    							Process Voltage Analyzer
    Model 235
    • Calibrate voltage instrumentsSimulate Inputs
    Check Outputs
    • Three ranges with 0.05% accuracy0.0 to 199.9 Millivolts
    0.00 to 19.99 Volts
    -25.0 to 199.9% of 1 to 5 Volts
    • Calibrate live voltage input receiversNo need to disconnect loop
    • “Quik-Chek®” switchSet HI & LO outputs to match your Span & Zero
    Instantly switch between HI & LO outputs
    • Automatic overload protectionWithstands 120V AC or DC without fusesCalibrate and check out all your DC voltage
    instrumentation with Altek’s Model 235 Process
    Voltage Analyzer. Source from millivolts to 20 VDC
    and measure your process voltage signals with this
    single pocket sized instrument
    Display your readings with 0.1% resolution from 0.0 to
    199.9 millivolts, 0.00 to 19.99 Volts or from -25.0 to
    199.9% of the 1 to 5 Volt signal. Accuracy is ±0.05%
    of span.
    Source voltage
    Built-in batteries calibrate high or low impedance
    voltage or millivolt instruments. Accuracy is
    maintained to devices requiring up to 30mA. Three
    nine volt alkaline batteries provide 100 hours of use
    into high impedance loads. An optional AC adapter
    plugs in for continuous bench use.
    Instantly recall three output voltages
    User adjustable Quik-Chek switch provides instant HI
    and LO settings in source mode. DIAL position selects
    a continuously adjustable potentiometer. Full 10 turns
    with high resolution allows fast, easy setting to any
    exact value.
    Calibrate live circuits
    Automatically calibrate live circuits without
    disconnecting wires. Circuit sinks up to 20mA to
    clamp test voltage in all ranges. 1 to 5 Volt devices in
    live 4 to 20mA loops can be calibrated without any
    effect on the other instruments in the loop.
    Measure voltages
    Precisely measure positive and negative voltages in
    the 100mV and 10V ranges. 0 to 100% is displayed in
    the 1 to 5 Volt range for checkout of process control
    instruments. High input resistance (>2 Megohms)
    minimizes loading effect on signals. Special protective
    circuitry withstands misconnection to 120V AC in any
    mode without fuses.
    Display digits are 0.35 (9mm) high for readability
    from across the room. Non-glare liquid crystal display
    is readable in any light…even in direct sunlight. The
    digital measuring circuit is independent and measures
    the actual input or output.
    The Altek Model 235 is rugged, yet lightweight and
    pocket sized. Backed by a three year warranty, the
    Model 235 is toolbox tough. Latest LSI circuitry and
    wide temperature range components make the Model
    235 ideal for use in the field, control room and shop.
    Benchtop accuracy in a toolbox calibrator assures
    fast, precise setting of trips, recorders, controllers,
    loggers, computers and analysis instruments. Altek
    brings you the handy Model 235 Process Voltage
    Analyzer at a cost low enough for every check-out
    and maintenance person.
    General description 
    						
    							PN 1886496 Rev A April 2002
    ©2002 Altek Industries, Inc. Specifications subject to change without notice. 
    All rights reserved. Printed in U.S.A.  
    Altek products are warranted to be free from defects
    in material and workmanship (excluding fuses,
    batteries and leads) for a period of three years from
    the date of shipment. Warranty repairs can be
    obtained by returning the equipment prepaid to our
    factory. Products will be replaced, repaired, or
    adjusted at our option. Altek gives no other
    warranties, including any implied warranty of fitness
    for a particular purpose.Also, Altek shall not be liable
    for any special, indirect, incidental or consequential
    damages or losses arising from the sale or use of its
    products.
    Warranty
    Unless otherwise Indicated, specifications are in % of Span @ 23° C
    Accuracy: ±(0.05% of Span + 1 Least Significant Digit)
    Display: Liquid Crystal; 31/2 digit, 0.35 (9.0mm) high
    Negative voltages: Measured on 100mV and 10V ranges
    “Quik-Chek” Factory preset at 0% and 100% (1 and 5V)
    ADJUSTMENT RANGE LO HI
    20 Volts 0.10V to +1.50V       0.75V to 11.00V
    Percent 1 to 5 V -25.0% to +12.0%    -10.0% to 199.9%
    200 Millivolts -2.0mV to +25.0mV  10mV to 199.9 mV
    Batteries: 3 x 9 Volt alkaline are included
    Battery life: Nominal 100 Hours, Sourcing into high impedance 
    loads; nominal 20 Hours at 20mA Drain. Batteries should be
    removed when storing the unit >3 months.
    Source current: 30mA Maximum
    Sink current: 20mA Maximum
    Output impedance: 2 Megohms
    Specifications
    Source resistance effect (Read mode): 0.1% error per 2000 Ohms
    Overload protection: Protected to 120 Volts AC or DC
    Overload indicator: Lamp indicates high current or misconnection
    Short circuit duration: Continuous
    Low voltage indicator: LO BAT arrow       turns on at 18 Volts
    (approximately 10 operating hours remain)
    Temperature effect: ±0.01%/°C (Based on 23°C ±23°C Recom-
    mended Range)
    Recommended temperature range: 32 to 122°F (0 to 50°C)
    Operating temperature range: -5 to +140  °F (- 20 to + 60 °C)
    Storage temperature range: -22 to +175°F (-30 to +80°C)
    Relative humidity: 10 to 90%, non-condensing
    Warm up time: 3 seconds to rated accuracy
    Overall size: 21
    /
    2x 25
    /
    8x 51
    /
    8inches (63.5 x 66.7 x 130 mm)
    Weight: 12.5 oz (0.35 kg)
    AC Adaptors: optional; 120 or 240V, 50/60 Hz
    Carrying case: optional, zippered with belt loop
    Operating instructions
    Calibrate voltage inputsDisconnect one or both input wires from the device to be check or
    calibrated. Attach the red (+) lead of the Model 235 to the plus input
    of the device to be calibrated, connect the black (-) lead to the
    minus terminal. Select the desired range and turn the
    SOURCE/OFF/READ mode selector switch to the SOURCE 
    position. Actual voltage sent to the receiving device is shown on the
    LCD.
    Output voltage is continuously adjustable with the QUIK-CHEK
    switch in the DIAL position. The source voltage can be set to any
    exact value from 0 to 200mV, 0 to 20V or -25 to 125% of 1 to 5V. 
    HI and LO values are user adjustable and can be instantly selected
    by the QUIK-CHEK switch.
    Measure voltage outputsPlace the leads of the MODEL 235 across the voltage signal to be
    measured. Place the MODE Switch in the READ position and 
    select the range to be measured. Choose the 100mV position for
    signals from -199.9mV to +199.9mV. Signals from -19.99V to
    +19.99V can be measured with the 10V range. 1 to 5V signals can
    be displayed in units of 0 to 100% to monitor live process signals.
    When measuring voltages from high impedance sources, the
    source resistance effect is 0.1% per 2000 Ohms. All ranges are ful-
    ly protected to 120V AC against misconnection.
    CONTROLLER
    (TYPICAL)
    TRANSMITTER
    4-20mA
    ADDITIONAL
    4-20 mA OR 1-5V
    INSTRUMENTSALTEK
    235
    SOURCE
    MODE RECORDER
    COMPUTER
    LOGGER
    ETC,
    1-5V DC250Ω
    ALTEK
    235
    READ
    MODEANALYZER
    VOLTAGE
    SOURCE
    mV SOURCE
    ALTEK
    235
    SOURCE
    MODECOMPUTER
    RECEIVER
    LOGGER
    TRANSMITTER
    Check 1 to 5V inputs in live loopAny 1 to 5V device in a 4 to 20mA loop can be calibrated without
    breaking the loop or turning off the signal current. Clip the red lead
    to the positive input terminal and the black lead to the negative of
    the device to be checked or calibrated. It is not necessary to 
    disconnect any associated 250 Ohm resistor. Make certain that
    changing the signal input will not disturb the process or cause 
    unexpected alarms when checking on-line instruments.
    It is important to remember the Model 235 drives only the device to
    which it is connected. It has no effect on other devices in the 4 to
    20mA loop.
    Set the RANGE switch to the 1 to 5 Volt range and the MODE
    switch to SOURCE. the LCD will display 0 to 100% corresponding
    to the 1 to 5 Volts required by the device being calibrated. The
    voltage is set to any exact value by the 10 turn dial. Adjust HI and
    LO trimmers to desired values for fastest repetitive calibration
    through the QUIK-CHEK switch.
    Additionally, SINK MODE will clamp the selected value in the
    100mV and 10V Ranges to the maximum sink current of 20mA. For 
    example, receivers requiring 0.25 to 1.25V may be calibrated in the
    10 Volt range. Connect the Model 235 across the 62.5 Ohm 
    resistor and set the LO and HI QUIK-CHEKS for 0.25 and 1.25V. 
    						