Have a look at the manual Adam Equipment GBC Bench Counting Scales GBCGFCREVB4 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 35 Adam Equipment manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
P a g e 19 © Adam Equipment Company 2014 Data Format-Normal Output: Data Format- Memory Recall Print: DATE 12/09/2006 TIME 14:56 GROSS Wt 1.234 Kg Net Wt. if net weight is displayed Unit Wt. 123 g Kg or g for metric weights and lb for pounds. Pcs 10 pcs Includes 2 line feeds with carriage return at beginning and end of the form DATE 12/09/2006 TIME 14:56 ------------------ TOTAL No. 5 Wt. 1.234 Kg Pcs 10 pcs Includes 1 line feed ------------------ 2 line feeds, carriage return
P a g e 20 © Adam Equipment Company 2014 Data Format- Continuous Print: In other languages the format is the same but the t ext will be in the language selected. See Section 9.2 Description ENGLISH FRENCH GERMAN SPANISH Print gross weight Gross Wt Pds Brut Brut-Gew Pso Brut Net weight Net Wt. Pds Net Net-Gew Pso Net Weight per unit counted Unit Wt. Pds unit Gew/Einh Pso/Unid Number of items counted Pcs Pcs Stck. Piezas Number of weighings added to subtotals No. Nb. Anzhl Num. Total weight and count printed Total Total Gesamt Total Print date Date Date Datum Fecha Print time Time Heure Zeit Hora ST,GS, 0.9080 kg U.W. 0.90798 g/pcs PCS 1000 pcs Includes 2 line feeds with carriage return between sets of data
P a g e 21 © Adam Equipment Company 2014 9.1 INPUT COMMANDS FORMAT The scale can be controlled with the following comm ands. The commands must be sent in upper case letters, i.e. “T” not “t”. Press the Enter key of the PC after each command. T Tares the scale to display the net weight. This is the same as pressing [Tare] key. Z Sets the zero point for all subsequent weighing. Th e display shows zero. Same as pressing the [Zero] key. P Prints the weight, unit weight and totals same as p ressing the [Print] key. R Recall and Print, Same as if first the [MR] key and then the [Print] key is pressed. Will display the current accumulate d memory and print the total results. C Same as pressing [MR] first and then the [CE] key to erase the current memory.
P a g e 22 © Adam Equipment Company 2014 9.2 RS-232 SETUP The RS-232 interface uses parameters set by the use r for language, baud rate and date format. Press and hold the [Print] key for 4 seconds to access the parameters. Press [U. Wt./Units] to scroll through the options and [Tare] to confirm the change and then advance to the next parameter. When a parameter is entered by pressing [Tare], the displays will guide you through the parameter selected and the options available. The parameters and their functions are: Displays Options Functions Weight Unit Weight Count Port on On oFF Enable or disable the RS- 232 interface. 9600 bPS 600 1200 2400 4800 9600 19200 Set baud rate. Print mAn Cont to PC Print Auto Print mAn Select printing options for continuous printing, print automatically or printing manually. AC mAn AC Auto AC mAn AC oFF Select the operation of accumulating automatically, manually or turned off. LAnG EnGLiS EnGLiS FrEnCH GErMAn SPAniS Select Language for printing.
P a g e 23 © Adam Equipment Company 2014 The scale will perform the following functions depe nding on the Accumulation and Print settings: ACCUMULATION FUNCTIONS PRINT FUNCTIONS AC Auto AC mAn AC oFF Print Auto Accumulate and print automatically Print automatically ; Accumulate and print when [M+] is pressed Print automatically, [M+] key has no function Print mAn Automatically Accumulate but not print, Print only when [Print] key pressed Accumulate and print when [M+] or [Print] is pressed Print when [Print] key is pressed, [M+] key has no function Cont to PC Print continuously and accumulate automatically when stable , [Print] key no function Print continuously and accumulate when [M+] is pressed, [Print] key no function Print continuously , [M+] and [Print] key have no function
P a g e 24 © Adam Equipment Company 2014 9.3 REAL TIME CLOCK SETUP The Real Time Clock (RTC) is used only for the RS-2 32 output. The Date and Time can be set as required. The scale will keep the clock running even when the power is off. Setting up the clock · Press and hold the [CE] key for 4 seconds, release when the date and time is displayed. The initial displays show the current d ate and time set. “ rtC “ “11,14,06” “16,41,35” · Press the [CE] key to change the date and time. The display will show the current time in the format, “H-m-S”. “timE” “H-m-S” “142929” or if no time has been set show “000000” · Enter the time using the numeric keys using a 24 ho ur clock format, 3:41PM is “154100”. “timE” “H-m-S” “154100” new time overwrites the old · Press the [Tare] key to accept the time. The display will show the current date format. “dAtE” “y-m-d” · Press the [ U.Wt./Unit] key to change the date format. Available formats a re: “Y-m-d” year, month, day “m-d-Y” month, day, year “d-m-Y” day, month, year · Press the [Tare] key to accept the chosen format and then enter the date in this format. “dAtE” “m-d-y” “051209” note current date is in current format, will show 000000 if no date has been set. · Press the [Tare] key to accept the date. An error code will be shown if the time (Err 1) or the date (Err 2) is not the permissible values. For example, 34th day of a month is an in valid entry.
P a g e 25 © Adam Equipment Company 2014 9.4 AUTO SLEEP FUNCTION This function may be enabled or disabled by the use r. If enabled, when the scale is not used for some time (as pre-set by the user unde r this function) it automatically switches off. To set this parameter- · Press and hold the [Zero] key for 4 seconds, release when the display shows “SLEEP nodE” . · Press [U/Wt/Unit] key to scroll through the auto sleep values. “0” Auto sleep mode disabled “1” Auto sleep after 1 minute “5” Auto sleep after 5 minutes “10” Auto sleep after 10 minutes · Press [Tare] to set the value. The scale returns to zero. 10.0 BATTERY AND BACKLIGHT OPERATION 10.1 BATTERY · The scales can be operated from the battery, if des ired. The battery life is approximately 70 hours. · When the battery needs charging the battery symbol will appear to be empty. The battery should be charged as possible. Once the “LO bAT” message is shown the scale will still operate for a bout 10 minutes after which it will automatically switch off to protect t he battery. · To charge the battery, simply apply power to the sc ale. The scale does not need to be turned on. · The battery should be charged for at least 12 hours for full capacity. · Just above the “Unit Weight” display is an LED to indicate the status of battery charging. When the scale is plugged into t he mains power, the internal battery will be charged. If the LED is gr een the battery is fully charged. If it is red, the battery is nearly disch arged and yellow indicates the battery should be charged longer, preferably ov ernight. · If the battery has not been used properly or it is used for a number of years it may eventually fail to hold a full charge. If the battery life becomes unacceptable then contact your supplier.
P a g e 26 © Adam Equipment Company 2014 10.2 BACKLIGHT FOR LCD · The backlight of the LCD can be set to be- “1” : ON at all the time, “2” : ON only when a weight is placed on the scale or “3” : Turned off. · To set the backlight press and hold [Pst] key for 4 seconds. · The weight display will show “EL xx” where xx is the current setting for the backlight. Press [U. Wt./Units] to scroll through the options. “EL on” Sets the backlight to be on at all times. “EL Au” Sets the backlight to operate automatically when a weight is placed on the scale or a key is pressed. “EL OFF” Sets the backlight to be off. Press the [Tare] key to store the value or press the [Zero] key to escape from this setting and return to weighing.
P a g e 27 © Adam Equipment Company 2014 11.0 ERROR CODES During the initial power-on testing or during opera tion, the scale may show an error message. The meaning of the error messages is desc ribed below. If an error message is shown, repeat the step that caused the message, turning the balance on, carry out the calibration or other func tions. If the error message is still shown contact your dealer for further support. ERROR CODE DESCRIPTION POSSIBLE CAUSES Err 1 Time input error. Tried to set an illegal time, i.e . 26hours Err 2 Date input error Tried to set an illegal date, i.e. 36th day Err 4 Initial Zero is greater than allowed (typically 4% of the maximum capacity) when power is turned on or when the [Zero] key is pressed, Weight is on the pan when turning the scale on. Excessive weight on the pan when zeroing the scale. Improper calibration of the scale. Damaged load cell. Damaged Electronics. Err 6 A/D count is not correct when turning the scale on. Platform is not installed. Damaged Load cell. Damaged Electronics. diSPEr Shows when a number is too large tot fit the display The accumulated totals for weight or count exceed 999999. The data on RS232 is still shown accurately
P a g e 28 © Adam Equipment Company 2014 12.0 SERVICE PARAMETERS 12.1 USING “0000” TO ENTER THE CALIBRATION PARAMETER · Press the [Tare] key during the display counting when turned on. Th e display will ask for a code number, “Pin “ on the Weight Display. · Enter the number 0000 when “Pin “ is displayed and then press [Tare]. If another PIN number has been previously stored en ter that number instead. · The displays will show the first parameters, called “F1” “CAL” . · To select another parameter press the [U.Wt./Units] key to advance through the parameters. · Press the [Tare] key to enter a parameter. · To exit a parameter, press the [Zero] key. · The Weight window will show the parameter number an d the Unit Price and Count windows will show the word describing the function. · When a parameter is entered by pressing the [Tare] key, the displays will guide you through the parameter selected and t he options available. The parameters available are: “F1” “CAL” To enter the Calibration “F2” “Pin” To set a new Pin number for access “F3” “Unit” Enable Weighing Units