Home > Altek > Calibrator > Altek 134 mA Source2Wire Simulator Manual

Altek 134 mA Source2Wire Simulator Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Altek 134 mA Source2Wire Simulator Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 Altek manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							ALTEK INDUSTRIES CORP
    ALTEK DATA SHEET
    MILLIAMP CALIBRATORS
    Models 46, 134
    134
    ALTEK’s pocket sized milliamp calibrators offer  2-Wire transmitter simu-
    lators as well as milliamp sources. Each model give you ease of use,
    as well as accuracy and ruggedness both.
    When you check out receivers, recorders or controllers in a 2-Wire 4 to
    20mA loop, simply disconnect the 2-Wire transmitter and hook up a
    Model 46 or 134 in 2-Wire mode. Flip a switch to get instant Full Scale
    and Zero outputs. The calibrators operate just like a
    2-Wire transmitter by modulating the external power supply
    voltage into a 4 to 20mA signal.
    In the control room, shop, or field, use the Model 134 as a precise
    4 to 20mA Source. Stroke valves, set trip points or calibrate receivers
    or recorders. Easily replaceable 9 Volt alkaline batteries will source 20
    milliamps 
    continuously for over 25 hours. Many users can go for months
    between battery changes!
    Used as a 2-Wire Transmitter Simulator the Model 134 receives
    all of its operating power from the loop. This lets you operate
    continuously for days when long term loop testing is required.
    Use it as a manual loading station when a controller needs repair
    or replacement. Minimize process shutdowns installing a new
    controller.
    High efficiency design typically gives you months of operation
    before replacing the batteries. Optional AC adapters are available for full-
    time bench use.
    OTHER PRODUCTSAltek designs and manufactures fast, accurate instruments for mea-
    surement, generation and simulation of virtually every process control
    signal. Consult our factory directly or contact your local stocking
    representative to order precise, low cost Milliamp Calibrators, Voltage
    Sources, Direct Thermocouple Sources, RTD Simulators and Fre-
    quency Sources. Altek also produces calibrators for custom ranges
    and unique applications. Additional models and ranges are frequently
    added to the Altek instrument family to meet all of your critical
    calibration requirements. Altek products are made in the USA.
    MODEL 46
    • 2-WIRE SIMULATOR
    Calibrate Receivers
    • 5 POINT CALIBRATION 4, 8 12, 16, 20mA
    • LOOP POWERED No Batteries Required
    • EASY TO USE
    One Hand Operation
    Simulate 2-Wire Transmitters to calibrate
    receivers, recorders, controllers, indicators
    and computers. Loop Power eliminates the
    need for batteries. Loop current is CLAMPED
    at the key value you select. Accuracy is
    ±0.08% of 4 to 20mA, even with changes in
    load and supply voltage.
    Switch settings of 0, 25, 50, 75 and 100%
    give you fast, full range checkout of instru-
    ment or system. The single switch and palm
    sized case have been designed for simple,
    one handed operation. Just clip the Built-in
    leads to the 2-Wire terminals in the Field,
    Junction Box or Control Room.Calibrate computers, receivers and control-
    lers with unparalled ease. The Model 134
    is also ideal for setting current trips and
    stroking valves. Built-in batteries supply
    20mA loads from 0 to 1000 Ohms (typically
    3 months of field use).
    Simulate any 2-Wire Transmitter without
    using internal batteries. Current stays at de-
    sired value even with changes in load and
    supply voltage.
    “ QUIK-CHEK ” switch gives you instant
    settings of 0 and 100% for both the mA
    Source and 2-Wire Simulator. Set any value
    in dial position using the 10 turn continuous
    adjustment.
    MODEL 134
    • MILLIAMP OUTPUT
    Sources 4-20mA
    • 2-WIRE SIMULATOR
    Field or Panel
    • “QUIK-CHEK
    ® ” SWITCH
    Instant 4.00 & 20.00mA
    • CONTINUOUS DIAL
    0.02% Resolution 
    						
    							000 062(5)  100  200  250  300  400  500 562(5)  600  700  750  800  900  000
    134 (LINEAR)4.00 5.00 5.60 7.20 8.00 8.80 10.4 12.0 13.0 13.6 15.2 16.0 16.8 18.4 20.0
    (0, 25, 50, 75, 100% D/P FLOW)0% 25% 50% 75% 100%
    OPERATING INSTRUCTIONS
    250
    ALTEK134BATTERYCHECK LOOP
    ON
    • mA OUT   100% •
    • OFF    DIAL •
    • 2-WIRE      00.0% •
    4-20mA
    00.0-100%
    mA SOURCE
    2 WIRE SIMULATOR0.1%ALTEK INDUSTRIES CORP.
    ROCHESTER, N.Y. 14623   U.S.A.
    COMPUTER RECEIVERLOGGER VA LV E
    MODEL
    134
    500
    ALTEK134BATTERYCHECK LOOP
    ON
    •  mA OUT   100%  •
    •  OFF    DIAL  •
    •  2-WIRE     00.0%  •
    4-20mA
    00.0-100%
    mA SOURCE
    2 WIRE SIMULATOR0.1%ALTEK INDUSTRIES CORP.
    ROCHESTER, N.Y. 14623   U.S.A.
    SENSORSPAN ZERO
    RECEIVER
    (Powers External
    2-Wire Transmitters)
    POWER
    SUPPLY
    MODEL 46, 134
    2-WIRE TRANSMITTER
    ORDERING INFORMATION:                 Part No.Model 46       2-Wire Transmitter Simulator          46-0420
    Model 134     4 to 20mA Source/2-Wire Simulator        134-0420
    AC Adapters for Models 134       120 VAC Adapter          28-0120
          240 VAC Adapter          28-0240
    Carrying Case          09-3781WARRANTYOur equipment is guaranteed against defective material and work-
    manship (excluding batteries) for a period of three years from date
    of shipment. Claims under guarantee can be made by returning the
    equipment prepaid to our factory. The equipment will be replaced,
    repaired or adjusted at our option. No responsibility is accepted for
    damage, loss or other expense incurred through sale or use of our
    equipment. Under no condition shall Altek be liable for any special,
    incidental or consequential damage.
    ACCURACY:
    MODEL 46 …±0.08%
    MODEL 134
    Dial:  ±0.1% from 1% to 99%
    “ QUIK-CHEK ” Switch:  ±0.08% at 0.00% and 100.0%
    TEMPERATURE EFFECT:  ±0.01%/Degree C
    OPERATING TEMPERATURE RANGE: -10 to 140 °F (-25 to +60 °C)
    STORAGE TEMPERATURE LIMITS: -40 to +160 °F (-40 to +70 °C)
    POWER SUPPLY EFFECT:  ±0.005%/Volt
    WARM UP TIME: 2 Seconds to rated accuracy
    WEIGHT:
    Model 46 …3 ounces (85 grams)
    Models 134 …12 ounces (340 grams)
    SIZE:
    Model 46 …11/2x21/8x31/4 inches (38x54x63 mm)
    Models 134 …21/8x4x21/4 inches (54x102x55 mm)
    LOOP VOLTAGE LIMITS:
    Models 46 & 134 …6 to 45 VDC
    2-WIRE SIMULATOR MODE
    Models 46,134
    1) Disconnect existing 2-Wire transmitter from the loop
    2) Put the Mode selector switch (Models 134) into the 2-Wire  position
    3) Connect the red lead of the calibrator to the plus (+) terminal of\
     the field connections and the black lead to the minus (-).
    The LED will glow steadily to indicate that the loop current is present.\
    If the LED doesn ’t glow, check the connections and the power supply.
    Select the 0.0% and 100.0% switch positions to quickly calibrate
    the endpoints of your receivers. Intermediate positions of 25, 50
    and 75% may be selected with the Model 46. Use the digital dial on
    the Models 134 to select any value from 0.0 to 100.0% (a setting of
    645 on the dial is equal to 64.5%).
    SOURCE MODE
    Models 134
    1) Disconnect one or both wires from the device to be calibrated
    2) Connect the red lead of the calibrator to the plus (+) input of th\
    e device and the black lead to the minus (-).
    3) Move the Mode to mA OUT
    The LED on the Model 134 will glow steadily only when current is
    flowing through the loop. If the LED doesn ’t light, check the batter-
    ies and connections.
    Output current is continuously adjustable with the  “QUIK-CHEK ”
    switch in the DIAL position. Exact 0.0 and 100.0% outputs are avail-
    able by using the 0.0 and 100.0%  “QUIK-CHEKs ”.
    SPECIFICATIONS
    OUTPUT DRIVE CAPABILITY:
    Model Fresh Batteries or AC Adapter        Battery Low Limit
    134 1000 Ohms 800 Ohms
    BATTERY LIFE:  (Nominal):
    Model 
     Nominal Life —Alkaline Battery134        25 hours at 20mA continuous output      (3-9 Volt)
    Batteries should be removed when storing the unit for >3 months.
    Battery Check:
    Model 131 …LED will glow when the power switch is held in the
    BATTERY CHECK position
    Model 134 …LED will pulse once when the power switch is
    moved to the mA OUT position
    (Unless otherwise indicated, specifications are in 
    ±% of Span @ 23 °C)
    DIGITAL DIAL SETTING vs MILLIAMPS
    AVAILABLE FROM:
    MODEL Reading on Dial Clockwise  ➨
    MILLIAMPS
    © June 1999 Altek Industries Corp 
                       Model 46:100893, Model 134:100885-900 REV:ASpecifications subject to change without notice
    SPECIFICATIONS 
    						