Page 1 Adam Equipment MDW – 300L HEALTH AND FITNESS SCALE with BMI ADAM EQUIPMENT CO. LTD. p.n. 700660159 Rev. A2, Apr 2014
Page 2
Page 2 CONTENTS 1.0 INTRODUCTION ...................................... ..................................................... 3 2.0 TECHNICAL SPECIFICATIONS .......................... ......................................... 3 3.0 DISPLAY AND KEY DESCRIPTIONS ...................... .................................... 4 4.0 SETTING UP THE SCALE ............................................................................ 5 5.0 FUNCTION...
Page 3
Page 3 1.0 INTRODUCTION · The MDW 300L scale is a Digital Health and Fitness Scale with Body Mass Index (BMI) readout. · It is an accurate electronic device with advanced design and stable performance. · It is designed to measure both the weight and the height of a person and compute the BMI index. 2.0 TECHNICAL SPECIFICATIONS Maximum capacity 300 kg / 660lb Minimum capacity 2 kg / 4lb Scale division 0.05kg / 0.1lb Height range 60 cm-212 cm / 24”-83”...
Page 4
Page 4 3.0 DISPLAY AND KEY DESCRIPTIONS KEYS FUNCTIONS [On/Off / Zero] To turn the scale on or off. To zero the scale if the display drifts from zero. [Tare] ¿ ¿¿ ¿ To tare the scale, if necessary. Accept settings as required [Units / BMI] To toggle the weighing unit between Kg and Lb. and to select the BMI function. Move active digit to the right when req uired. [Print / Hold] To lock the reading even if the person to be weighe d is moving and also to print...
Page 5
Page 5 4.0 SETTING UP THE SCALE · Take the scale out of the box. 1. Locate the 4 screws (2) 2. Place the cable (4) through the pillar (3). 3. Insert the 4 screws (2) through the holes on the pillar bracket and secure the bracket to the indicator (1), 4. Make sure the screws are securely tightened. 5. Place the cable (4) into the tube support (6) on the scale base. 6. Insert the pillar (3) into the tube support (6). 7. Rotate the pillar to position the...
Page 6
Page 6 9. Place the indicator cable (4) and the load cell cable (13) together near the hole on the base (16). 10. Plug the two connectors (15) together. 11. Position the connector inside the base (16). 12. The cables should be fixed to the base using the clip provided (14). 13. Locate the fixing bolt and washer (10) used to secure the height rod to the scale base (11) 14. Locate the height rod (7) onto the fixing points (9) of the upper and lower brackets (8) & (12)....
Page 7
Page 7 5.0 FUNCTION 5.1 WEIGHING · Place the Scale on an even floor and press the [On/Off] key. · The instrument performs a self-test after which it is ready for operation. · Press the [On/Off] key and the machine switches off. · The person to be weighed can step on to the platfo rm once the scale shows 0.0 on the display. The weight will be display in K g. or lb. depending on the units chosen by the user. · If the weighing value is to be...
Page 8
Page 8 5.2 MEASURING HEIGHT · To measure someones height, it is necessary to ext ract the measuring rod and place the height arm on top of the head of the pers on being measured. · For shorter heights the height arm can be released by pressing the red button on the top of the height rod, and then moving the h eight arm to the lower section of the height pillar. · The height is read on the bar at the top of the su pport. The height can be read in either centimetres or...
Page 9
Page 9 6.0 CALIBRATION Before calibrating the scale, you should ensure that you have a suitable known weight for calibration. 1. When in normal weighing mode with the scale at zero press and hold down [TARE] and [ON / OFF] keys to enter the calibration mode. 2. If the calibration switch is in the off position on the main PCB inside the scale, the indicator will show “ CAL.OFF” and then exit this mode. If the indicator shows “ CAL-? ”, the scale is...
Page 10
Page 10 7.0 USER PARAMETERS This indicator has 4 parameter settings that can be selected. 1. When the scale is in normal weighing mode, press an d hold down the [ON / OFF] key and the [UNIT] key for 3 seconds until ‘ Setup’ is shown on the display. 2. When in the SETUP mode, press the [Print/Hold] key to change the flashing digits, and [TARE] key to confirm the flashing digits and move to the next parameter setting. Press the [ON / OFF] key to exit the set up mode....