    							DATA SHEET 530
    • 4000 COUNT (33/4digits) DIGITAL DISPLAY
    • 50 SEGMENT ANALOG SCALE 
    • BACKLIT DISPLAY
    • TEST VOLTAGES FROM 250 to 1000 VDC
    • AUTORANGING INSULATION RESISTANCE MEASUREMENT TO  4000M ΩON DIGITAL
    DISPLAY AND 10G Ω(10,000M Ω) ON ANALOG
    SCALE
    • AC VOLTAGE CHECK TO 600V
    • BATTERY POWERED AND AUTO OFF
    • LOCKING TEST BUTTON FOR DIELECTRIC ABSORPTION TESTING
    • RS-232 INTERFACE FOR RECORDING DATA
    • AUTOMATIC DISCHARGE OF ITEM UNDER TEST
    • DESIGNED TO IEC 61010-1 SAFETY STANDARDS
    GENERAL INFORMATION
    ALTEK
    AUTORANGING INSULATION TESTER MODEL 530
    Testing Insulation
    Virtually everywhere you go, electric wire can be
    found. Motors, transformers, generators, cables, etc.
    all have wiring covered with some type of insulating
    material. A variety of problems can occur if the
    insulation is allowed to deteriorate. The most
    common causes of deterioration in insulation are
    caused by environmental and/or physical damage.
    Moisture and extremes in temperature, vibration,
    abrasions, cuts or other mechanical damage, dust,
    dirt, grease or oil, plus normal electrical stress all
    contribute. The best way to evaluate insulation
    quality is to take measurements at regularly
    scheduled intervals so trends can be monitored.
    Planned reconditioning of motors and equipment can
    then be accomplished without risk of serious injury or
    catastrophic failures in your manufacturing process.
    Autoranging Insulation Tester
    The Altek Model 530 is an autoranging insulation
    tester. Use the 530 as the standard tool for testing your insulation. The reliable, straightforward
    megohmmeter is simple to use and easy to
    read. The quality backlight illuminates the
    display, making it easy to read even in the dark.
    The 530 provides its rated test voltages into a
    1mA load, so youre sure that the full voltage is
    applied to the tested device. The instrument
    also offers continuity testing, with a short circuit
    current of at least 200mA and an open circuit
    voltage of 12.8V for your safety.
    Data Logging
    If record keeping is important to you, use the
    RS-232 serial port and the provided software on
    a Windows based computer to record your
    measurements. The insulation tester will
    automatically power off after 30 minutes of
    inactivity to preserve battery life. The 530 can be used for routine preventive
    maintenance as well as acceptance testing, QA
    testing in manufacturing, and diagnostic testing.
    ALTEK INDUSTRIES CORP 
    						
    							SPECIFICATIONS
    Unless otherwise indicated, specifications are for 1 year in ±(% of reading + number of digits) @ 23°C
    General
    Display: 3.1 x 2.0 LCD. 3 
    3/4digits (4,000 counts) Digital
    readout with an analog bar indication. Backlighting available
    Measurement Range: 600VAC, 40Ω, 4000M Ω/250V,
    4000M Ω/500V, 4000M Ω/1000V
    Sampling Rate: 2.5 times/second for digital reading
    10 times/second for analog indication
    Auto Power Off: Approx. 30 minutes of inactivity
    Over Range Indicator: OL will be displayed
    Operating Temperature  Range: 32°F to 104°F (0°C to
    40°C)
    Storage Temperature  Range: 14°F to 140°F (-10°C to
    60°C)
    Operating Relative Humidity: 0% to 80% (noncondensing)
    Storage Relative Humidity: 0% to 70% (noncondensing)
    Altitude Range: Up to 6,500 feet
    Power Supply: 12VDC, (8 x 1.5V Size AA batteries)
    Power Consumption:
    Approx. 90mA (4000M Ω/1000V range) (output open circuit)
    Approx. 60mA (4000M Ω/500V range) (output open circuit)
    Approx. 45mA (4000M Ω/250V range) (output open circuit)
    Approx. 20mA (ACV and OHMS range)
    Safety Standards: Designed to IEC 61010-1 
    Overall Size: 7.5L x 5.5W x 3.0H (190mm x 140mm x 77mm)
    Weight: Approx. 2Lbs (900g), includes battery
    Specifications subject to change without notice
    © August 2000 Altek Industries Corp 
    100912-901 Rev A
    ADDITIONAL INFORMATION
    LIFETIME WARRANTY  PROGRAM
    Our equipment is guaranteed against defective material and workmanship
    (excluding batteries and test leads) for a period of one year. For a lifetime
    warranty: (1) Simply register the unit upon receipt by completing and
    returning the products Warranty Registration Card. (2) Once a year, send
    the unit to any Altek authorized distribution center for factory re-
    certification. Claims under warranty can be made by returning the
    equipment prepaid to any authorized Altek distributor. Altek will repair, re-
    calibrate or replace free of charge any unit found to be defective or out of
    calibration at any time between re-certification intervals. The liability of
    Altek is restricted to that given under our guarantee. No responsibility is
    accepted for damage, loss or other expense incurred through sale or use
    of our equipment. Under no condition shall Altek be held liable for any
    special, incidental or consequential damage.
    OTHER PRODUCTS
    Altek designs and manufacturers fast, accurate instruments for
    measurement, generation and simulation of virtually every process control
    signal. Consult our factory directly or contact your local stocking
    representative to order precise, low cost Milliamp Calibrators, Voltage
    Sources, Thermocouple Sources, RTD Simulators, Frequency Calibrators
    and Pressure Pumps, Indicators & Calibrators, as well as electrical test
    instruments. Altek also produces calibrators for custom ranges and unique
    applications. New models are frequently added to the Altek family to meet
    all of your critical calibration requirements.
    ORDERING INFORMATION Part No.
    Autoranging Insulation Tester 530
    Included with each Model 530 are: Instruction Manual 100912-900
    Altek 530 Windows Software (Diskette)
    Serial Cable (Optical to Female DB-9)
    8 x 1.5V Size AA Batteries
    Certificate of Conformity and Lifetime Warranty
    AVAILABLE FROM
    OHMS & Continuity Beeper
    Max. Open   Max. ShortCircuit          Circuit       Overload
    Range   Resolution   Accuracy    Voltage         Current     Protection
    40 Ω 0.01Ω 1%+30        12.8V           280mA* 220Vrms
    * Minimum Value: 200mA
    Max. OpenCircuit Overload
    Range   Resolution        Beeper Active            Voltage Protection
    Beeper 0.01 ΩResistance
    						