    							PneuCal III
    The 2500T offers
    simplistic set-up to
    make your job
    easier.
    The Altek PneuCal III is a highly accurate, digital calibrator for pneumatic field
    instrumentation, including valve actuators, P/I transmitters, controllers, gauges,
    switches and recorders. It is especially suitable for checking 3 to 15 PSI 
    systems. Its dual precision regulators enable output of set and variable 
    pressures to control devices, while the switching manifold allows fast selection
    among the pressure ports. The unit simultaneously displays pressure and mA,
    and has a built-in loop power supply. Accuracy is ± 0.035% of full scale for
    all ranges. Optional external sensors measure other pressure ranges.
    Each Unit Includes: Built - In Switch Testing, AC Adapter/Charger, Test Leads,
    Hard Shell Carrying Case, Instructions, Built -in Pressure and Vacuum Pumps.
    663 PNEUMATIC CALIBRATOR SPECIFICATIONS
    Features:
    nTwo Built-In Pumps Ð
    Vacuum and Pressure
    n High Accuracy Ð 0.035% FS
    n1-51 mA Inputs
    n Switch Test Inputs
    n 100 PSID Internal Sensor
    nRead up to 10,000 PSI With
    Optional External Sensor 
    nBuilt-In 24 VDC Loop Power
    nTemperature Compensated
    From -10û to 50ûC 
    (13û to 122ûF)
    nSelectable Engineering Units
    nJust 18 Pounds
    nDual Digital Display Ð
    PSI and mA
    nGauge, Differential, or 
    Vacuum Modes
    Unless otherwise indicated, all specifications are referred to an ambient temperature of 
    23ûC ±1ûC (72ûF ±2ûF).
    Pressure Ranges/Resolution: -5.00 to 100.00 PSI, -34.47 to 680.00 kPa, VAC to 
    10,000 PSI with optional modules
    Pressure Accuracy: ±0.035% FS PSI (over -10 to +50ûC range)
    Pressure Engineering Units: PSI; Pa; kPa; MPa; BAR; mBAR; atm; kgf/cm
    2; mm Hg; 
    in Hg @ 0ûC and 15ûC, in H
    2O: 4ûC,15ûC, 20ûC, 23ûC; mm 
    H20 @ 4ûC and 23ûC; cm H2O @ 4ûC and 23ûC
    Electrical Engineering Units: mA; % of 4-20 mA; % of 10-50 mA
    Electrical Input Connections: Double insulated, full-size banana jacks
    mA Input Range: 1-51 mA
    mA Resolution: 0.001mA
    mA Accuracy: ±0.01% FS
    Temperature Effect: 0.005%/ûC (mA Input only)
    Loop Power: 24 VDC, ±20%, 25 mA max. (When loop power is enabled, 
    the discharge rate on the batteries exceeds the charging 
    capacity of the 16 VDC adapter.)
    Switch Test: Dry contacts only. (Display will freeze when switch test inputs 
    change state. Press RESET key to resume normal operation.)
    Operating Temperature: -10ûC to 50ûC (13ûF to 122ûF)
    Storage Temperature: -30ûC to 85ûC (-22ûF to 185ûF)
    Zero Adjustment: Via front panel push-button (pressure only)
    Display: 2-line,16 character alpha-numeric display
    Power: Internal Ni-Cd batteries or 16 VDC adapter. (To charge battery 
    the unit has to be powered off.)
    Battery Life (typical): 23 hrs. (1 hr. with 24 VDC transmitter power at 20 mA)
    Dimensions (h x w x d): 7x18.5x15.5     
    PHONE: 1-800-322-5835 ¥ FAX: 716-349-3510E-MAIL: [email protected] ¥ WEB: www.altekcalibrators.com 
    						
    							PRESSURE MODULE SPECIFICATIONS
    MODPAK 91 MODULE HOLDER
    MODULES
    Gauge Pressure Modules
    Model Range Accuracy Media
    90-5G 5 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible
    90-10G 10 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible
    90-30G 30 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible
    90-100G 100 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible
    90-300G 300 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible
    90-500G 500 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible
    90-1,000G 1,000 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible
    90-2,500G 2,500 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible
    90-5KG 5,000 PSIG ±0.1% FS 316SS Compatible
    90-10KG 10,000 PSIG ±0.1% FS 316SS Compatible
    ÒDual SensorÓ Pressure Module
    Model Range Accuracy Media
    90-10G/2500G 10/2,500 PSIG ±0.025% Rdg +0.01%FS 316SS Compatible
    Differential Pressure Modules
    Model Range Accuracy Media
    90-10WD ±10 H
    2O Diff. ±0.025 in. H2O Clean Dry Gasses
    90-5D 5 PSID ±0.025% Rdg +0.01% FS 316SS High/Dry Gas Low
    90-15D 15 PSID ±0.025% Rdg +0.01% FS 316SS High/Dry Gas Low
    90-30D 30 PSID ±0.025% Rdg +0.01% FS 316SS High/Dry Gas Low
    90-100D 100 PSID ±0.025% Rdg +0.01% FS 316SS High/Dry Gas Low
    Absolute Pressure Modules
    Model Range Accuracy Media
    90-15A 15 PSIA ± 0.05% FS 316SS Compatible
    90-30A 30 PSIA ± 0.05% FS 316SS Compatible
    Vacuum Modules
    Model Range Accuracy Media
    90-15V VAC to 0 PSIG ± 0.05% FS 316SS Compatible
    90-15C VAC to 15 PSIG ± 0.05% FS 316SS Compatible
    PHONE: 1-800-322-5835 ¥ FAX: 716-349-3510E-MAIL: [email protected] ¥ WEB: www.altekcalibrators.com 
    						