    							Transmitter Simulator
    Model 46
    • 2-wire simulatorCalibrate Receivers
    • 5 point calibration4, 8 12, 16, 20mA
    • Loop poweredNo Batteries Required
    • Easy to useOne Hand OperationSimulate 2-wire transmitters to calibrate receivers,
    recorders, controllers, indicators and computers. Loop
    power eliminates the need for batteries. Loop current
    is clamped at the key value you select. Accuracy is
    ±0.08% of 4 to 20 mA, even with changes in load and
    supply voltage.
    Switch settings of 0, 25, 50, 75 and 100% give you
    fast, full range checkout of instrument or system. The
    single switch and palm sized case have been
    designed for simple, one handed operation. Just clip
    the built-in leads to the 2-wire terminals in the field,
    junction box or control room.
    When you check out receivers, recorders or
    controllers in a 2-wire 4 to 20mA loop, simply
    disconnect the 2-wire transmitter and hook up a
    Model 46. Flip a switch to get instant full scale and
    zero outputs. The calibrator operates just like a 2-wire
    transmitter by modulating the external power supply
    voltage into a 4 to 20mA signal.
    The Model 46 receives all of its operating power from
    the loop. This lets you operate continuously for days
    when long term loop testing is required. Use it as a
    manual loading station when a controller needs repair
    or replacement. Minimize process shutdowns
    installing a new  controller.
    General Information 
    						
    							PN 1886437 Rev A April 2002
    ©2002 Altek Industries, Inc. Specifications subject to change without notice. 
    All rights reserved. Printed in U.S.A.  
    Operating Instructions
    Specifications
    (Unless otherwise indicated, specifications are in ±% of 16 mA span
    @ 23°C)
    Accuracy: ±0.08%
    Temperature effect: ±0.01%/Degree C
    Operating temperature range: -10 to 140°F (-25 to +60°C)
    Storage temperature range: -40 to +160°F (-40 to +70°C)
    Power supply effect: ±0.005%/Volt
    Warm up time: 2 Seconds to rated accuracy
    Weight: 3 ounces (85 grams)
    Size: 11
    /
    2x21
    /
    8x31
    /
    4inches (38x54x63 mm)
    Loop voltage limits: 6 to 45 VDC
    Simulate 2-wire transmitters
    1) Disconnect existing 2-Wire transmitter from the loop
    2) Connect the red lead of the calibrator to the plus (+) terminal of
    the field connections and the black lead to the minus (-).
    The LED will glow steadily to indicate that the loop current is
    present. If the LED doesn’t glow, check the connections and the
    power supply.
    Select the 0.0% and 100.0% switch positions to quickly calibrate the
    endpoints of your receivers. Intermediate positions of 25, 50 and
    75% may be selected.
    Altek products are warranted to be free from defects
    in material and workmanship (excluding fuses,
    batteries and leads) for a period of three years from
    the date of shipment. Warranty repairs can be
    obtained by returning the equipment prepaid to our
    factory. Products will be replaced, repaired, or
    adjusted at our option. Altek gives no other
    warranties, including any implied warranty of fitness
    for a particular purpose.Also, Altek shall not be liable
    for any special, indirect, incidental or consequential
    damages or losses arising from the sale or use of its
    products.