Panasonic Ew Bu35 Operating Instructions
Have a look at the manual Panasonic Ew Bu35 Operating Instructions online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 10737 Panasonic manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
HOW TO USE English 20 Storage 1. Unplug the AC adapter from the AC outlet. 2. Unplug the plug from the main unit. 5. Place the tube into the cuff. Do not wrap the tube around the main ● unit. It may get damaged by excessive force on the tube. Be careful not to bend the tube. ● 3. Close the hook and loop fastener of the cuff. 1 23 4. Fold the spare cuff material inside the cuff. 6. Place the main unit and the cuff into the accessory pouch. To prevent from unexpectedly operations ● during delivery, do not put the cuff or the accessories on the main unit. 71 8 123 8 BU35.indb 202011/06/14 14:28:28
APPENDIX English 21 Care and maintenance Do not apply excessive force to the instrument or drop the instrument. ● - Doing so may cause damage. Measurements may be affected if the device is used close to televisions, microwave ● oven, X-ray equipment or other devices with strong electrical fields. Measure blood pressure at a sufficient distance from such devices or switch them off. If the instrument is stored at temperatures below the freezing point, do not use it ● immediately. Leave it in a warm place for at least an hour before use. - If the instrument is not allowed to warm up, it may not pressurize. Do not insert dust or foreign objects into the instrument. ● - Doing so may result in damage. If the instrument becomes dirty, clean it with a soft cloth moistened with warm water ● or soapy water. Do not use alcohol, benzene, or paint thinner. - Use of such chemicals may result in cracking or discoloration. For storage, avoid high temperatures, high humidity, and direct sunlight. ● - Storage under such conditions may result in damage. The cuff cannot be washed. ● Specifications Power source DC 6 V (4 AA/LR6 size alkaline batteries) AC 120 V 60 Hz (when using AC adapter) Display Digital LCD Method of measurement Oscillometric system Measurement range Pressure: 0 mmHg to 280 mm Hg Pulse rate: 30 to 160 beats per minute Accuracy Pressure: Within ±3 mmHg Pulse rate: Within ±5 % Operation temperature/humidity range 50 °F to 104 °F (10 °C to 40 °C), 30 % to 85 % RH Storage temperature/humidity range –4 °F to 140 °F (–20 °C to 60 °C), 10 % to 95 % RH Measurable arm circumference Approx. 7-3/4" to 15" (20 cm to 38 cm) Weight Approx. 10.9 oz. (310 g ) (not including batteries) Dimensions 3" × 5" × 4-1/2" (7.9 cm × 13.0 cm × 11.5 cm) Protection against electric shock Internally powered equipment, Type BF applied part APPENDIX The blood pressure monitor might not perform to specifications if stored or used outside the specified temperature and humidity ranges. Input AC 120 V 60 Hz Output DC 6 V 500 mA AC adapter BU35.indb 212011/06/14 14:28:28
APPENDIX English 22 Explanation of rating plate symbols Read the operating instructions carefully before use. Protection against electric shock Type BF applied part (Blood pressure monitor) Protection against electric shock Class II equipment (AC adapter) AC DC BU35.indb 222011/06/14 14:28:29
APPENDIX English 23 Q&A Q Why are the blood pressure values measured at home different from those measured at the doctor’s office? A Blood pressure varies 24 hours a day. In addition, it can change greatly depending on the weather, your mental state, exercise, etc. . In particular, there is the phenomenon of “White coat hypertension” at doctor’s offices, where the blood pressure measured in a doctor’s office is higher because of stress or worry. In addition, if the blood pressure measured at home is sometimes higher and sometimes lower than the blood pressure measured at the doctor’s office, the following can be considered. Please pay attention to these points when taking measurements. 1. Is the cuff wrapped firmly?Make sure that the cuff is not wrapped too loosely or too tightly. ● Is the cuff wrapped around the upper arm? Make sure it is not partly covering the elbow. ● See page 8 for how to wrap the cuff correctly. 2. Were you worried or upset during measurements? Before taking measurements, take 2 or 3 deep breaths and relax, so that your blood pressure stabilizes. Q When should I measure my blood pressure? A We recommend three measurements a day. First .......... Upon Waking (Because, with almost no influence from body movement, the blood pressure is closest to that at night.) Second ... Before Lunch (Because, with no influence from the meal, it provides blood pressure values close to the highest of the day.) Third ......... At Bedtime (So as to obtain data close to your average blood pressure values during the day.) This way, by taking measurements morning, noon and night, you can use the unit to learn the patterns of and monitor your blood pressure. Use of the "Comparing with Old Measurement" function can aid in the detection of a ● blood pressure pattern known as “early morning surge” in which blood pressure rises rapidly in the morning. (See page 18.) Q Since the indicator was displayed, I retried the measurement by setting my arm in the proper position. But the indicated value is still higher than usual. A The indicator is just a guide for the proper measurement method. If you correct your posture when the indicator is displayed but the measured value is still higher or lower than usual, retry the measurement. The following causes are suspected. Did you measure properly (in proper posture)? ● Measure as indicated on pages 12–13. ● Measured immediately after smoking. ● Measured when irritated or emotional. ● Measured in cold environments. ● For persons who have a physical constitution with less fluctuation in the pulse rate. ● If you are concerned about the results, retry the measurement after staying quiet for 4 to ● 5 minutes. BU35.indb 232011/06/14 14:28:29
APPENDIX English 24 Q The indicator does not light when an arm is moved. A The indicator is a function to notify you that your arm is bent or to indicate extra loads on the cuff. Therefore, it may not indicate motions (bending an arm slightly, etc.) that do not affect the cuff. If you are concerned about the result, retry the measurement. Q&A (continued) Troubleshooting Display Status before error Cause and solution is displayed. The indication disappears in the middle of the operation. (the mark is displayed.)Batteries are exhausted. (See page 10.) is displayed. Your arm or body moved. Has the unit detected movement? (See page 15.) Pressurization to more than 280 mm Hg was performed. (Cuff pressurizes several times.) Were measurements taken according to the proper procedure and in the correct posture? (See page 13.) Pressure decreased suddenly. Is the cuff wrapped properly? ● Is the cuff plug inserted correctly into ● the socket? (See page 12.) Cuff does not inflate. The measurement is not possible though the cuff inflates. Cuff pressurizes several times. is displayed. Failure of the main unitTake it to the store where it was purchased for testing and repairs. BU35.indb 242011/06/14 14:28:30
APPENDIX English 25 Troubleshooting (continued) SymptomPossible cause Systolic or diastolic blood pressure value is high or low. The position of the cuff is higher or lower than your ● heart when measuring. Cuff is not wrapped correctly. ● You moved or talked during measurements. (See page ● 13.) Value is different from that measured by the doctor. Blood pressure varies in response to minute changes ● in your mental state, such as your reaction to a doctor’s presence. Sit quietly for 4 to 5 minutes and then take another ● measurement. (See page 22.) Measurement values are different with each measurement. No display appears even when the OFF/ ON & START button is pressed. Batteries are installed with the polarities wrong.(See ● page 10.) Batteries are exhausted. (See page 10.) ● The indicator flashes when readings are being taken. There was a significant fluctuation in pulse when readings were being taken. (See pages 15–16.) The indicator flashes when readings are being taken. Your arm or body moved or the buttons are pressed during the measurement. (See page 15.) The flashes even though I did not move my arm or body. In some cases this will be displayed when you have tensed the muscles in your hand. Relax the muscles in your hand and take the reading again. (See page 15.) Instrument is broken. Take it to the store where it was purchased for testing and repairs. BU35.indb 252011/06/14 14:28:31
APPENDIX English 26 Federal Communication Commission Interference Statement This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one of the following measures: Reorient or relocate the receiving antenna. ● Increase the separation between the equipment and receiver. ● Connect the equipment into an outlet on a circuit different from that to which the ● receiver is connected. Consult the dealer or an experienced radio/TV technician for help. ● FCC Caution: To assure continued compliance, install and use in accordance with provided instructions. Any changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate this equipment. FCC STATEMENT Product Information for FCC This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. BU35.indb 262011/06/14 14:28:31
For questions or assistance with your blood pressuremonitor, call us at 1-800-338-0552. Panasonic Corporation of North AmericaOne Panasonic Way 1H-1 Secaucus, NJ 07094 1-800-338-0552 (USA only) for questions and comments No.1 EN, SP ¢ žÝæ§ £ © Panasonic Electric Works Co., Ltd. 2011 1 W9030BU35201 Issued date: 2011-06-14 Printed in China Impreso en China BU35.indb 292011/06/14 14:28:43