    							ALTEK INDUSTRIES CORP
    ALTEK DATA SHEET
    The Model 334 may also be used when a controller is temporarily
    removed for repair or replacement. Simply substitute the Model
    334 for the controller and use it as a manual loading station.
    SIMULATE 2-WIRE TRANSMITTERS by modulating external power
    to pass 4 to 20mA. The Altek Model 334 uses any loop power from
    2 to 100 Volts DC. True current design maintains set current
    independent of voltage or load changes.
    READ MILLIAMPS OR PERCENT of the 4 to 20mA signal. Maxi-
    mum read current is limited to (nominally) 55mA to minimize the
    possibility of damaging the device to be calibrated.
    MEASURE DC Volts range displays from-99.99 to +99.99 VDC with
    10mV resolution. Built in voltmeter is used to check loop power
    supplies, I/V converters, 1 to 5 Volt signals and battery voltages.
    The Model 334 is protected in all ranges to 135 Volts AC or DC.
    Special protective circuitry withstands accidental misconnection in
    any mode 
    without fuses.
    Display digits are 0.4 (10.2 mm) high for readability from across
    the room. Non-glare Liquid Crystal Display is readable in any light.
    . . even in direct sunlight. The digital measuring circuit is indepen-
    dent of the current adjustment and measures the actual input
    or output.
    Benchtop accuracy in a toolbox calibrator assures fast, precise set-
    ting of current trips, recorders, controllers, loggers, computers and
    final control elements. Altek brings you the handy Model 334 Loop
    Calibrator at a cost low enough for every bench and toolbox.
    Altek
    ’s Model 334 Loop Calibrator has been designed to check,
    calibrate and trouble shoot all the signals present within a standard
    4 to 20 milliamp process control loop. A 24 mA Source, mA meter,
    2-Wire Transmitter Simulator and a DC Volt meter are combined
    into one easy-to-operate instrument.
    Dual current ranges allow display of each current function in
    percent of 4 to 20mA signal, as well as directly in milliamps. In
    SOURCE MODE, percent range displays -25.0% to +125.0% with
    0.1% resolution while the milliamp range displays from 0.00 to 24.00
    mA with 10 µA resolution. In READ MODE the milliamp range
    displays from 0.00mA to 52.00mA, or percent.
    SOURCE MODE uses built in batteries to provide 0.00 to 24.00 mA
    into any load from 0 to 1200 Ohms. Three 9 volt alkaline
    batteries provide over 30 hours of continuous 12 mA output . . .
    more than 3 months of typical use. An optional AC adaptor plugs
    in for continuous bench use. True current source maintains set
    output independent of load,  “QUIK-CHEK ” switch provides instant
    zero (4.00mA) and full scale (20.00mA) in both SOURCE and 2
    WIRE modes. Output in the PERCENT range displays 0.0% and
    100.0% for ”QUlK-CHEK ” settings. ADJUST position selects a
    continuously adjustable 10 turn potentiometer to allow fast, easy
    setting to any exact value from 0.00 to 24.00mA. No range change
    at critical 20.00mA eliminates hysteresis error.
    POWER & MEASURE 2-Wire Transmitters simultaneously in Source
    Mode. The Model 334 supplies the required voltage and displays
    the current as regulated by the 2-Wire Transmitter.
    MILLIAMP LOOP CALIBRATOR MODEL 334
    • ALL 4 TO 20mA LOOP FUNCTIONS
    Source 0.00 to 24.00mA or-25.0 to 125.0%
    Read mA or Percent to 52.00mA
    Simulate 2-Wire Transmitters
    Measure -99.99 to 99.99 DC Volts
    Power and Read 2-Wire Transmitters
    • “QUIK CHEK
    ® ” SWITCH
    Instant Zero and Span at 4.00 & 20.00mA
    • FULL 4 DIGIT ACCURACY    0.05% of 20.00mA Span
       0.03%  “QUIK-CHEK ”
    • STANDARD 9 VOLT BATTERIES Months of use, no nightly recharging
    •  OVERLOAD PROTECTED
    Withstands 135 Volts AC/DC without fuses
    334
    +
    +
    OTHER PRODUCTSAltek designs and manufactures fast, accurate instruments for mea-
    surement, generation and simulation of virtually every process control
    signal. Consult our factory directly or contact your local stocking
    representative to order precise, low cost Milliamp Calibrators, Voltage
    Sources, Direct Thermocouple Sources, RTD Simulators and Fre-
    quency Sources. Altek also produces calibrators for custom ranges
    and unique applications. Additional models and ranges are frequently
    added to the Altek instrument family to meet all of your critical
    calibration requirements. Altek products are made in the USA.
    WARRANTY
    Our equipment is guaranteed against defective material and work-
    manship (excluding batteries) for a period of three years from date
    of shipment. Claims under guarantee can be made by returning the
    equipment prepaid to our factory. The equipment will be replaced,
    repaired or adjusted at our option. The liability of Altek is restricted
    to that given under our guarantee. No responsibility is accepted for
    damage, loss or other expense incurred through sale or use of our
    equipment. Under no condition shall Altek be liable for any special,
    incidental or consequential damage.
    GENERAL DESCRIPTION 
    						
    							SOURCE MILLIAMPS
    1) Disconnect one or both input wires from the device to be cali-brated
    2) Connect the red lead of the calibrator to the plus (+) input of th\
    e device and the black lead to the minus (-)
    3) Move the power switch to mA or %
    Output current is continuously adjustable with the  “QUIK-CHEK ” switch
    in the ADJUST position. Exact 0.0% (4.00mA) and 100.0% (20.00mA)
    outputs are available by using the 0.0 and 100.0%  “QUlK-CHECKs.
    ”
    SIMULATE 2-WIRE TRANSMITTERS
    1) Disconnect existing 2-Wire transmitter from the loop
    2) Put the Mode selector switch into the 2 WIRE position
    3) Connect the red lead of the calibrator to the plus (+) terminal of\
    the field connections and the black lead to the minus (-)
    4) Move the power switch to mA or %
    Select the 0.0% (4.00mA) and 100.0% (20.00mA)  “QUIK-CHEK ” switch
    positions to quickly calibrate the endpoints of your receivers. Use the
    ADJUST position and dial to select any value from -25.0 (0.00mA) to
    125.0% (24.00 mA).
    RECEIVER
    (Powers External
    2-Wire Transmitters)
    POWER
    SUPPLY 2 to
    100 VDCZEROSENSORSPAN
    2-WIRE TRANSMITTER (Disconnected)
    20.00
    334
    4.00
    334
    VA LV E
    POWER
    SWITCH
    MODE
    SWITCH
    ADJUST “
    QUIK-
    CHEK ”
    SWITCH OR
    RECEIVER
    I/P
    Specifications subject to change without notice
    READ MILLIAMPS
    1) Open the current loop at any convenient point along the signal
    path
    2) Connect the red lead of the calibrator to the more positive point of the break and the black lead to the more negative
    3) Put the Mode selector switch into the READ position
    4) Move the power switch to mA or%. The Model 334 will read current in the loop from 0.00mA to 52.00mA or-25.0% to 300%.
    Loop current is automatically limited to approximately 55mA to
    avoid any damage to other devices within the loop. If the Model
    334 is accidently connected in the wrong polarity, the display will
    read 0.00 mA or-25.0% and no current will flow in the loop.
    Simply reverse the leads for correct operation.
    Note:  “QUIK-CHEK ” switch is disabled when readingPOWER & MEASURE 2-WIRE TRANSMITTERS
    1) Disconnect one or both output wires from the 2-Wire Transmitter
    to be calibrated
    2) Connect the red lead of the calibrator to the plus (+) and the black lead to the minus (-) power connections of the 2-Wire
    Transmitter
    3) Connect an appropriate sensor or calibrator to the input of the
     2-Wire Transmitter
    4) Move the power switch to mA or %, the  “QUIK-CHEK ” switch to
    ADJUST, and rotate the knob fully clockwise (see shaded areas
    on face of unit)
    The Model 334 will supply a (nominal) 24 Volts DC to the 2-Wire
    Transmitter. The milliamp current passed by the transmitter will be
    accurately displayed. Calibrate the transmitter in the usual manner
    and disconnect the Model 334.
    20.00
    334
    T/C
    RTD mA
    mV
    SENSOR OR
    CALIBRATOR
    SENSOR OUTPUT
    2-WIRE
    TRANSMITTER
    60.5
    334
    COMPUTER
    CONTROLLER RECEIVER
    POWER
    SUPPLY
    SENSOR OUTPUT
    MEASURE VOLTS DC
    1) Put the mode switch to READ
    2) Move the power switch to READ DCV
    3) Connect the red and black leads across the voltage to be measured
    Any DC Voltage from -99.99 to +99.99 Volts may be measured. Loop
    power supplies, signal voltages at receivers, batteries and transmitter
    voltage drops may be measured. Signals exceeding 
    ±100 VDC will be
    indicated by OVRLD on the LCD.
    Note:  “QUIK-CHEK ” switch is disabled when reading
    24.00
    334
    DC
    POWER
    SUPPLY OR
    BATTERY
    GENERALACCURACY:  ±
    0.05%
    “ QUIK-CHEK ”, Source & 2-Wire Modes, Accuracy  ±
    0.03%
    DISPLAY: Liquid Crystal; 4 Digit, 0.4 (10.2mm) high
    TEMPERATURE EFFECT: Milliamp Ranges  ±
    0.008%/ ±
    C: Voltage Ranges  ±
    0,012%/ °CBUILT-IN BATTERIES: 3 x 9 Volt Alkaline
    AC ADAPTORS: Optional, 120 or 240 VAC, 50/60 Hz
    BATTERY LIFE:
      Batteries should be removed when storing the unit >3 months.
       SOURCE MODE: Nominal 33 hours at 12mA, Nominal 20 hours
    at 20mA output
       2-WIRE SIMULATOR or READ MODES: Nominal 150 hours
    LOW BATTERY INDICATION: BAT indicated on LCD when approxi-
    mately 1 hour remains driving 12mA into 800 Ohms
    OVERVOLTAGE PROTECTION: Protected to 135 Volts AC or DC in
    all ranges without fuses
    OVERLOAD PROTECTION:
       SOURCE & 2-WIRE SIMULATOR MODES: Current limited to 25mA   READ MODE: Current limited to 55mA
    OPERATING AMBIENT TEMPERATURE: -20 to +60 °C (-5 to +140 °F)
    STORAGE TEMPERATURE: -30 to +70 °C (-22 to +160 °F)
    RELATIVE HUMIDITY: 10 to 90%, non-condensing
    WARM UP TIME: 3 seconds to rated accuracy
    OVERALL SIZE: 2 1/2 x 2 5/8 x5 1/8 inches (63.5 x 66.7 x 130mm)
    WEIGHT: 12 oz. (0.35 kg)
    SOURCE MODERANGES:
       0.00 to 24.00 mA with 10  µA resolution
       -25.0 to 125.0% of 4-20mA with 0.1% resolution
    TYPICAL DRIVE CAPABILITY: 1200 Ohms with fresh batteries or
    adaptor, 800 Ohms at low battery indication (BAT)
    POWER TO EXTERNAL 2-WIRE TRANSMITTER: Nominal 24 VDC
    with fresh batteries or adaptor
    (Unless otherwise indicated, specifications are in ± (% of reading + 1 least significant digit) at 23 °C)
    © June 1999 Altek Industries Corp 
                                                                        1008\
    81-900 REV:A
    AVAILABLE FROM
    READ MODERANGES:
       0.00 to 52.00mA with 10  µA resolution
       >24 mA read  ±3 least significant digits
       -25.0 to 300.00% of 4-20mA with 0.1% resolution
        -99.99 to 99.99 VDC with 10mV resolution, OVRLD at +100 VDC
    INPUT IMPEDANCE: VDC range, >2 Meg Ohms
    VOLTAGE BURDEN: 0.9V at 4.00mA, 1.2 V at 20mA, 1.9 V at 50.00mA
    2-WIRE TRANSMITTER SIMULATOR MODERANGES:
       0.00 to 24.00mA with 10  µA resolution
       -25.0 to 125% of 4-20mA with 0.1% resolution
    LOOP VOLTAGE LIMITS: Minimum, 2 VDC; Maximum 100 VDC
    ORDERING INFORMATION      Part No.MODEL 334: LOOP CALIBRATOR                               334-2050
    AC Adaptor          28-0120
    AC Adaptor          28-0240
    CARRYING CASE: Optional, zippered with belt loop      09-3781
    OPERATING INSTRUCTIONS
    SPECIFICATIONS 
    						