    							20
    Setting (checking) the measurement notification time 
    19
    M/R
    Calling up readings from memory
    Readings can be recalled after being saved without
    having to switch the unit on.
    1. Press the Memory/Recall button.
    •The average readings of all recorded measurements
    are displayed.
    •The  mark will be displayed (Fig. 11-a).
    •When, for example, readings from 90 measurements
    have been saved the average of the readings from 90
    measurements will be displayed.
    2. Press the Memory/Recall button to display readings
    starting with from the most recent measurement.
    •Each time the Memory/Recall button is pressed
    readings are displayed in order of the most recent
    measurement (Fig. 11-c, 11-d).
    •No readings will be displayed when there are none in
    the memory.
    3.Press the On/Off and Start button to switch off the
    unit.
    If you forget to switch off the unit it will do so
    automatically after approx. 30 seconds.
    Deleting all data stored in memory
    (1) Recall stored data using the Memory/Recall button.
    (2)Press the Memory/Recall button again and hold it (for approx. 3
    seconds) until all readings are deleted.
    NOTE: Readings from individual measurements cannot be deleted.
    (3)When data has been deleted the  mark and  will be
    displayed on the screen.
    Fig. 11-a
    Fig. 11-c Fig. 11-b
    Fig. 11-dTo let you take measurements at the same time every
    day, 3 times per day can be set as notification times. 
    1. Press the Set button on the top surface of the blood
    pressure monitor (Fig. 12-a). 
    •Notification Time 1 will blink (Fig. 12-b).
    *If date and time have not been set, the display will
    change to the date/time setting display.
    2. Press the Set button again.
    •The  and  will blink (Fig. 12-c).
    3. Press the Adjust button to display  . 
    •The  and  will blink (Fig. 12-d). 
    •Notification sound will sound. 
    •Notification sound will not sound. 
    •When  is displayed and the Set button is 
    pressed, the time becomes  and the 
    and will light.
    4.(1)Press the Set button. The hour value will blink
    (Fig. 12-e). 
    (2)Press the Adjust button repeatedly to set the
    desired hour.Fig. 12-a
    Fig. 12-c Fig. 12-b
    Fig. 12-d
    Fig. 12-e (After approx.
    one second.)
    &864@&OH   Ö”´ 
    						
    							22 21
    5. (1) Oprima el botón Puesta. El valor de los minutos
    destellará (Fig. 12-f).
    (2)Oprima el botón Ajustar repetidamente para
    establecer los minutos deseados.
    •Para ajustar sólo la Hora de Notificación 1,
    oprima el botón Arrancar/Parar para comletar el
    ajuste.
    •Para ajustar las Horas de Notificación 2 y 3, siga
    los pasos 6 y 7 de más abajo.
    6Oprima el botón Puesta. La Hora de Notificación 2
    destellará (Fig. 12-g). Repita los pasos 2 a 5 para
    establecer la hora de notificación deseada.
    7Cuando se ha completado el ajuste de la Hora de
    Notificación 2, oprima el botón Puesta. La Hora de
    Notificación 3 destellará. Repita los pasos 2 a 5 para
    establecer la hora de notificación deseada.
    8Cuando el ajuste de la Hora de Notificación 3 ha sido
    completado, oprima el botón Puesta. Las horas
    ajustadas para las Horas de Notificación 1 a 3 serán
    mostradas automáticamente a intervalos de 2
    segundos (Fig. 12-h).
    9Para completar el ajuste, oprima el botón
    Arrancar/Parar.To change the notification time setting: 
    1. Press the Set button on the top surface of the blood pressure monitor.
    (Notification Time 1 will blink.) 
    •To change the setting for Notification Time 2 or 3, press the Change
    button to display Notification Time 2 or 3. 
    2. Press the Set button twice. The hour value will blink. 
    3. Press the Adjust button repeatedly to set the desired hour. 
    4. (1) Press the Set button. The minute value will blink 
    (2) Press the Adjust button repeatedly to set the desired minute. 
    5. Press the Start/Stop button to complete the setting. 
    To cancel the measurement notification sound: 
    1. Press the Set button on the top surface of the blood pressure monitor.
    (Notification Time 1 will blink.) 
    •To change the setting for Notification Time 2 or 3, press the Adjust button
    to display Notification Time 2 or 3. 
    2. Press the Set button once. The  and  will blink. 
    3. Press the Adjust button.  will be displayed. 
    4. Press the Start/Stop button to complete the setting. 
    •When the set time is reached, the measurement notification sound will
    sound to notify you that its time to take a measurement. 
    •If the set time is reached in the middle of taking a measurement, the
    measurement notification sound will not sound. 
    •When the measurement notification sound sounds, it can be stopped by
    pressing the Start/Stop button. If the button is not pressed, the sound
    stops after 1 minute. 
    •If the blood pressure monitor is put into a bag, desk, etc., it may not be
    possible to hear the measurement notification sound. 
    Advice: 
    If a record table is made of the measurement values taken around the same 
    time, you can easily see your own daily blood pressure changes. Fig. 12-f
    Fig. 12-g
    Fig. 12-h
    Fig. 12-i
    &864@&OH   Ö”´ 
    						
    							24 23
    Troubleshooting
    Display
    The “E” blinks on
    the display 
    Status before error
    Pressurization to above
    280 mmHgwas
    performed.
    (Re-pressurization was
    performed several
    times.)
    Pressure decreased
    suddenly and 
    appeared in the display.
    mark only blinks a
    few times, or not at all.
    Cuff does not inflate.
    Check points
    Check whether
    measurements are
    being taken correctly
    and whether you are
    assuming the correct
    posture. Make sure
    the unit is on the palm
    side of the wrist. 
    (See pages 12-13.)
    Check whether the
    pressure cuff is
    attached to your wrist
    correctly. Make sure
    the unit is on the palm
    side of the wrist.
    (See page 12.)
    Care and Maintenance
    1. Use a dry, soft cloth to clean the device. If desired, use a cloth lightly
    dampened with tap water and a mild detergent to clean the device. 
    DO NOT use alcohol, benzine, thinner or other harsh chemicals to clean
    the device or cuff.
    2. When the unit will not be used for a long time (30 days or more), be sure
    to remove the batteries. Otherwise, the batteries may leak and damage
    the unit.
    Storing the Monitor
    1. Fold the pressure cuff as indicated.
    (Fig. 13-a)
    2.Put the blood pressure monitor in
    the storage case. (Fig. 13-b)
    •Be sure not to pinch the pressure
    cuff or 2cm gauge when closing the
    case.
    1
    2
    0
    8
    0
    If the 2cm gauge is attached, 
    straighten it out.
    Fig. 13-a
    Fig. 13-b
    NOTE:When the batteries are removed, the date, time, and measurement
    record are erased. Set the date and time again the next time you
    use the unit.
    &864@&OH   Ö”´ 
    						
    							26 25
    Specifications
    Method of measurement:
    Display:
    Measuring range:
    Measurement accuracy:
    Power source:
    Operating environment:
    Storage environment:
    Wrist circumference
    measurement range:
    Dimensions (H#W#D):
    Weight:
    Oscillometric System
    Digital LCD
    Blood pressure: 0 - 280 mmHg
    Pulse rate: 30 - 160 pulse/min.
    Pressure: ±3 mmHg
    Pulse rate: ±5%
    Two 1.5-volt alkaline (AAA/LR03) batteries
    10°C to 40°C (50°F to 104°F)
    30% to 85% RH
    -20°C to 60°C (-4°F to 140°F)
    10% to 95% RH
    12.5 to 22.0 cm (5 to 8-3/4)
    54.0#83. 0#26.0 mm 
    (2-1/8#2-1/4#1-1/32)
    90 g(3.2 oz.) not including batteries
    If the unit still appears to provide unusual or erroneous readings,
    consult your physician. If the unit does not appear to be functioning
    properly, contact Panasonic at 1-800-338-0552
    .
    Specifications are subject to change without notice.
    Symptom
    The systolic pressure
    value or diastolic
    pressure value is high. 
    The systolic pressure
    value or diastolic
    pressure value is low. 
    The measured values
    differ significantly.
    The measured value
    differs from that
    measured by the doctor.
    The measured value is
    different for each
    measurement.
    Varying sounds and
    pressurization speeds
    occur during inflation of
    the cuff. 
    The measured value
    obtained is different
    from that measured at
    the upper arm.
    Possible cause
    •The hand with the cuff was held too low and
    not at heart level. (See page 12)
    •The cuff was not attached properly. (See
    page 12.)
    •You moved your hand or body or you spoke
    during measurement. (See page 12.)
    •The hand with the cuff was held too high and
    not at heart level. (See page 12.)
    •You moved your hand or body or you spoke
    during measurement. (See page 12.)
    •Your posture was not the same for all
    measurements, or measurements were
    taken without sufficient time in between.
    Always wait approximately 5 minutes before
    taking another reading, and above all, relax.
    •Blood pressure varies in response to
    minute changes in your mental state, such
    as your reaction to your doctor taking the
    measurement. One way to check this is to
    bring the unit to your doctor’s office and
    directly compare the readings taken by
    your physician.
    •Relax for at least 5 minutes and then take
    another measurement.
    •Changes in operating sounds and
    pressurization speeds are normal and occur
    due to the fact that measurements are
    being made as the cuff is being inflated.
    •For persons with certain circulation problems,
    there may be substantial differences in
    blood pressure values measured at the
    upper arm versus at the wrist.