    							ALTEK20 MILLIAMP SIGNAL ANALYZER
    MODEL 234
    • TRIPLE FUNCTION, 0 TO 20mA & -25 TO +125%
    Source, Read & 2-Wire simulate
    • PRECISE 3-1/2 DIGIT LCD
    ±0.1% of 4-20mA SPAN in Milliamps or Percent
    • “QUIK-CHEK® 
    ” ZERO AND FULL SCALE
    Switch 4.00, 20.00mA or Continuous Dial
    • POCKET SIZE & LIGHTWEIGHT
    12.5 oz. (0.35 kg) with 3 Month Batteries
    • THREE YEAR WARRANTY
    Tough enough for your ToolboxGENERAL DESCRIPTION
    ALTEK’s triple function MODEL 234 SIGNAL ANALYZER combines
    a self contained 0-24mA SOURCE, a 2-WIRE SIMULATOR and a
    large LCD DIGITAL READOUT in a single pocket sized instrument.
    DUAL  RANGE  allows  display  of  each  function  in  PERCENT  of
    4-20mA signal, as well as directly in MILLIAMPS. PERCENT range
    displ ays  minus  25.0%  to  plus  125.0%  with  0.1%  resolultion.
    MILLIAMP range displays from 0.00 to 19.99mA with 10 microampresolution. ACCURACY is ±0.1% of 4-20mA ±1/2 least significant digit.“QUIK-CHEK” switch provides instant zero (4.00mA) and full scale
    (20.00mA) settings in both  SOURCE and 2-WIRE modes.  Output
    in  the  PERCENT  range  displays  00.0%  and  100.0%  for  “QUIK-
    CHEK”  settings.  DIAL  position  selects  a  continuously  adjustable
    10 turn potentiometer to allow fast, easy setting to any exact value.
    SOURCE MODE uses built-in batteries to provide 0-24mA into any
    load from 0-1000 ohms. 3 nine volt alkaline batteries provide over
    20 hours of continuous 20mA output…more than three months of
    typical use. An optional AC adapter plugs in for continuous bench
    us e.   True  current  source  maintains  set  output  independent  of  load.
    2-WIRE TRANSMITTERS can be POWERED and MEASURED in this mode.2-WIRE  SIMULATOR  mode  modulates  external  power  to  pass
    4-20mA. The  ALTEK  MODEL  234  uses  any  LOOP  POWER  from
    2  to  100  volts  D C.  True  current  design  maintains  set  current
    independent of voltage or load changes.
    READ mode displays either MILLIAMPS or PERCENT of 4-20mA
    signal.  Maximum  READ  current  is  limited  to  (nominally)  25mA  to
    minimize the possibility of damaging the calibrator or the loop to be
    cal ibrat ed.  S pec ial   protec tiv e  ci rcui try  w iths tand s  acc ide ntal
    misconnection in any mode without fuses.
    DISPLAY digits are 0.350 (9mm) high for  readability from across the
    room.  Non-glare  LIQUID  CRYSTAL  DISPLAY  is  readable  in  any
    light…even in direct sunlight. The digital measuring circuit is indepen-
    dent of the current adjustment and measures the actual input or output.The ALTEK MODEL 234 is rugged, yet lightweight and pocket sized.
    Latest LSI circuitry and wide temperature range components make
    the Model 234 ideal for use in the field, control room and shop.
    BENCHTOP  ACCURACY  in  a  TOOLBOX  CALIBRATOR  assures
    fast, precise setting of current trips, recorders, controllers, loggers,
    computers and final control elements. ALTEK brings you the handy
    MODEL  234  SIGNAL  ANALYZER  at  a  cost  low  enough  for  every
    bench and toolbox.SPECIFICATIONS ALTEK MODEL 234
    (Unless otherwise indicated, specifications are in % of 16mA span at 23°C)ACCURACY: ±0.1% (±1/2 least significant digit)
    DISPLAY: Liquid Crystal; 31/2 digit, 0.35 (9.0mm) high
    DISPLAY RANGES: 0.00 to 19.99 mA (-25.0 to +125.0%)
    RESOLUTION: 0.1% (percent mode) 0.01mA (mA mode)
    READ PROTECTION: Limited to nominal 25mA against accidental
    misconnection
    OVERVOLTAGE  PROTECTION:  Protected  to  120 Volts  AC  or  DC
    in all ranges without fuses
    TEMPERATURE EFFECT: ±0.01%/Degree C
    OU TP U T/2- W IR E   R A N GE S :  0  to  2 4mA   D C   (-25   to  + 125 %)
    continuous
    “QUIK-CHEK” SETTINGS: 4.00 and 20.00mA (00.0 and 100.0%)
    OUTPUT  DRIVE  CAPABILITY:  1000  ohms  with  fresh  batteries  or
    adapter, 800 ohms at first low battery warning
    BUILT-IN  BATTERIES: 3 x  9 volt alkaline, included
    POWER  TO  EXTERNAL  2-WIRE  TRANSMITTER:  24  volts  with
    fresh batteries or adapter
    BATTERY LIFE: SOURCE; Nominal 20 hours at 20mA continuous.
    READ and 2-WIRE mode; Nominal 150 hours. 
    Batteries should be
    removed when storing the unit >3 months.
    AC ADAPTERS: optional; 120 or 240V, 50/60 Hz
    LOW VOLTAGE INDICATORS: LO BAT ARROW         turns on at 18
    volts; decimal point turns off at 9 volts
    LOOP   VOLTAGE   LIMITS : 2-W i re  S imul ator  Mode;  mi ni mum,
    2V DC; maximum, 100V DC
    OPERATING  AMB IENT  TEMP ERATURE : Minus  5  to  plus  140
    degrees F (minus 20 to plus 60 degrees C)
    STORAGE   TEMPERATURE:  minus  22  to  plus  175  degrees  F
    (minus 30 to plus 80 degrees C)
    RELATIVE HUMIDITY: 10 to 90%, non-condensing
    WARM UP TIME: 3 seconds to rated accuracy
    OVERALL SIZE: 21/2 x 25/8 x 51/8 inches (63.5 x 66.7 x 130 mm)
    WEIGHT: 12.5 oz (0.35 kg)
    CARRYING CASE: Optional, zippered with belt loop.
    Specifications subject to change without notice 
    						
    							OPERATING INSTRUCTIONSSOURCE MILLIAMPS
    Disconnect one or both input wires from the device to be checked
    or calibrated. Attach the red (+) lead of the ALTEK MODEL 234 to
    the plus input of  the device  to be calibrated, connect the black (-)
    lead to the minus terminal. Turn the SOURCE/READ/2-WIRE mode
    selector switch to the SOURCE position. Turn on the Model 234 to
    the  PERCENT  or  mA  position,  as  desired.  Actual  current  sent
    through the loop is seen as  a PERCENT or MILLIAMP reading in
    the LCD display.
    Output  current  is  continuously  adjustable  with  the  “QUIK-CHEK”
    (0%/D IA L/ 10 0%)  sw i tch  i n  the  D IA L  p os i ti on.  Th e  S OU R C E
    current can be set from 0.00mA (-25.0%) to 24 mA (125.0%). In the
    mA mode, the display reads up to 19.99 mA. For display of higher
    outputs,  use  the  PERCENT  position.  Exact  4.00  and  20.00mA
    output is instantly available in either mode by the use of the “QUIK-
    CHEK” switch.SIMULATE 2-WIRE TRANSMITTERS
    Disconnect  any  existing  2-wire  transmitter  from  the  loop  to  be
    checked or calibrated. (Only one of the two wires need be discon-
    nected.) Leave the power source and the receiver in place.
    Put the mode selector in the 2-WIRE position. Connect the red lead
    of the ALTEK Model 234 to the plus (+) terminal of the field connec-
    tions. Connect the black lead to the minus (-) terminal and turn the
    234 on. The LCD display will read the loop current in PERCENT or
    mA ,  as   s el ec ted .  0%  (4 .0 0mA )  D IA L   adj us tm ent   and   100 %
    (20.00mA)  are  selected  with  the  right  hand  “QUIK-CHEK”  switch.
    Instant ZERO and FULL SCALE checks are made with the 0% and
    100%  positions. The  DIAL  mode  allows  continuous  adjustment  of
    t he  2 -w i re  c ur re nt   f rom  0 . 00m A   (-2 5. 0%)   t o  24 mA   ( 125 %).
    Complete any necessary recalibration  and turn the ALTEK Model
    234 off. Reconnect any wires removed for checkout.READ MILLIAMPS
    Open the current loop at any convenient point along the signal path.
    Connect  the  ALTEK  Model  234  Red  (+)  lead  to  the  more  positive
    point  of  the  break.  Connect  the  Black  (-)  terminal  to  the  more
    negative side of the break. Put the left hand mode select switch in
    the  “READ”  position  and  turn  the  unit  on  in  either  MILLIAMP  or
    PERCENT display position. The Model 234 will read current in the
    loop from 0 to 19.99mA or -25 to +125% (24mA).
    LOOP  CURRENT  is  automatically  limited  to  approximately  25mA
    to  avoid  any  damage  to  the  calibrator  or  other  devices  within  the
    loop.  If  the  Model  234  is  accidentally  connected  in  the  wrong
    polarity, the display will read 0.00mA or -25.0% and no current will
    flow in the loop. Simply reverse the leads for correct operation.POWER & MEASURE 2-WIRE TRANSMITTERS
    Connect the Model 234 (+) and (-) leads to the power leads of the
    2-wire  transmitter  to  be  powered  and  measured.  Connect  the
    appropriate  signal  to  the  input  of  the  2-wire  transmitter.  Set  the
    “QUIK-CHEK” switch to the DIAL position. Turn the DIAL fully clock-
    wise  (maximum  position).  Set  the  mode  switch  to  the  SOURCE
    position. Turn the 234 on to either the mA or the PERCENT display
    modes. 24 volts (nominal) will be supplied to the 2-wire transmitter
    by the 234. The milliamp current passed by the transmitter will be
    accurately  displayed. Calibrate  the  2-wire transmitter in  the  usual
    manner and disconnect the 234.ORDERING INFORMATION:Part No.
    MODEL 234: 20 Milliamp Signal Analyzer      234-0420
    AC Adapter: 120V AC       28-0120
    AD Adapter: 240V AC       28-0240
    Carrying Case       09-3781
    © June 1999 Altek Industries CorpALTEK
    234
    SIMULATOR
    MODE
    RECEIVERPOWER
    SOURCE
    TRANSMITTER
    OR
    CONTROLLER
    ALTEK
    234
    READ
    MODE
    RECEIVER
    COMPUTER
    RECORDER
    ALTEK
    234
    SOURCE
    MODE2-WIRE
    TRANSMITTERSENSOR OR
    CALIBRATOR
    INPUT
    ALTEK
    234
    SOURCE
    MODECOMPUTER
    RECEIVER
    LOGGER
    VALVE
    ALTEK
    234
    SOURCE
    MODEI/PVALVE 
    						