ADS PrimeWorks User Manual 45 C H A P T E R 6 PrimeLog Operation PrimeLog is used to program loggers and/or read data from loggers. Th is chapter describes how to use PrimeLog. Before using PrimeLog, the user m ust connect the logger to the serial port. Select the proper com munication (com) port from the PrimeLog tab on the P ropert ies dialog . Refer to Port on page 19.
ADS PrimeWorks User Manual 46 Program The Program option l aunches the PrimeLog Program dialog. The PrimeLog Program function enables the user to program the logger with all the necessary parameters. Th is program is very flexible , allowing the user to load a data file, program a logger, read a logger, change the setting s, save data to a file for future logger programming , and perform other essential tasks . PrimeLog Program window PrimeLog Program Buttons The buttons located at the bottom of the dialog perform the following functions: New Config Sets up a new configuration file Load Config Loads a saved configuration file Save Config Save s the configuration stored in memory to a file Read Config Read s the configuration of the logger Program Now Writes the program data to the logger
ADS PrimeWorks User Manual 47 New Config Selecting the New Config button displays the New Config dialog with a warning message indicating that creating a new config uration file eliminates any changes made since saving the co nfig uration file. To continue with the new config uration file, click on the Yes button; to abandon the new file , click on the No button. New Config dialog Load Config Selecting the Load Config button displays the Browse For Folder dialog for navigating to a previously saved logger configuration file . Browse For Folder dialog After locating and selecting the file , click on the OK button to read the configuration file into the program. Configuration files have the following filename format : [ site name ].setup.xml . Save Config Selecting t his button sav es the configuration data currently stored in memory to a file . PrimeWorks provides the file with the filename [ site name ].setup.xml and save s the file to a folder with the same site name. The Save Config dialog
ADS PrimeWorks User Manual 48 indicates the configuration has been successfully saved. Click on the OK button to return to the main program. Save Config dialog Read Config Selecting t his button displays the Read Configuration dialog to enable the user to read data from the log ger. The logger must be connected to the PC’s serial port. Read Configuration dialog Clicking OK initiates the reading process and displays the Read Configuration window with a scrolling bar indicatin g the progress of the data reading process. Rea d Configuration progress window When reading is complete, the fields in the PrimeLog Program dialog will be populated with the data read from the logger. Program Now Selecting this button displays the Program Logger dialog instructing the user to verify th at the logger is connected to the PC serial port.
ADS PrimeWorks User Manual 49 Program Logger dialog Once this connection is confirmed, click on the OK button to program the logger with the current settings. If the user attempts to program the logger before all the data is read ba ck from the logger, PrimeWorks will display a warning. Readback Error dialog Clicking on the Yes button will program the logger without reading back some of the logged data stored in the logger. This data will be lost. Clicking on the No button will di scontinue logger programming. Logger Clock The Logger Clock section on the Logger Information tab enables the user to configure the software to check the logger clock and synchronize it to the user’s PC clock when necessary or based on a user -defined diffe rence. Logger Clock section on Logger Information A warning displays when the software detects a time difference between the PC and the logger clocks. Channel Parameter s Selecting t he Channel Information tab displays the channel data.
ADS PrimeWorks User Manual 50 PrimeLog Program dialog ( Channel Information tab) The Channel drop-down list on the Channel Information tab lists the c hannels available on the selected logger. Two channel types are available: Analog and Digital . The selected channel type determines the categories of logging that are available and the settings/parameters that display on the tab. For example, a user can log data from a bi-directional digital flow meter by select ing the appropriate meter type from the Digital 1a options. A user can log uni -directional flows using two meters at independent locations in a single flow channel by selecting the meters separ ately from the Digital 1a and Digital 1b options. Digital Screen Description Four types of logging are available throu gh PrimeLog: Flow Meter This involves standard , periodic logging. The other three are event -driven, recording the times at which pulses occur rather than the number of pulses. Meter Type Select the type of flow meter. This is critical to ensuring the proper powering of the meter. FC al Enter/select the amount per pulse , and then select the units upon which the amount is based. Refer to the Chapter 9, Appendi x – Meter Scaling Tables , beginning on page 75, for the proper amount to enter.
ADS PrimeWorks User Manual 51 Flow Total Index Enter the reading from the meter at installation so the total in PrimeLog will be consistent with the meter. Low Flow s Select this checkbox if the meter will send less than one pulse per minute. Logging Functions Modify the logging function selection to setup the appropriate function. Average & Total data can be logged independently. Show Events Select this checkbox to program the PrimeLog to record/provide notification based on internal (threshold) events or to trigger an output on the logger based on control output events. Rain Gauge This involves logging only when a specific amount of rainfall accumulates in a rain gauge tipping bucket . The user must designate the amount of rainfall required to cause the tipping mechanism in the bucket to tip . Pump This involves logging only when a user-defined high or low indicates a p ump is on or off. The user also must de signate the flow rate at which the pump turns on. Overf low This involves logging only when a user -defin ed high or low indicates an overfl ow state. Analog Screen Description Select the type of sensor ( Transducer/Voltage/Current). Transducer Enter the PCal for the transducer . Select the A utozero on Program checkbox to remove the atmospheric offset when programming the logger. Voltage Enter the minimum scale and maximum scale value s i n user units and the offset in volts that the minimum scale represents. Current Enter the m inimum scale and full scale values in user units and the offset in mA that the minimum scale represents. Note: When using a nalog inputs on a PrimeLog , use the correct cable to connect to the logger source. Current and voltage cables cannot be interchanged. Improper use could cause internal damage to the logger. It also is important to use a voltage cable of the appropriate rating. Do not apply a 10-volt source using a 2- volt or 5-volt cable. Th is will overload the logger. The user can select an application f or all analog sensors. This will determine what is logged by the logger. When using a depth transducer to measure open channel flow, both the depth and the derived flow will be saved during readback.
ADS PrimeWorks User Manual 52 Readback The Readback function enables the user to download data from a logger into the PrimeWorks database. Selecting the Readback option from the PrimeLog menu displays the PrimeLog R eadback dialog. Empty fields indicate that no data is available because readback has not yet occurred. PrimeLog Readback dialog The radio buttons on the dialog allow the user to designate whether to readback all or only a portion of the data from the logger: Read All Data Selecting this options reads back all data contained in the logger. If the logger configuration has just been read, the range of data availabl e in the logger will display in the start date/time and end date/time fields adjacent to the Read Data Between radio button. Read Since Last Selecting this option reads back all data since the last readback for the selected logger. Read Data Between Selecting this option allows the user to designate the date/time range for which to readback the data from the logg er. If the logger configuration has just been read, the range of data available in the logger will display in the corresponding start date/time and end date/time fields. Designate the start and end dates/times using the data selection tool accessible by clicking on the down arrows associated with each date/time field. Current Values Selecting this checkbox expands the dialog to reveal additional data fields. The se fields are filled with current values from the data logger. These values will update every 10 seconds.
ADS PrimeWorks User Manual 53 Expanded PrimeLog Readback dialog Readback C licking on this button displays the Readback Confirmation dialog instructing the user to verify that the logger is connected to the PC serial port. Readback Data dialog Once this connection is confirmed, click on the OK button to initiate the readback process from the logger. A progress bar displays showing the status as data is read from the logger. Please note that large amounts of data may take additional time to readback. Once readback is complete and the designated data is added to the database, the data will display in the database navigation pane .
ADS PrimeWorks User Manual 54 Pressure AutoCal The Pressure AutoCal function is used to set the local air pressure value as the reference for calibration. By default, t he calibration pressure is set to standard atmospheric pressure. Note: To AutoCal a logger, the pressure connection pipe for the appropriate channel must be disconnected and the logger sensor must be open to the ambient air pressure that wi ll be used as the calibration value. To run the AutoCal function, select the Pressure AutoCal option from the PrimeLog menu. PrimeWorks opens the Read Configuration dialog instructing the user to verify that the logger is connected to the PC’s serial port. Read Configuration dialog Connect the logger to the serial port , and click on the OK button. The progress will display in the lower right section of the main window. Once communication is complete, the Pressure AutoC al dialog will display . Note: Autocal is available only for loggers recording data from pressure or depth transducers . The dialog displays t he serial number, s ite name, and logger type. S elect the checkbox corresponding to the Sensor Information channel on which to perform the AutoCal process. Verify that the logger pressure transducer pipe is not connected to the logger , and then c lick on the AutoCal Now button to initiate the AutoCal process. Once the process is complete, a n AutoZero Setting dialog will display the depth value representing the calibration factor that has been measured and set. AutoZero Setting dialog indicating a local zero value has been set Note: Remember to reconnect the logger pressure measuring pipe as required.