Altek Process Calibrator TechChek 820 Manual
Have a look at the manual Altek Process Calibrator TechChek 820 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 Altek manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
19 READ DC VOLTAGE OUTPUTS V, m V Choose this function to measure from 0.00 to 10.25 DC Volts. For checking low level sensor outputs and other low levels change the input range to display from 0.00 to 110.00 millivolts. Use the high voltage connection to read from 0.0 to 200.0 VDC 1) Connect the red READ (+) lead and the black READ (-) lead of the calibrator across the voltage to be measured. Signals above or below those available for the currently selected range will be indicated by OVER and UNDER on the display. OPERATING INSTRUCTIONS !! 1 4 52 3 5V MAX INPUT 45V MAX INPUT ! 300V MAX INPUT 45V MAX INPUT!T/CVoltage Output Signal Controller Transmitter Power SupplyRED BLACK Connection for millivolts and Volts below 10.25 VDC Connection for Volts to 200.0 VDC 18 READ MILLIAMP OUTPUTS mA, mA % (Percent of 4 to 20 mA) Choose this function to measure from 0.00 to +24.00 milliamps or -25.0 to 125.0%. 1) Open the current loop at any convenient point along the signal path 2) Connect the red READ (+) lead of the calibrator to the more positive point of the break and the black READ lead (-) to the more negative Display the present reading, Maximum or Minimum by moving the toggle switch from READ to MAX or MIN. If TechChek 820 is connected in the wrong polarity, the word POLARITY will appear in the display. Simply reverse the leads for correct indication. OPERATING INSTRUCTIONS !! 1 4 52 3 5V MAX INPUT 45V MAX INPUT ! 300V MAX INPUT 45V MAX INPUT!T/CMilliamp Output Signal Controller Transmitter P/I DCSRED BLACK
21 READ RESISTANCE Ohms Choose this function to measure resistance from 0.0 to 1000.0 Ohms. 1) Connect the red READ (+) lead and the black READ (-) lead of the calibrator across the resistance to be measured. Signals above or below those available for the currently selected range will be indicated by OVER and UNDER on the display. OPERATING INSTRUCTIONS !! 1 4 52 3 5V MAX INPUT 45V MAX INPUT ! 300V MAX INPUT 45V MAX INPUT!T/CRED BLACK Receiver InputOhms or 2-Wire RTD Controller Transmitter Computer 20 MEASURE THERMOCOUPLE SENSORS Choose this function to read a thermocouple. The input of the TechChek 820 is automatically cold junction compensated. 1) Disconnect the thermocouple from any instrument. 2) Press the RANGE/TYPE push-button until the desired T/C type and temperature scale appear. 3) Use the proper thermocouple wire and corresponding minia- ture thermocouple connector to connect TechChek 820 to the thermocouple. OPERATING INSTRUCTIONS !! T/C 1 4 52 3 5V MAX INPUT 45V MAX INPUT ! 300V MAX INPUT 45V MAX INPUT!+-
23 MEASURE RTD SENSORS Choose this function to read an RTD. Three wires must be use for both 2 and three wire RTDs. 1) Disconnect the RTD from any instrument. 2) Press the RANGE/TYPE push-button until the desired RTD type and temperature scale appear. 3) Connect using 3 wires as in the diagrams below. Spade lugs are recommended to minimize any contact resistance. OPERATING INSTRUCTIONS !! 1 4 52 3 5V MAX INPUT 45V MAX INPUT ! 300V MAX INPUT 45V MAX INPUT!T/C RED BLACKBLACK BLACK BLACK RED3 2 1RTD Element CHECK CONTINUITY Choose this function to check continuity. A tone will sound and a sound symbol will appear on the display when the resistance between the leads is less than 100 Ohms. 1) Plug the leads into the TechChek 820 as shown below. 2) Turn the selector knob to OHMs 3) Press the SOURCE/READ pushbutton until the word READ appears on the display. 4) Press the RANGE/TYPE pushbutton until the word CONTINUITY appears on the display 22 OPERATING INSTRUCTIONS !! 1 4 52 3 5V MAX INPUT 45V MAX INPUT ! 300V MAX INPUT 45V MAX INPUT!T/CRED BLACK
25 READ AC VOLTAGES Choose this function to measure from 0.0 to 250.0 V True RMS. CAUTION: Care should be used when measuring AC voltage. The included safety test probes or safety alligator clips should be used. Do not exceed voltage limits shown on calibrator. AC Voltage Signal Mains Line Voltage !! 1 4 52 3 5V MAX INPUT 45V MAX INPUT ! 300V MAX INPUT 45V MAX INPUT!T/C BLACKRED OPERATING INSTRUCTIONS 24 COUNT FREQUENCIES Choose this function to use the 820 as a frequecny counter. Availible ranges are from 0.01 to 10.00 kHz, 1 to 1000 Hz and from 1 to 1000 CPM (Counts-Per-Minute). To measure waveforms with amplitudes between 1 V and 10.25 V RMS use the low level inputs. Use the high voltage connection to read waveforms with amplitudes from 10.25 to 250.0 V RMS OPERATING INSTRUCTIONS !! 1 4 52 3 5V MAX INPUT 45V MAX INPUT ! 300V MAX INPUT 45V MAX INPUT!T/CFrequency Output Signal Flowmeter Flow Sensor Variable Speed Drive Controller TransmitterRED BLACK Frequency Output Signal Flow Meter Line Frequency Variable Speed Drives !! 1 4 52 3 5V MAX INPUT 45V MAX INPUT ! 300V MAX INPUT 45V MAX INPUT!T/C BLACKRED Connection for signals with amplitudes below 10.25 V RMS Connection for signals with amplitudes to 250 V RMS
POWER & MEASURE 2-WIRE TRANSMITTERS RANGES & ACCURACY:Same as for MILLIAMP SOURCE OUTPUT CURRENT: up to 24.00 mA TYPICAL DRIVE CAPABILITY:1200 Ohms @ 20.00 mA COMPLIANCE VOLTAGE:nominal 25 VDC @ 20 mA 2-WIRE TRANSMITTER SIMULATOR RANGES: 1.00 to 24.00 mA; -18.8 to 125.0% of 4 to 20 mA ACCURACY:Same as for MILLIAMP SOURCE LOOP VOLTAGE LIMITS:Mininum, 3 VDC; Maximum 45 VDC OVERLOAD PROTECTION:Current limited to 25 mA nominal MILLIAMP READ RANGES: 0.00 to 24.00 mA; -25.0 to 125.0 % of 4 to 20 mA ACCURACY:Same as for MILLIAMP SOURCE OVERLOAD PROTECTION: Current limited to 25 mA nominal VOLTAGE BURDEN: 0.9V at 4 mA, 1.2V at 20 mA, 1.9V at 24 mA DC VOLTAGE SOURCE RANGES: 0.00 to 110.00 mV; 0.00 to 10.25V ACCURACY: ±(0.05% of 110 mV+ 0.01mV) = ±0.07 mV ±(0.05% of 10.25 V + 0.01V) = ±0.02V SOURCE CURRENT:>20 mA SINK CURRENT:>20 mA OUTPUT IMPEDANCE:
ACCURACY RTD: ±1°C (At 1 mA Excitation Current) RTD Type: Pt 100 Ohm DIN/IEC 751, Alpha = 1.3850 (0.00385) RANGE RTD: -100 to 850°C; -148 to 1562°F TEMPERATURE EFFECT:±((0.035 mV/°C)*(1/mA Excitation Current)) ALLOWABLE EXCITATION CURRENT: 0.125 to 2.0 mA continu- ous DC READ RTD & OHMS RANGE OHMS: 0.0 to 1000.0 Ohms ACCURACY: ±(0.05% of 1000.0 Ohms + 0.1 Ohm) = ±0.6 Ohms RTD RANGE & ACCURACY: Same as for SOURCE RTDs EXCITATION CURRENT SUPPLIED:1 mA, nominal FREQUENCY SOURCE RANGES: 1 to 1000 CPM (Count-Per-Minute); 1 to 1000 Hz, 0.01 to 10.00 kHz ACCURACY: ±(0.05% of 1000 CPM + 1 CPM) = ±2 CPM ±(0.05% of 1000 Hz + 1 Hz) = ±2 Hz ±(0.05% of 10.00 kHz + 0.01 kHz) = ±0.02 kHz OUTPUT WAVEFORM:Square Wave, Zero Crossing, -1V to +5V ±10% RISETIME:Hz 1 Meg Ohm to 10.25V, >5 Meg Ohm to 200V SOURCE RESISTANCE EFFECT: 0.01% per 100 Ohms SOURCE THERMOCOUPLES RANGES: Type J -100 to 1200°C; -148 to 2192°F Type K -100 to 1371°C; -148 to 2500°F Type T -100 to 400°C; -148 to 752°F Type E -100 to 1000°C; -148 to 1832°F ACCURACY: 1°C for temperatures above 0°C 2°C for temperatures below 0°C COLD JUNCTION ACCURACY: ±1°C COLD JUNCTION EFFECT: within 0.05°C per °C change OVERLOAD PROTECTION:Fused OUTPUT IMPEDANCE:20 mA READ THERMOCOUPLES RANGES & ACCURACY: Same as for SOURCE THERMOCOUPLES INPUT IMPEDANCE:> 1 Meg Ohm OPEN THERMOCOUPLE DETECTION: 450 millisecond pulse. Nominal threshold, 10 K Ohms. SOURCE RTD & OHMS RANGE OHMS: 0.0 to 400.0 Ohms ACCURACY:±0.05% of Full Scale + 0.075 mV/mA Excitation Current ACCURACY OHMS: ±(0.05% of 400.0 Ohms + 0.1 Ohm = ±0.3 Ohms (At 1 mA Excitation Current) 28 SPECIFICATIONS
3130 NOTESWARRANTY Altek products are warranted to be free from defects in material and workmanship (excluding fuses, batteries and leads) for a period of three years from the date of shipment. Warranty repairs can be obtained by returning the equipment prepaid to our factory. Products will be replaced, repaired, or adjusted at our option. Altek gives no other warranties, including any implied warranty of fitness for a particular purpose.Also, Altek shall not be liable for any special, indirect, incidental or consequential damages or losses arising from the sale or use of its products.
32 ORDERING INFORMATION Part No. MODEL 820 TechChek 820 PROCESS CALIBRATOR Included with each Model 820 are: Carrying Case with belt loop and shoulder strap Test Lead Kit NOTES
PN 1839053 Rev A April 2002 ©2002 Altek Industries, Inc. Specifications subject to change without notice. All rights reserved. Printed in U.S.A.