Home > Altek > Calibrator > Altek Model 11 Step RTD Simulator Manual

Altek Model 11 Step RTD Simulator Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Altek Model 11 Step RTD Simulator Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 Altek manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							RTD Simulator
    Model 11
    • Direct temperature outputInstant simulation of 11 temperature points
    • 0.1°C accuracy25, 50, or 100 degree resolution
    • Simple to useReplaces cumbersome decade box
    • Worldwide Pt 100 Ohm RTD CurveDIN/IEC 751, Alpha=1.3850Altek Model 11 Resistance Temperature Detector
    (RTD) Simulators provide 11 precise temperatures for
    inputs to transmitters, recorders, controllers, alarms,
    data acquisition and computer systems. The Model 11
    allows 2, 3 or 4 wire connections. Factory calibration
    includes compensation for the resistance of the built-
    in lead wires. Conformity to the particular RTD curve
    is in accordance with the latest standards for exact
    temperature simulation.
    Simulate temperature into any instument that
    measures a Pt 100 ohm RTD based on the worldwide
    DIN/IEC 751 standard curve with an alpha of 1.3850
    (sometimes displayed as 0.00385 or 385). Resolution
    is 25, 50 or 100 degrees, corresponding to full scale
    output of 250, 500 and 1000 degrees respectively.
    Positive switch selection allows quick, easy settings
    for any output. Calibrated accuracy is ±0.1°C.
    The low cost Model 11 is a complete compact
    simulator for checkout and calibration of all RTD
    instruments in the field, shop or control room.
    General description  
    						
    							PN 1886547 Rev A April 2002
    ©2002 Altek Industries, Inc. Specifications subject to change without notice. 
    All rights reserved. Printed in U.S.A.  
    Altek products are warranted to be free from defects
    in material and workmanship (excluding fuses,
    batteries and leads) for a period of three years from
    the date of shipment. Warranty repairs can be
    obtained by returning the equipment prepaid to our
    factory. Products will be replaced, repaired, or
    adjusted at our option. Altek gives no other
    warranties, including any implied warranty of fitness
    for a particular purpose.Also, Altek shall not be liable
    for any special, indirect, incidental or consequential
    damages or losses arising from the sale or use of its
    products.
    Warranty
    (Unless otherwise indicated, all specifications are referred to at an
    ambient temperature of 23˚C)
    Accuracy: ±0.1°C (±0.2°F)
    Temperature drift:  ±0.01%/°C
    Allowable excitation current:  0 to 15 milliamps, steady or intermittent
    Operating temperature range:  -25 to 55°C (-10 to +130°F)
    Storage temperature range:  -40 to 70°C (-40 to 180°F)
    Relative humidity:  10 to 90%, non-condensing
    Size:  2 1/8 x 4 x 2 1/4 inches (54 x 102 x 55 mm)
    Weight:  3 ounces (0.12 kg)
    Specifications
    Connections
    Select the Model 11 for correct temperature scale (F or C), and range.
    Connect the leads tightly to the input terminals of the instrument to be
    tested, the field mounted head or the junction box terminals. Set the
    temperature selection switch to the desired temperature. End point
    calibration temperatures of the instrument are then selected on the
    Model 11 and any required adjustments are made. Intermediate points
    may be selected to verify instrument linearity or check critical points.
    Operating instructions
    Order Model   
    11-250F
    11-250C
    11-500F
    11-500C
    11-1000F
    25° steps
    50° steps 100° steps 
    						
    All Altek manuals Comments (0)