Home > Altek > Calibrator > Altek Model 40A Frequency Source Manual

Altek Model 40A Frequency Source Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Altek Model 40A Frequency Source Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 Altek manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Frequency Source
    Model 40A
    • Frequency output40 precise steps 0.5 Hz to 60 KHz
    • Tuning fork crystal0.0008% accuracy
    • Laboratory stability±0.001% per deg C
    • Zero based & zero crossing square waveSwitch selected outputs
    • Pocket sized
    Altek Model 40A Frequency Source provides 40
    precise frequencies from 0.5 to 60,000 Hertz. Output
    is a square wave* of 15 volt peak-to-peak amplitude
    which can be set to any lower value with a built-in
    attenuator. A switch selects zero based or zero center
    outputs for standard or zero crossing inputs.
    The Model 40A checks and calibrates the electronics
    of turbine meters, vortex shedding meters, positive
    displacement meters, oscilloscopes, electronic
    counters, tachometers and many other frequency
    input devices. LED flashes to calibrate optical and
    fiber optic devices.
    Frequency accuracy is ±8 parts per million. Long term 
    stability is ±3 parts per million per year.
    Instant turn-on allows long life from the two 9 volt
    alkaline batteries (200 hours continuous). Frequency
    is adjusted with the 8 position frequency selector
    switch and a 5 position range multiplier switch.
    The low cost Model 40A is a complete, compact
    source for checkout and calibration of all frequency
    instruments in the field, shop, lab and control room.
    General description
    *20 KHz = 1:3 ratio     
    12 KHz = 2:5 ratio 
    						
    							PN 1887794 Rev A April 2002
    ©2002 Altek Industries, Inc. Specifications subject to change without notice. 
    All rights reserved. Printed in U.S.A.  
    Altek products are warranted to be free from defects
    in material and workmanship (excluding fuses,
    batteries and leads) for a period of three years from
    the date of shipment. Warranty repairs can be
    obtained by returning the equipment prepaid to our
    factory. Products will be replaced, repaired, or
    adjusted at our option. Altek gives no other
    warranties, including any implied warranty of fitness
    for a particular purpose.Also, Altek shall not be liable
    for any special, indirect, incidental or consequential
    damages or losses arising from the sale or use of its
    products.
    Warranty
    (Unless otherwise indicated, specifications are at 23°C)
    Accuracy: ±0.0008% (8 parts per million)
    Frequencies: 5, 6, 10, 12, 15, 20, 30 and 60 Hz times multiplier
    Multiplier: 0.1, 1, 10, 100 and 1000 times frequency scale
    Output: Nominal 15V peak-to-peak square wave (1:3 ratio 20KHz,
    2:5 ratio 12KHz)
    Waveform: Switch selected zero based or zero crossing squarewave
    Attenuator: Sliding potentiometer, logarithmic taper
    Output impedance: 2500 ohms maximum
    Output drive current: 6mA at max output voltage
    Rise time: 1 microsecond nominal into resistive load
    Operating temperature range: 14 to 158° F 
    (-10 to +70°C)
    Storage temperature range: -22 to +185° F (-30 to +85°C)
    Temperature effect: ±0.001%/°C
    Batteries: 2 x 9-volt Alkaline
    Battery life: Nominal 200 hours of operation. Batteries should be
    removed when storing the unit for >3 months.
    Battery indication: LED flashes with frequency if batteries OK
    Size: 2-1/8 x 4 x 2-1/4 inches (54x102x55mm)
    Weight: 7 oz. (0.2 kg)
    Specifications
    Multiplier
    Switch Output frequency
    0.1 0.5 0.6 1 1.2 1.5 2 3 6
    1 5 6 101215203060
    10 50 60 100 120 150 200 300 600
    100 500 600 1K 1.2K 1.5K 2K 3K 6K
    1000 5K 6K 10K 12K 15K 20K 30K 60K
    5 6 10 12 15 20 30 60
    Hx (x Multiplier) Switch
    Connect the Model 40A to the frequency input of the
    device to be checked or calibrated. Select the
    frequency and multiplier for the initial frequency
    desired. Turn the power switch “on” and observe the
    glow in the light emitting diode (LED).
    If the diode glows, the battery is at a sufficient level to 
    provide proper operation. At low frequencies, the LED
    flashes in synchronism with the selected frequency.
    Adjust the attenuator for proper level. Check at all
    desired frequencies, turn off and disconnect the
    Model 40A.
    Operating instructions
    RECEIVER
    FREQUENCY
    INPUT ALTEK
    40A 
    						
    All Altek manuals Comments (0)