    							4 1  • Print switch mode / Command mode Function setting “ prt 2 ” 
     When the 
     PRINT 
      switch is pressed while the weight value is stable (STABLE indicator is on), the 
    scale transmits the value. 
    • Auto-print mode +/- data / Command mode Function setting “ prt 3 ” 
      The scale transmits the weight value when the display is stable (STABLE indicator is on) and the value 
    is more than +4d or less than -4d. The next output can be obtained after the value returns to be more 
    than -4d and less than +4d. 
    d = minimum display (Refer to “Specifications” of the scale instruction manual.) 
    Even the counting mode uses “d” for judgment. 
    • Auto-print mode + data / Command mode Function setting “ prt 4 ” 
      The scale transmits the weight value when the display is stable (STABLE indicator is on) and the value 
    is more than +4d. The next output can be obtained after the value returns to be less than +4d. 
    d = minimum display (Refer to “Specifications” of the scale instruction manual.) 
    Even the counting mode uses “d” for judgment. 
     
     
    3.4. Command mode 
    In the command mode, the scale is controlled by commands that come from the external device such 
    as a personal computer. 
     Command list 
    • “ Q ” command  Command to request the current weight value. 
    Command Q C
    R L
    F 
     
    Reply S T  ,  + 0 0 1 2 3
     
    . 
    4 
    5 
    _ 
    k 
    g C
    R LF 
     
    • “ Z ” command  Same operation as the 
     ZERO 
     switch. 
    Command Z C
    R L
    F 
     
    • “ T ” command  Same operation as the 
     TA R E  
     switch. 
    Command T C
    R L
    F 
     
     
    Reply to the command 
    •  When the command cannot be executed, for example, because the scale is unstable, “ I ” will be sent. 
    Reply I C
    R L
    F 
     
    •  If the received command is not for the SC/SE series scale, the scale will send “ ? ”. 
    Reply ? C
    R L
    F 
     
    • When “aCk 0” is selected, there is no reply except the “ Q ” command. 
     
    SCE-03 RS-232C Serial Interface and   
    Comparator Relay Output 
    Instruction Manual 
    1WMPD4002182A  
     1. Features 
    •  This interface allows the scale to be connected to a printer or a personal computer. •  It also allows outputting the comparison results of 
     HI
     ,  OK
      or 
     LO
       obtained using the comparator 
    function as a relay signal. 
    Note: SCE-02 and SCE-03 can not be installed at the same time. 
     
     
    2. Installation Procedure Note:  Before installing the SCE-03 board, be sure to remove 
    the batteries from the scale. Step 1
      Pass the cable to connect an external device through the hole 
    on the casing and connect it to the terminal block on the 
    interface board. 
    Step 2
      Insert the interface board into the casing. 
    Step 3
      Place the waterproof packing on the casing, aligning its holes 
    with the casing screw holes. 
    Step 4
      Remove the four screws on the option slot panel located on 
    the bottom of the display. 
    Step 5
      Insert the interface board into the option slot. 
    Step 6
      Secure the casing to the display, using the four screws 
    provided with the casing. 
    Step 7
      Seal the screw heads with the tabs attached to the waterproof 
    packing. 
    • Cable connection using optional cables (sold separately)  
      When connecting to an external device with hardware flow 
    control, communication will be impossible using a cable 
    without RTS and CTS connected. In that case, connect RTS 
    and CTS. This will disable hardware flow control, but enable 
    communication. 
    •  When the AX-KO3285-320 cable is used, RTS and CTS are 
    internally connected and the above operation id unnecessary. 
    •  When the connector of an external device is a D-sub 9-pin 
    connector, pin 7 is RTS and pin 8 is CTS. 
    AX-KO3285-320 
     
     
     
     
      3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo  170-0013  JAPAN
    Telephone: [81] (3) 5391-6132  Fax: [81] (3) 5391-6148
     
    Cable (AWG 26~16) to 
    connect an external device  Waterproof packing Seal the screw heads  Casing 
    Screws on the 
    option slot 
    6 mm (AWG 26~16) Terminal block 
    FG
    SG
    DSR
    TXD
    RXD
    COM
    LO
    O
    KHIThe hole for 
    the cable 
    Interface board
    Cable (AWG 26~16) to 
    connect an external device
    The hole for 
    the cable 
    1. Brown (DCD)
    2. Red  (RXD) 
    3. Black  (TXD) 
    4. Yellow (DTR) 
    5. Green  (SG) 
    6. Blue  (DSR) 
     Shield RXD 
    TXD 
    DSR 
    SG 
    FG
     
    SCE-03 
    Terminal block J1
     
    Optional cable unit to 
    connect with a PC (3 m)
      
    						
    							2 3 
    5 Frame ground  FG 
    4 Signal ground  SG 
    3  Data set ready  DSR 
    2 Transmit data  TXD 
    1 Receive data  RXD  Terminal block J1 
    4 Relay output common  COM 
    3 Relay output  LO 
    2 Relay output   OK 
    1 Relay output   HI 
    Terminal block J2   
    D-Sub 9 Pin Connector 
    D-Sub 9 pin Connector  Connections  SEC-03, Terminal block J1
    Pin  Signal name    Pin Signal name 1 DCD       2 RxD    2 TxD 3 TxD    1 RxD 4 DTR       5 SG    4 SG 6 DSR    3 DSR 7 RTS     8 CTS Shorted  
      9 -    Blank heat shrink tubing covered wire    5 FG 
     
     
    AX-KO3341-320 
     
     
     
     
     
     
     
     
     
     3. SCE-03 RS-232C Serial Interface 
    •  The RS-232C serial interface has four output modes; Stream mode, Command mode, Print switch 
    mode and Auto-print mode. For details, refer to “3.3. Data output mode”. 
    •  Set the parameters of the functions “ bp5 ”, “ prt ” and “ aCk ” as necessary. 
     
    3.1. Interface specifications 
    •  Data format  Baud rate:  2400, 4800, 9600 bps 
        Data bits:  7 bits + parity 1 bit (even) 
       Start bit:  1 bit 
        Stop bit:  1 bit 
       Code:  ACII 
       Terminator:  C
    R LF  (C
    R: 0Dh, L
    F: 0Ah) 
     
      
    LSB 0  
    1  
    2  
    3  
    4  
    5
    MSB6    
     
     
     
     
     
    •  Relay output maximum rating 
       Maximum voltage:  50V DC 
        Maximum current:  100mA DC 
       Maximum ON resistance: 8Ω  
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
    3.2. Data format 
    S 
    T 
    , 
    + 
    0 
    0 
    0 
    0 
    0 
    . 
    0 
    0 
    _ k g C
    R L
    F 
     
     
    •  Four types of headers are available: 
    ST  : Stable weight data US  : Unstable weight data (including counting data) 
    QT  : Stable counting data OL  : Out of weighing range 
    •  The data is always 9 digits including a sign and a decimal point. 
    •  Four units are available: 
    _kg  : Weight data in “kilograms” (kg) _lb  : Weight data in “decimal pounds” (lb) 
    _PC  : Counting data in “pieces” (pcs) _oz  : Weight data in “decimal ounces” (oz) 
    •  The terminator is always C
    R LF. 
    •  Example of output data: 
     
    Weight data “kg” S
     
    T 
    , 
    + 
    0 0 1 2 3  .  4 5 _ k g C
    R L
    F 
     
    Counting data “pcs” Q
     
    T 
    , 
    + 
    0 0 0 1 2 3 4 5 _ P C C
    R L
    F 
     
    Out of weighing range “kg” (+) O
     
    L 
    , 
    + 
    9 9 9 9 9  .  9 9 _ k g C
    R L
    F 
     
    Out of weighing range “pcs” (-) O
     
    L 
    , 
    -9 9 9 9 9 9 9 9 _ P C C
    R L
    F 
     
     3.3. Data output mode 
    • Stream mode / Command mode Function setting “ prt 0 ” 
      The scale outputs the current display data in sync with the display update of approximately 10 times per 
    second. The scale does not output data while it is in the setting mode. 
    • Command mode only Function setting “ prt 1 ” 
      The scale replies only to a command. For details, refer to “3.4. Command mode”.
     
    1 (-15V~-5V) 
    0 (5V~15V)   
    Stop bi
    t 
    Parity bit 
    Data bits 
    Start bit 
    HeaderData 
    Unit 
    Terminator 
    Separator Note:  “_” indicates a space.
     
    1.Brown 
    2.Red 
    3.Orange 
    4.Yellow 
    5.Green 
    6.Blue 
    7.Purple 
    8.Gray 
    9.White 
    10.Black RXD
    TXD 
    DSR 
    SG 
    FG
     
    SCE-03
    Terminal block J1
     
    Optional cable unit   
    to connect with   
    AD-8121B printer (3 m)
      
    						