    							Thermcouple Source
    Model 22
    • Direct temperature output22 precise steps
    • 0.1% accuracy25, 50, or 100 degree resolution
    • Rugged designField, shop and control room use
    • Pocket sized1 year battery life
    • Models for types B, E, J, K, R, S, T°C & °F cold-junction compensatedAltek Model 22 Thermocouple Sources provide 22
    precise temperatures for inputs to transmitters,
    recorders, controllers, alarms, data acquisition and
    computer systems. The Model 22 provides
    thermocouple wire output and is cold-junction
    compensated for ambient temperature variations.
    Conformity to the particular thermocouple vs millivolt
    curve is in accordance with the latest ASTM and IPTS
    standards for exact temperature simulation. 
    Thermocouple types B, E, J, K, R, S and T are
    available in both degrees F and C. Resolution is 25,
    50 or 100 degrees, corresponding to full scale output
    of 500, 1000, 1700, 2100 and 3100 degrees
    respectively.
    Dual ranges, with an individual “ON” position for each
    range, allow quick, easy settings for any output.
    Calibrated accuracy is ±0.1% of span ±1 degree.
    Negative temperatures add ±2 degrees.
    Two built-in AA cells provide power for one year of
    everyday use. A front panel LED pulses every time the
    Model 22 is turned on to indicate proper battery
    voltage.
    The low cost Altek Model 22 is a complete, compact
    source for checkout and calibration of all
    thermocouple instruments in the field, shop or control
    room.
    General description
    Select the Altek Model 22 for the correct thermocouple type, the
    desired temperature scale (F or C), and range. Connect the Model
    22 to the input terminals of the instrument to be tested, the field
    mounted head or junction box terminals.
    Cold-junction compensation and instantaneous automatic
    standardization is built into the Model 22.
    Set the temperature selection switch to the desired temperature
    and slide the “ON” switch to the selected scale. The Battery Check
    LED will pulse once indicating proper battery voltage. If no pulse is
    seen, the batteries should be replaced with 2 “AA” cells. Alkaline
    cells are preferred for longest life and widest operating temperature
    range.
    End point calibration temperatures of the instrument are then
    selected on the Model 22 and any required adjustments are made.
    Intermediate points may then be selected to verify instrument
    linearity or check critical points.
    Operating instructions  
    						
    							PN 1886665 Rev A April 2002
    ©2002 Altek Industries, Inc. Specifications subject to change without notice. 
    All rights reserved. Printed in U.S.A.  
    Altek products are warranted to be free from defects
    in material and workmanship (excluding fuses,
    batteries and leads) for a period of three years from
    the date of shipment. Warranty repairs can be
    obtained by returning the equipment prepaid to our
    factory. Products will be replaced, repaired, or
    adjusted at our option. Altek gives no other
    warranties, including any implied warranty of fitness
    for a particular purpose.Also, Altek shall not be liable
    for any special, indirect, incidental or consequential
    damages or losses arising from the sale or use of its
    products.
    Warranty
    (Unless otherwise indicated, specifications are in ±% of span@ 23˚C)
    Accuracy:  Positive temperatures ±0.1% of span ±1 degree. 
    Negative temperatures ±0.1% of span ±3 degrees
    Cold junction compensation:  Built-in for specified T/C type
    Cold junction temperature effect:  Within ±0.25 degree at 75°F
    (20°C) ±0.025 degrees/degree change in ambient
    Operating temperature range:  -10 to +130°F (-25 to +55°C)
    Ambient temperature effect: Zero:  Included in cold junction effect
    Span:  ±0.01% of span/degree
    Specifications
    Thermocouple types and ranges
    TYPE B,R,S DEG C TYPE E,J,K,T DEG F TYPE J,K DEG F TYPE E,J,K DEG F TYPE B,R,S DEG F
    TYPE E,J,K,T DEG C TYPE E,J,K DEG C
    Storage temperature limits:  -40 to +160°F (-40 to +70°C)
    Output impedance:  Fixed, 50 ohms nominal
    Batteries:  2 AA alkaline cells provide 1 year of use at nominal 4
    hours each workday. Batteries should be removed when storing the
    unit longer than 3 months.
    Battery indicator:  LED light pulse at turn-on in either range
    Size:  2 1/8 x 4 x 2 1/4 inches (54 x 102 x 55 mm)
    Weight:  6 oz. (0.15 kg)
    Specify thermocouple type, scale range and °F or °C to insure
    prompt delivery of all the above standard ranges. A 22J 1000F
    completely describes an Altek 22 position source providing a
    maximum output of 1000°F for a type J thermocouple
    Chromel and Alumel are registered trademarks of 
    Hoskins Manufacturing Company.
    BPt 30% Rh - Pt 6% Rh
    EChromel®- Constantan
    JIron - Constantan
    KChromel
    ®- Alumel®
    RPt - Pt 13% Rh
    SPt - Pt 10% Rh
    TCopper - Constantan
    500
    1000
    1700Maximum
    2100Scale
    3100ReadingF -Degrees Fahrenheit
    C -Degrees Celsius
    22