11 5. SETTING THE HOUR The time is set using AM or PM. Press the MEMORY button to advance by increments of one hour. Press the Date/Time Setting button to set the current hour. The minutes flash on the display. 6. SETTING THE MINUTE Press the MEMORY button to advance by increments of one minute. Press the Date/Time Setting button to set the current minute. The year flashes on the display. Press the START/STOP button to turn the display off. SETTING THE DATE AND TIME HEM-741CAN_EN_FR_r1.qxp 5/26/09 9:13 PM Page 10
12 1. Make sure the air plug is securely insertedin the main unit. 2. Remove tight-fitting clothing from your upper arm. 3. Sit in a chair with your feet flat on the floor. Place your arm on a table so the cuff is level with your heart. APPLYING THE ARM CUFF HEM-741CAN_EN_FR_r1.qxp 5/26/09 9:13 PM Page 11
13 4. Apply the cuff to your left upper arm so the blue arrow is centered on the inside of your arm and aligned with your middle finger. The air tube runs down the inside of your arm. The bottom of the cuff should be approximately 1/2" above your elbow. APPLYING THE ARM CUFF 1/2 inch (1 - 2 cm) Air Tube Arrow APPLYING THE CUFF ON THE RIGHT ARM When taking a measurement using the right arm use this instruction. Apply the cuff to your right upper arm so the blue arrow and the air tube runs along the inside of your elbow. The bottom of the cuff should be approximately 1/2'' above your elbow. NOTE: Be careful not to rest your arm on the air tube. This will restrict the flow of air to the cuff. 5. Wrap the cuff firmly in place around your arm using the closure strip. HEM-741CAN_EN_FR_r1.qxp 5/26/09 9:13 PM Page 12
14 The monitor automatically determines your ideal inflation level. In rare circumstances when a higher inflation may be necessary, the monitor automatically re-inflates the cuff up to 30 mmHg higher than the initial inflation and restarts the measurement. 1. Press the START/STOP button. All display symbols appear on the screen. The cuff starts to inflate automatically. As the cuff begins to inflate, the monitor automatically determines your ideal inflation level. Because this monitor detects the pulse even during inflation, do not move your arm and remain still until the entire measurement process has been completed. NOTE: To stop the inflation or measurement, push the START/STOP button. The monitor will stop inflating, start deflating, and will turn off. 2. Inflation stops automatically and the measurement is started. As the cuff deflates, decreasing numbers appear on the display. The Heartbeat Symbol ( ) flashes at every heartbeat. TAKING A MEASUREMENT HEM-741CAN_EN_FR_r1.qxp 5/26/09 9:13 PM Page 13
15 You3blood pressure and pulse rate are displayed. 4. Press the START/STOP button. The monitor turns off. NOTE: The monitor will automatically turn off after five minutes. INSTRUCTIONS FOR SPECIAL CONDITIONS If your systolic pressure is known to be more than 220 mmHg, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your suspected systolic pressure. NOTES: • Do not apply more pressure than necessary. • The monitor will not inflate above 299 mmHg. TAKING A MEASUREMENT 3. When the measurement is complete, the arm cuff completely deflates. Your blood pressure and pulse rate are displayed. HEM-741CAN_EN_FR_r1.qxp 5/26/09 9:13 PM Page 14
16 Every time you complete a measurement the monitor stores the blood pressure and pulse rate in the memory. The blood pressure monitor automatically stores up to 30 measurement values. When 30 sets of measurement values are stored in the memory, the oldest record is deleted to save the most recent measurement values. AVERAGING FUNCTION The monitor also calculates an average reading based on the values of the three most recent measurements that were taken within 10 minutes of the most recent reading. NOTES:• If only two readings are stored in the memory for the10-minute period, the average is based on the two readings. • If one reading is stored in the memory for the 10-minute period, this reading is displayed as the average. If the date and time is not set, the measurement values and the average value will not be stored in the memory correctly. 1. Press the MEMORY button The Average Value Symbol will display on the screen with the average reading. USING THE MEMORY FUNCTION HEM-741CAN_EN_FR_r1.qxp 5/26/09 9:13 PM Page 15
TO DELETE ALL VALUES STORED IN THE MEMORY You cannot partially delete records stored in the memory. All records will be deleted. 1.Press the MEMORY button. 2. While holding the MEMORYand the START/STOP button simultaneously for more than 2 seconds all records will be deleted. 17 TO DISPLAY THE MEASUREMENT VALUES The date and time is alternately displayed with the measurement values. 1. Press the MEMORY button to display the most recent measurement values on the screen. NOTES: • Press the button repeatedly to display the next values. • Press and hold the button to display the values faster. 2. Press the START/STOP button to turn the power off. USING THE MEMORY FUNCTION Alternating display HEM-741CAN_EN_FR_r1.qxp 5/26/09 9:13 PM Page 16
18 To keep your digital blood pressure monitor in the best condition and protect the unit from damage follow the directions listed below: Do not forcefully bendthe arm cuff or air tube. Do not fold tightly. Clean the monitor with a soft dry cloth. Do not use any abrasive or volatile cleaners. Do not attempt to clean the cuff. Never immerse the monitor or any components in water. Store the monitor in a safe and dry location. Do not subject the monitor to extreme hot or cold temperatures, humidity or direct sunlight. Avoid subjecting the monitor to strong shocks , such as dropping the unit on the floor. Remove the batteries if the unit will not be used for three months or longer. Always replace all the batteries with new ones at the same time. Use the unit consistent with the instruction provided in this manual. Use only authorized parts and accessories. CARE AND MAINTENANCE HEM-741CAN_EN_FR_r1.qxp 5/26/09 9:13 PM Page 17
19 ERROR INDICATORS SYMBOLCAUSECORRECTION Cuff under-inflated. Remove the arm cuff. Read, “Taking a Measurement”. Wait 2-3 minutes.Take another measurement. Cuff over-inflated. Movement during measurement.Remove the arm cuff. Wait 2-3 minutes.Take another measurementand remain still until the measurement is complete. Batteries are worn.Replace the four batteries. Refer to “Battery Installation”. Blinking or Steady HEM-741CAN_EN_FR_r1.qxp 5/26/09 9:13 PM Page 18
20 TROUBLESHOOTING TIPS PROBLEMCAUSES AND SOLUTIONS No power. No display appears on the unit.Replace worn batteries. Check the battery installation for proper placement of the battery polarities. Measurement values appear too high or too low. Blood pressure varies constantly. Many factors including stress, time of day, how you wrap the cuff, may affect your blood pressure. Review the sections “Before Taking a Measurement” and “Taking a Measurement.” HEM-741CAN_EN_FR_r1.qxp 5/26/09 9:13 PM Page 19