This is a hazard mark (caution mark). © 2006 No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of A&D Company Ltd. The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice.
Page 3
1 Warning Definitions The warnings described in this manual have the following meanings: Caution A potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. Note An important description for proper use of the instrument. Precautions When handling the balance with the rechargeable battery unit installed, observe the following precautions. Caution Do not place the unit in water, heat it, or...
Page 4
2 Introduction The rechargeable battery unit (FXi-09, HRA-09) is exclusively designed for A&D-manufactured balances (of FX-i series, HR-A series). Using the battery unit, a balance can be operated in a location where an AC power source is not available. The battery unit consists of a battery panel and a built-in rechargeable battery that have been installed inside the balance prior to delivery. To install the battery unit into an balance that was not equipped with this option, contact...
Page 5
3 *1: Operation time of the battery depends on operating conditions. *2: The FXi-02, FXi-08 and FXi-09 can not be used at the same time. In addition, the HRA-02, HRA-08 and HRA-09 can not be used at the same time. *3: The battery unit is neither dust-tight nor waterproof. Therefore, upon installing the battery unit onto the FX-iWP and FZ-iWP series dust-tight and waterproof series, the balances will no longer comply with dust-tight and waterproof specifications. Notes For The Battery Unit...
Page 6
4 Charging the Battery Note Charging the battery before it is completely discharged will shorten the battery life. Charge the battery fully until the CHARGE LED indicator goes out. This procedure will take about 10 hours. During the charging operation, the balance can not be used. Procedure: 1. Make sure that the battery is completely discharged before recharging. How to determine when the battery is completely discharged: When the power switch on the rear of the balance is...
Page 7
5 Using the Balance Powered by the Battery Use the battery to supply power to the balance as follows: Procedure: 1. After charging is complete, disconnect the AC adapter from the balance. 2. Turn the power switch on the rear of the balance ON. The battery will start supplying power to the balance. 3. When the balance is not to be used, turn the power switch on the rear of the balance OFF. Under this condition, the battery is not used. Note When the power switch is turned...
Page 8
6 Using the Balance Powered by the AC Adapter Use the AC adapter to supply power to the balance as follows: The balance can be used powered by the AC adapter even if the battery is depleted. While the balance is used powered by the AC adapter, the battery is not used and charging is not performed. Procedure: 1. Turn the power switch on the rear of the balance ON. Note Be sure to turn the power switch ON. If the power switch is turned OFF and the AC adapter is connected to the...