Adam Equipment 
     
 
 
 
MDW – 300L
 
HEALTH AND FITNESS SCALE with BMI 
 
 
 
 
 
 
   
ADAM EQUIPMENT CO. LTD.   
p.n. 700660159 Rev. A2, Apr 2014 
      

Page 2

Page 2 
   
CONTENTS    
1.0 INTRODUCTION ...................................... ..................................................... 3 
2.0 TECHNICAL SPECIFICATIONS .......................... ......................................... 3 
3.0 DISPLAY AND KEY DESCRIPTIONS ...................... .................................... 4 
4.0 SETTING UP THE SCALE ............................................................................ 5 
5.0 FUNCTION...

Page 3

Page 3 
1.0  INTRODUCTION 
   
·   The MDW 300L scale is a Digital Health and Fitness  Scale with Body Mass 
Index (BMI) readout. 
  
·  It is an accurate electronic device with advanced  design and stable 
performance. 
 
·  It is designed to measure both the weight and the  height of a person and 
compute the BMI index.  
     
2.0  TECHNICAL SPECIFICATIONS 
    
Maximum capacity  300 kg / 660lb 
Minimum capacity  2 kg / 4lb 
Scale division  0.05kg / 0.1lb 
Height range  60 cm-212 cm / 24”-83”...

Page 4

Page 4 
3.0  DISPLAY AND KEY DESCRIPTIONS 
  
   
KEYS FUNCTIONS 
[On/Off / 
Zero]  To turn the scale on or off.  
To zero the scale if the display drifts from zero. 
[Tare] 
¿
¿¿
¿
  To tare the scale, if necessary. Accept settings as
 required 
[Units / 
BMI] 

  To toggle the weighing unit between Kg and Lb. and 
to select the BMI 
function.   Move active digit to the right when req uired. 
[Print / 
Hold] 

  To lock the reading even if the person to be weighe
d is moving and 
also to print...

Page 5

Page 5 
4.0 SETTING UP THE SCALE  
   
·  Take the scale out of the box.  
 
 
  
 
  
1. Locate the 4 screws (2)    
2. Place the cable (4) through the pillar 
(3).  
3. Insert the 4 screws (2) through the 
holes on the pillar bracket and secure 
the bracket to the indicator (1),   
4. Make sure the screws are securely 
tightened. 
     
5. Place the cable (4) into the tube 
support (6) on the scale base.  
6. Insert the pillar (3) into the tube 
support (6).  
7. Rotate the pillar to position the...

Page 6

Page 6 
 
   
 
 
  
9. Place the indicator cable (4) and the  
load cell cable (13) together near the hole 
on the base (16).  
10. Plug the two connectors (15) together.   
11. Position the connector inside the base 
(16).  
12. The cables should be fixed to the base 
using the clip provided (14).  
13. Locate the fixing bolt and washer 
(10) used to secure the height rod to 
the scale base (11)  
14. Locate the height rod (7) onto the 
fixing points (9) of the upper and lower 
brackets (8) & (12)....

Page 7

Page 7 
5.0  FUNCTION 
 
5.1  WEIGHING 
 ·  Place the Scale on an even floor and press the  [On/Off] key.  
 
·  The instrument performs a self-test after which it  is ready for operation.  
 
·  Press the  [On/Off] key and the machine switches off.  
 
·  The  person to be  weighed  can  step  on  to  the  platfo rm once  the  scale  shows 
0.0  on  the  display.  The  weight  will  be  display  in  K g.  or  lb.  depending  on  the 
units chosen by the user.  
 
·  If  the  weighing  value  is  to  be...

Page 8

Page 8 
5.2  MEASURING HEIGHT  
 ·  To measure someones height, it is necessary to ext ract the measuring rod and 
place the height arm on top of the head of the pers on being measured.   
 
·  For shorter heights the height arm can be released  by pressing the red button 
on  the  top  of  the  height  rod,  and  then  moving  the  h eight  arm  to  the  lower 
section of the height pillar.  
 
·  The height is read on the bar at the top of the su pport.  The height can be read 
in either centimetres or...

Page 9

Page 9 
6.0  CALIBRATION 
  
Before calibrating the scale, you should ensure that you have a suitable known weight for 
calibration.    
1.  When in normal weighing mode with the scale at zero  press and hold down [TARE] and 
[ON / OFF]  keys to enter the calibration mode.  
  
2.  If  the  calibration  switch  is  in  the  off  position  on   the  main  PCB  inside  the  scale,  the 
indicator  will  show  “ CAL.OFF”  and  then  exit  this  mode.  If  the  indicator  shows  “ CAL-? ”, 
the scale is...

Page 10

Page 10 
7.0  USER PARAMETERS 
  
This indicator has 4 parameter settings that can be selected. 
 
1. When  the  scale  is  in  normal  weighing  mode, press  an d  hold  down  the [ON  /  OFF] key 
and the  [UNIT] key for 3 seconds until ‘ Setup’ is shown on the display.  
 
2. When in the SETUP mode, press the  [Print/Hold] key to change the flashing digits, and  
[TARE]  key to confirm the flashing digits and move to the  next parameter setting. Press 
the  [ON / OFF]  key to exit the set up mode....