    C H A P T E R  6  
    PrimeLog Operation  
    PrimeLog is used to program loggers and/or read data from loggers.  Th is 
    chapter  describes how to use PrimeLog.   
    Before using PrimeLog, the user m ust connect the logger to the serial port.  
    Select  the proper com munication (com)  port from the PrimeLog  tab on the 
    P ropert ies dialog .  Refer to  Port on page 19.   
    						
    							 ADS PrimeWorks User Manual  
    46 
     
    Program 
    The Program  option l aunches the  PrimeLog Program  dialog.  The PrimeLog 
    Program function enables the user to program the logger with  all the necessary  
    parameters.   
    Th is program is very flexible , allowing the user to load a data file, program a 
    logger, read a logger, change the setting s, save data to a file for  future logger 
    programming , and perform other essential tasks . 
     
    PrimeLog Program  window  
    PrimeLog Program Buttons  
    The buttons located at the bottom of the dialog perform the following functions:   
      New Config   Sets up a new configuration file  
     Load Config   Loads a saved configuration file  
     Save Config  Save s the configuration stored in memory to a file  
     Read Config  Read s the configuration of the logger  
     Program Now   Writes the program data to the logger   
    						
    							ADS PrimeWorks User Manual  
    47 
     New Config  
    Selecting the  New Config button displays the New Config  dialog with a 
    warning message indicating that creating a new config uration file eliminates 
    any changes made since saving the  co nfig uration  file.  To continue with the  
    new config uration file, click on the  Yes button; to abandon the new file , click 
    on the  No button. 
     
    New Config  dialog  
     Load Config 
    Selecting the Load Config button displays the Browse For Folder dialog  for 
    navigating to a previously saved logger configuration file . 
     
    Browse For Folder dialog  
    After locating and selecting the file , click on the OK button to read  the 
    configuration file into the program.   Configuration files have the following 
    filename format : [ site  name ].setup.xml . 
      Save Config 
    Selecting t his button sav es the configuration data  currently stored in memory  to 
    a file .  PrimeWorks  provides the file with the filename [ site name ].setup.xml  
    and save s the file to  a folder with  the same site  name.  The Save Config dialog  
    						
    							 ADS PrimeWorks User Manual  
    48 
     
    indicates the configuration has been successfully saved.  Click on the OK  
    button to return to the main program.  
     
    Save Config dialog  
     Read Config 
    Selecting t his button displays the  Read Configuration dialog to enable the user 
    to read  data from the log ger.  The logger must be  connected to the PC’s serial 
    port.   
     
    Read Configuration  dialog  
    Clicking OK  initiates the reading process and displays the  Read Configuration 
    window with a scrolling bar indicatin g the progress of the data reading process.    
     
    Rea d Configuration progress window  
    When reading is complete, the fields in the  PrimeLog Program dialog will be 
    populated with the data read from the logger.  
      Program Now  
    Selecting this button displays the  Program Logger dialog instructing the user 
    to verify th at the logger is connected to the PC serial port.    
    						
    							ADS PrimeWorks User Manual  
    49 
     
    Program Logger dialog  
    Once this connection is confirmed, click on the OK  button to program the 
    logger  with the current settings.  
    If the user attempts to program the logger before all the data is read ba ck from 
    the logger,  PrimeWorks  will display a warning.  
     
    Readback Error dialog  
    Clicking on the Yes  button will program the logger without reading back some 
    of the logged data stored in the logger.  This data will be lost.  Clicking on the 
    No  button will di scontinue logger programming.  
    Logger Clock 
    The Logger Clock  section on the Logger Information  tab enables the user to 
    configure the software to check the logger clock and synchronize it to the user’s 
    PC clock  when necessary or based on a user -defined diffe rence.   
       
    Logger Clock section on Logger Information   
    A warning displays when the software detects a  time difference between the PC 
    and the logger  clocks. 
    Channel Parameter s 
    Selecting t he Channel Information  tab displays the channel data.    
    						
    							 ADS PrimeWorks User Manual  
    50 
     
     
    PrimeLog Program dialog ( Channel Information  tab)  
    The Channel  drop-down list on the Channel Information  tab lists the 
    c hannels available on the selected logger.   Two channel types  are available: 
    Analog and Digital .  The selected channel type determines the categories of 
    logging that are available and the settings/parameters that display on the tab.    
    For example, a user can log data from  a bi-directional digital flow meter by 
    select ing the appropriate meter type from  the Digital 1a options.  A user can log 
    uni -directional flows using two meters  at independent locations in a single flow 
    channel  by selecting the meters separ ately from the Digital 1a and Digital 1b 
    options. 
      Digital Screen  Description 
    Four types of logging are available throu gh PrimeLog: 
      Flow Meter   This involves standard , periodic logging.  The other three 
    are event -driven, recording  the times at which pulses occur  rather than 
    the number of pulses.  
      Meter Type   Select the type of flow meter.  This is critical to 
    ensuring the proper powering of the meter.  
      FC al  Enter/select the amount per pulse , and then select the units 
    upon which the amount is based.  Refer to the Chapter 9, Appendi x 
    – Meter Scaling Tables , beginning on page 75, for the proper 
    amount to enter.   
    						
    							ADS PrimeWorks User Manual  
    51 
     Flow Total Index   Enter the reading from the meter at installation 
    so the total in PrimeLog  will be consistent with the meter.  
      Low Flow s  Select this checkbox if the meter will send less than 
    one pulse per minute.  
      Logging Functions   Modify the logging function selection to 
    setup the appropriate function.  Average & Total data can be logged 
    independently.  
      Show  Events   Select this checkbox to program the PrimeLog to 
    record/provide notification based on internal (threshold) events or to 
    trigger an output on the logger based on control output events. 
      Rain Gauge   This involves logging only when  a specific amount of  
    rainfall accumulates in a  rain gauge tipping bucket .  The user must 
    designate the amount of rainfall required to cause the tipping 
    mechanism  in the bucket  to  tip . 
      Pump  This involves logging only when  a user-defined high or low 
    indicates a p ump is on or off.  The user also must de signate the flow rate 
    at which the pump turns  on. 
      Overf low  This involves logging only when a user -defin ed high or low 
    indicates an overfl ow state. 
      Analog Screen Description 
    Select the type of sensor ( Transducer/Voltage/Current). 
      Transducer   Enter the PCal for  the transducer .  Select the A utozero 
    on Program  checkbox to remove the atmospheric offset when  
    programming the logger.  
     Voltage   Enter the minimum scale and maximum scale value s i n user 
    units  and the offset in volts  that the minimum scale represents.  
     Current   Enter the m inimum scale and full scale values in user units 
    and the offset in mA that  the minimum scale represents.  
    Note:  When using a nalog inputs on a PrimeLog , use the correct cable 
    to connect to the logger source.  Current and voltage cables cannot be 
    interchanged.  Improper  use could cause internal  damage to the logger. 
    It also  is important to use a voltage cable of the  appropriate rating.  Do 
    not apply a 10-volt source using a 2- volt or 5-volt  cable.  Th is will 
    overload the logger.  
    The user can select an application f or all analog sensors.  This will determine 
    what is logged by the logger.  When  using a depth transducer to measure open 
    channel flow, both the depth and the derived flow will be saved during  
    readback.   
    						
    							 ADS PrimeWorks User Manual  
    52 
     
    Readback 
    The Readback function enables the user to download  data from a logger into the 
    PrimeWorks  database.  
    Selecting the  Readback  option from the PrimeLog  menu displays the 
    PrimeLog R eadback dialog.  Empty fields indicate that no data is available 
    because readback has not yet occurred.  
     
    PrimeLog Readback dialog 
    The radio buttons on the dialog allow the user to designate whether to  readback 
    all or only a portion of the data from the logger:  
     Read All Data  Selecting this options reads back all data contained in 
    the logger.  If the logger configuration has just been read, the range of 
    data availabl e in the logger  will display in the start date/time and end 
    date/time fields adjacent to the Read Data Between  radio button. 
     Read Since Last   Selecting this option reads back all data since the 
    last readback for the selected logger.  
     Read Data Between   Selecting this option allows the user to 
    designate the date/time range for which to readback the data from the 
    logg er.  If the logger configuration has just been read, the range of data 
    available in the logger  will display in the corresponding start date/time 
    and end date/time fields.   Designate the start and end dates/times using  
    the data selection tool accessible by clicking on the down arrows 
    associated with each date/time field.  
      Current Values  
    Selecting this checkbox expands the dialog to reveal  additional data fields.  
    The se fields are filled with current values  from the data logger.  These values 
    will update every  10 seconds.  
    						
    							ADS PrimeWorks User Manual  
    53 
     
    Expanded PrimeLog Readback  dialog 
     Readback  
    C licking on this button displays  the Readback Confirmation  dialog instructing 
    the user to verify that the logger is connected to the PC serial port. 
     
    Readback Data  dialog 
    Once this connection is confirmed, click on the OK  button to initiate the 
    readback process from the logger.  
    A progress bar displays showing the status as  data is read from the logger.  
    Please note that large amounts of data may take  additional time to readback. 
    Once readback  is complete and the designated data is added to the database, the 
    data will display  in the database navigation pane .  
    						
    							 ADS PrimeWorks User Manual  
    54 
     
    Pressure AutoCal 
    The Pressure AutoCal  function is used to set  the local air pressure value as the 
    reference for  calibration.  By default, t he calibration pressure is set to standard 
    atmospheric pressure.  
    Note:  To  AutoCal a  logger, the pressure connection pipe for the 
    appropriate channel must be disconnected and the logger sensor must be 
    open to the ambient air pressure that wi ll be used as the calibration 
    value. 
    To run the AutoCal function, select the  Pressure AutoCal option from the 
    PrimeLog  menu.  PrimeWorks  opens the Read Configuration  dialog 
    instructing the user to verify that the logger is connected to the PC’s serial port.  
     
    Read Configuration dialog  
    Connect the logger to the serial port , and click on the OK button.  The progress 
    will display  in the lower right  section of the main window. 
    Once communication  is complete, the Pressure AutoC al dialog  will display . 
    Note:   Autocal is available only for loggers recording data from 
    pressure or depth transducers .  
    The dialog displays t he serial number, s ite name, and logger type.  S elect the 
    checkbox corresponding to  the Sensor Information channel on which to 
    perform the AutoCal  process. 
    Verify that  the logger pressure transducer pipe is not connected to the logger , 
    and then c lick on the AutoCal Now  button to initiate the AutoCal process.  
    Once the process is complete, a n AutoZero Setting dialog will display the 
    depth value representing  the calibration factor that has been measured and set.  
     
    AutoZero Setting dialog indicating a local zero value has been set  
    Note:  Remember to reconnect the logger pressure measuring pipe as 
    required.   
    						