Mains
    Line Voltage
    !!
    1 4
    52 3
    5V MAX
    INPUT 45V MAX
    INPUT
    !
    300V
    MAX
    INPUT
    45V
    MAX
    INPUT!T/C
    BLACKRED
    OPERATING INSTRUCTIONS
    24
    COUNT FREQUENCIES
    Choose this function to use the 820 as a frequecny counter.
    Availible ranges are from 0.01 to 10.00 kHz, 1 to 1000 Hz and
    from 1 to 1000 CPM (Counts-Per-Minute).
    To measure waveforms with amplitudes between 1 V and 10.25 V
    RMS use the low level inputs. Use the high voltage connection to
    read waveforms with amplitudes from 10.25 to 250.0 V RMS
    OPERATING INSTRUCTIONS
    !!
    1 4 52 3
    5V MAX
    INPUT 45V MAX
    INPUT
    !
    300V
    MAX
    INPUT
    45V
    MAX
    INPUT!T/CFrequency Output Signal
    Flowmeter
    Flow Sensor
    Variable Speed Drive
    Controller
    TransmitterRED
    BLACK
    Frequency Output Signal
    Flow Meter
    Line Frequency
    Variable Speed Drives
    !!
    1 4
    52 3
    5V MAX
    INPUT 45V MAX
    INPUT
    !
    300V
    MAX
    INPUT
    45V
    MAX
    INPUT!T/C
    BLACKRED
    Connection for signals with amplitudes below 10.25 V RMS
    Connection for signals with amplitudes to 250 V RMS 
    						
    							POWER & MEASURE 2-WIRE TRANSMITTERS
    RANGES & ACCURACY:Same as for MILLIAMP SOURCE
    OUTPUT CURRENT: up to 24.00 mA
    TYPICAL DRIVE CAPABILITY:1200 Ohms @ 20.00 mA
    COMPLIANCE VOLTAGE:nominal 25 VDC @ 20 mA
    2-WIRE TRANSMITTER SIMULATOR
    RANGES:
    1.00 to 24.00 mA; -18.8 to 125.0% of 4 to 20 mA
    ACCURACY:Same as for MILLIAMP SOURCE
    LOOP VOLTAGE LIMITS:Mininum, 3 VDC; Maximum 45 VDC
    OVERLOAD PROTECTION:Current limited to 25 mA nominal
    MILLIAMP READ
    RANGES:
    0.00 to 24.00 mA; -25.0 to 125.0 % of 4 to 20 mA
    ACCURACY:Same as for MILLIAMP SOURCE
    OVERLOAD PROTECTION: Current limited to 25 mA nominal
    VOLTAGE BURDEN: 0.9V at 4 mA, 1.2V at 20 mA, 1.9V at 24 mA
    DC VOLTAGE SOURCE
    RANGES:
    0.00 to 110.00 mV; 0.00 to 10.25V
    ACCURACY:
    ±(0.05% of 110 mV+ 0.01mV) = ±0.07 mV
    ±(0.05% of 10.25 V + 0.01V) =  ±0.02V
    SOURCE CURRENT:>20 mA
    SINK CURRENT:>20 mA
    OUTPUT IMPEDANCE: 
    						
    							ACCURACY RTD: ±1°C (At 1 mA Excitation Current)
    RTD Type: Pt 100 Ohm DIN/IEC 751, Alpha = 1.3850 (0.00385)
    RANGE RTD: -100 to 850°C; -148 to 1562°F
    TEMPERATURE EFFECT:±((0.035 mV/°C)*(1/mA Excitation Current))
    ALLOWABLE EXCITATION CURRENT: 0.125 to 2.0 mA continu-
    ous DC
    READ RTD & OHMS
    RANGE OHMS: 0.0 to 1000.0 Ohms
    ACCURACY: ±(0.05% of 1000.0 Ohms + 0.1 Ohm) = ±0.6 Ohms
    RTD RANGE & ACCURACY: Same as for SOURCE RTDs
    EXCITATION CURRENT SUPPLIED:1 mA, nominal
    FREQUENCY SOURCE
    RANGES: 1 to 1000 CPM (Count-Per-Minute); 1 to 1000 Hz,
    0.01 to 10.00 kHz
    ACCURACY:
    ±(0.05% of 1000 CPM + 1 CPM) = ±2 CPM
    ±(0.05% of 1000 Hz + 1 Hz) = ±2 Hz
    ±(0.05% of 10.00 kHz + 0.01 kHz) = ±0.02 kHz
    OUTPUT WAVEFORM:Square Wave, Zero Crossing, -1V to +5V
    ±10%
    RISETIME:Hz 1 Meg Ohm to 10.25V, >5 Meg Ohm to
    200V
    SOURCE RESISTANCE EFFECT: 0.01% per 100 Ohms
    SOURCE THERMOCOUPLES
    RANGES:
    Type J  -100 to 1200°C; -148 to 2192°F
    Type K  -100 to 1371°C; -148 to 2500°F
    Type T -100 to 400°C; -148 to 752°F
    Type E  -100 to 1000°C; -148 to 1832°F
    ACCURACY:
    1°C for temperatures above 0°C
    2°C for temperatures below 0°C
    COLD JUNCTION ACCURACY: ±1°C
    COLD JUNCTION EFFECT: within 0.05°C per °C change
    OVERLOAD PROTECTION:Fused
    OUTPUT IMPEDANCE:20 mA
    READ THERMOCOUPLES
    RANGES & ACCURACY: Same as for SOURCE THERMOCOUPLES 
    INPUT IMPEDANCE:> 1 Meg Ohm
    OPEN THERMOCOUPLE DETECTION: 450 millisecond pulse.
    Nominal threshold, 10 K Ohms.
    SOURCE RTD & OHMS
    RANGE OHMS: 0.0 to 400.0 Ohms
    ACCURACY:±0.05% of Full Scale + 0.075 mV/mA Excitation Current
    ACCURACY OHMS: ±(0.05% of 400.0 Ohms + 0.1 Ohm = ±0.3
    Ohms (At 1 mA Excitation Current)
    28
    SPECIFICATIONS 
    						
    							3130
    NOTESWARRANTY
    Altek products are warranted to be free from defects in material
    and workmanship (excluding fuses, batteries and leads) for a
    period of three years from the date of shipment. Warranty repairs
    can be obtained by returning the equipment prepaid to our factory.
    Products will be replaced, repaired, or adjusted at our option. Altek
    gives no other warranties, including any implied warranty of fitness
    for a particular purpose.Also, Altek shall not be liable for any
    special, indirect, incidental or consequential damages or losses
    arising from the sale or use of its products. 
    						
    							32
    ORDERING INFORMATION Part No.
    MODEL 820 TechChek 820
    PROCESS CALIBRATOR
    Included with each Model 820 are:
    Carrying Case with belt loop and shoulder strap 
    Test Lead Kit 
    NOTES 
    						
    							PN 1839053 Rev A April 2002
    ©2002 Altek Industries, Inc. Specifications subject to change without notice. 