Home > A&D > Balance > A&D Balance FS HC02iFS03iFS04i User Manual

A&D Balance FS HC02iFS03iFS04i User Manual

Here you can view all the pages of manual A&D Balance FS HC02iFS03iFS04i User Manual. The A&D manuals for Balance are available online for free. You can easily download all the documents as PDF.

Page 11

- 10 - 
Reply O K
, + 0 0 1000C
LF  
FS-15Ki : Target weight 1.000 kg will be set. 
 
‰ Set a HI limit value or upper limit weight.    (No reply for F20-1.) 
F07-0 or F07-1 is set, the command should have “+” and 6 digit number without 
decimal point. 
Command H I
, + 0 0 0200C
LF  Decimal point follows the weight display. 
Reply H I
, + 0 0 0200C
LF 
FS-15Ki : 
F07-0: 0.200 kg will be set as an upper limit weight. 
F07-1: 0.200 kg will be set as a HI limit weight. 
 
F07-2 is set, the command should...

Page 12

- 11 - 
Command M L
, 0 1 
, +001000
,+ 0 0 0 2 0 0 
, 
  + 0 0 0 1 0 0C
LF      
Reply   M  L
, 0 1 
, +001000
,+ 0 0 0 2 0 0 
, 
  + 0 0 0 1 0 0C
LF     
FS-15Ki : Store target weight 1.000 kg, upper limit weight +0.200 kg and lower limit 
weight +0.100 kg into the memory number “01”. 
 
F07-2 is set, the command should have 2 digit memory number, target weight that 
has a polarity and 6 digit number without decimal point, and HI/LO limit % that has 
“+” and 5 digit number without decimal point....

Page 13

- 12 - 
‰ The maximum number of text data is 300 characters. 
 
‰ To use the UFC function, the text data has to be sent to the scale from the 
computer using “PF” command in the command mode (F06-1) in advance. 
Then, connect the scale with the printer and set the F06-2, F06-3 or F06-4. 
‰ When the 
 PRINT   key is pressed or by auto-print mode (F06-2, F06-3 or F06-4), 
the scale will send the stored text data with the parameters replaced by the original 
data. 
 
Store Text Data into the Scale Memory...

Page 14

- 13 - 
If you will send more than 2 lines of data, attach “&” to the end of the first line. Then, 
the scale considers the data to be continued. 
‰ A “Space” or “,” will be used to separate these data. You can skip them to 
reduce the number of characters, but you cannot skip “,” after “PF”. You 
must start with “PF,”. 
 
Data Format for the Scale Data    (“_” shows “Space” (20H).) 
Parameters for the scale data will be replaced by the format below when the scale 
sends them out. 
‰ Data has a fixed...

Page 15

- 14 - 
 
3. OP-04 RS-422 / 485 / RELAY OUTPUT 
 
This interface allows a personal computer to connect and control up to 16 FS-i scales, 
and the relay outputs for comparator result are obtained.   
‰ The OP-04 unit includes an interface board, a connector cable, two cable glands 
and two screws (M3x6). 
‰ The relay output specifications are same as the OP-03. See “2-2. OP-03 
Specifications”. 
 
3-1. Installation 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Ten hexagon...

Page 16

- 15 - 
 
‰ Installation is similar to the OP-03. See “2-1. Installation”. 
‰ Set the function parameters F04, F05, F06, F18, F19 and F20 according your 
application. 
‰ The function setting F19-1 must be set for RS-422 and F19-2 must be RS-485. To 
connect more than one scale with a computer, set a different address to each scale 
by F18-##. 
 
3-2. OP-04 Specifications 
 RS-422/485 Specifications 
Transmission form  Asynchronous, bi-directional, half-duplex 
Data format  Baud rate:  2400, 4800, 9600...

Page 17

- 16 - 
 
 
Example of connection 
RS-422 RS-485 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
‰ The polarity (A, B) of the host computer signal depends on model. Check the 
technical manual of the computer. 
 
3-3. Data output 
Data and commands for the RS-422/485 are same as the RS-232C except following. 
‰ When used with the setting F19-1 (RS-422) or F1-2 (RS-485), set a different 
address F18-## (## = 01 ~ 99) to each scale.   
‰ Add “@##” (## is the address of the scale to send a command) at...

Page 18

- 17 - 
  Examples of command and reply    (“_” shows “Space” (20H).) 
Following examples are for F20-0 (Reply to command is sent.) 
The address ## = 23 (F18-23). 
‰ Request a weight data. 
Command @ 2 3 Q C
LF  
Reply @ 23 S T , +0012
.345_kg C
LF Stable data 
 @ 23 U S , +0007 . 890_kg C
LF Unstable data
  @ 23 O L  ,  +9999 . 999_kg C
LF “E” display 
 
‰ Zero the scale.    (No reply for F20-1.) 
Command @ 2 3 Z C
LF   
Reply @ 2 3 Z C
LF    The scale is within the zero range and stable. 
 
‰ Send a...
Start reading A&D Balance FS HC02iFS03iFS04i User Manual

Related Manuals for A&D Balance FS HC02iFS03iFS04i User Manual