41 | SPECTRAVIEW II - USER’S GUIDE Calibrating and Profiling the Displa\f If more than one supported display monitor is connected, select the display to be calibrated from the Display listbox. Note that each display has an independent Target. Click the Calibrate button to start the calibration process. The software will provide instructions for placing the calibration sensor on the screen and provide updates as the calibration progresses. Once the calibration process has been completed, an ICC/ColorSync profile will be generated if the Generate profile after calibration option in the Preferences dialog is selected. The profile will automatically be associated with the system’s Color Management System. The results of the calibration process will then be displayed in the Information window. It is also good practice to use the Test Patterns to quickly visually verify the integrity of the calibration. Profiling the Displa\f An ICC/ColorSync monitor profile of the display can be generated without calibrating the display by selecting Generate ICC Profile.. from the File menu. Non-SpectraView displays can be profiled but not calibrated by SpectraView. Validating the current Calibration The current state of the display monitor can be measured and compared to the currently selected Target in order to quickly determine if the current calibration is still accurate, or if the display should be recalibrated. A series of measurements will be taken on the display and a brief report will be shown of the results. The Information window will also be updated with a more detailed analysis of the results. Note: Several factors can influence the results of the validation process and a Delta-E values of 3-5 do not always indicate that the display needs to be recalibrated. Even very slight differences in the positioning of the sensor from when it was calibrated, and the length of time the monitor has been powered on, and the basic measurement repeatability accuracy of the sensor itself, all impact the results. Also keep in mind that the Delta-E value is comprised of the luminance and color difference. Differences in color are much more critical than a difference in relative luminance.

42 | SPECTRAVIEW II - USER’S GUIDE Using ICC/ColorS\fnc Monitor Profiles The ICC/ColorSync monitor profiles generated by SpectraView are used to inform the Color Management System of the color characteristics of the display monitor. Color Management Systems use this information to determine how to display images on-screen. Depending on the image editing software package being used, profiles for printers and scanners can also be utilized to create accurate image prints and scans. The exact method of loading ICC monitor profiles into the Color Management System will depend on the software application being used. Most software applications rely on the operating system (Mac OS or Microsoft Windows) to provide the current profile information. However some applications may have a color management setup where the display monitor profile can be selected. SpectraView will automatically select and associate the generated ICC/ColorSync monitor profiles with the system’s Color Management System. Refer to the instructions for the particular software package being used to determine how to correctly load and use the ICC/ColorSync monitor profiles generated by SpectraView. Adobe Photoshop Overview When combined with accurate printer profiles, Photoshop can perform accurate CMYK previews or soft proofs of what an image will look like when printed. The hardware based calibration of the SpectraView display monitor will provide superior results compared to the basic visual calibration offered by the Photoshop or ColorSync Control Panels. Note that it is not necessary to perform any of the monitor settings in the Adobe Gamma Control Panel since all of the necessary monitor calibration and measurements will be done by SpectraV iew. In Photoshop, images are manipulated in what is known as an image color workspace. This color workspace is independent of the display monitor color characteristics, allowing images to be transferred between systems with different monitor settings without loosing any color information or restricting the color gamut of the image. Photoshop will convert colors on the fly when it displays the image on the SpectraView monitor, so the colors you see accurately reflect the file’s working color space. The selection of an RGB color workspace will depend on the application. For details of the standard color workspaces included in Photoshop, consult the online help in Photoshop. Note that typically the color workspace will be different to that of the display monitor. Do not attempt to load the ICC/ColorSync monitor profile as the color workspace in Photoshop. When editing an image in CMYK format, Photoshop uses the monitor color characteristics to perform the CMYK- to-RGB conversion necessary to display the image accurately on the display monitor, which uses a RGB device dependent color space. When editing or viewing an image in RGB format, Photoshop also uses the display monitor color characteristics, together with a set of printer separation tables or printer ICC/ColorSync profile to perform the RGB-to-CMYK conversion necessary to preview or soft proof a print the image. Printer separation tables are usually provided by the printer manufacturer or other third party, or can be generated using a 3rd party printer profiling application. 5 Ou\fsi\be of Spec\fraView

43 | SPECTRAVIEW II - USER’S GUIDE Calibrati\fn Question 1: How often do I need to calibrate my monitor? Answer: The answer will depend on how critical a color accuracy is necessary, how long the monitor is used per day, environmental conditions and the characteristics of each individual monitor. Some users working in highly critical color applications may calibrate their displays every couple of days. Others may only need to do it once a month. The recommended and default re-calibration period is every 2 weeks. The Validate Calibration feature can always be used to verify the display’s white point and Intensity and determine if the monitor needs to be re- calibrated. Always remember to allow the monitor sufficient time to warm up and reach optimum color performance before performing any color critical work or calibrating the display . Question: ( Mac OS only) Why can’t I use the ColorSync control panel to switch between different display calibrations like I can do with other monitors? Answer: When used with the SpectraView software, NEC display monitors store all of the necessary color adjustments internally, including the 10 or 12 bit Look Up Tables. The ColorSync utility switches the display color settings by changing the Look Up Tables in the video graphics adapter. The ColorSync profiles generated by SpectraView contain linear Look Up Tables for the video graphics adapter. Therefore the SpectraView software must be used to update the display monitor. Question 2: ( Mac OS only) I’m using a wide color gamut display together with the standard gamut screen on my MacBook. Colors in Safari look oversaturated when the window is on the wide gamut display . What is wrong? Answer: The color management in Safari and some other Mac applications will only apply color correction for the primary display (the display with the menu bar). When a window is moved to the secondary display, the correct color correction for that display is not applied. This causes the images to be oversaturated on the secondary display. To display colors correctly on the wide color gamut display it must be made the primary display. Use the Mac OS System Preferences - Displays to configure the display as the primary. Question 3: I’m using two different displays and calibrating both to the same target, however the white points do not match exactly. What can I do to improve this? Answer: This issue can be caused by the different spectral outputs of the two displays causing the human eye to see color differences differently than the color sensor used to calibrate the displays. In this case the numerical measurements from the color sensor are identical on both displays, but they do not visually match. Manually visually matching one display to another can usually avoid this issue. The basic steps to take are as follows: 1. Calibrate both displays. 2. Pick one of the displays to be a reference. Call this display #1. 3. Select display #2 in SpectraView, open the Edit Calibration Target dialog then open the Custom White Point dialog by clicking on the Edit... button. Use the Visual Match tool to adjust the white color so it visually matches white shown on display #1. 4. When a visual match is achieved, put the sensor over the measurement area and click Measure. This will enter the value that the sensor perceives into the White Point. Save this as a new Target. 5. Recalibrate display #2 with this new Target. From now on always use the respective Target to calibrate each 6 Frequen\fly Aske\b Ques\fions

44 | SPECTRAVIEW II - USER’S GUIDE display. Question 4: I’m trying to match my display to my prints in order to soft proof, however after calibrating using a standard Target white point such as D65 or D50, the screen still appears too pink/magenta/green compared to the print. What can I do to make them match? Answer: Depending on many factors such as the lighting conditions used to view the prints, paper type, and the printer profiles used for the soft proofing, the display white point may not visually match that of the paper. It is important to remember that the preset white points such as D65 are not magic numbers that must be used regardless. The calibration Target white point can be adjusted as necessary in this scenario. This is especially true if a standard viewing environment such as a light box is not being used to view the prints. Adjusting the white point using a Kelvin or Daylight value alone is adjusting the white point along the fixed Black Body curve as viewed on a CIE xy diagram, and can vary the white point from cool to warm. However, the color cast between the print and display may be at right angles to this curve, so in this case no Kelvin value will be able to fully compensate. A white point that lies off the standard Black Body curve can be entered by specifying it in terms of CIE xy instead of as a Kelvin or Daylight value. To get an idea of how CIE xy values relate to the actual White Point color, try zooming in on the Color Gamut diagram on the Information window. In the example below, the Black Body curve is shown. If for example the display appears to be too green, then the white point should be moved slightly towards the lower right - this corresponds to increasing the CIE x value, and decreasing the y value slightly. Modify the Target White Point values by making small adjustment (for example increments of +/-0.003) and recalibrate the display. Several iterations may be required to find the best values. In addition to this method, the Visual Match tool described in the previous answer can be used to interactively find a visual match between the print and display . Black Body curve

45 | SPECTRAVIEW II - USER’S GUIDE Question 5: Each time I calibrate my monitor I receive a different Delta-E result. Should I recalibrate again if I receive a value that is high? Answer: For Delta-E values of less than 2 there is almost certainly no need to recalibrate. For values higher than 3 it is important to understand what the cause is, and if it is significant. The calibration and adjustment process has inherent variability due to many factors such as the repeatability of color sensor measurements, and changes within the display as it is adjusted. The basic accuracy and repeatability specifications for most sensors is typically +/- 0.002 or higher for CIE xy, and +/- 2% for luminance. These variances alone can give rise to Delta-E values of up to 2 or 3. When a specific Intensity value is being used for the calibration, the Delta-E value is comprised of the luminance and color differences. Differences in color are much more critical than a difference in relative luminance. Look at the results of the calibration shown in the Information window to determine if the high Delta-E value is caused mainly by a shift in color or in luminance. High Delta-E values can also occur if a specific Intensity value is being used for the calibration and the display is not capable of achieving the specified Intensity value. The calibration will always aim to achieve the specified White Point, even if the Intensity value could not be reached. If the display can not be calibrated to achieve the specified Intensity value modify the value accordingly. In some situations, the display Intensity will drift slightly for several minutes after it is adjusted. Since the measured value used to calculate the final Delta-E result is taken up to one or two minutes after the display was adjusted, a higher Delta-E value may be shown, even though the calibration process achieved the specified Target values during the calibration process. Selecting the Extended luminance stabilization time option in the Preferences can be used to try and minimize this effect. Question 6: The Delta-E result values shown in SpectraView are higher than those shown in other display calibration applications. Does that mean they are better at calibrating the display than SpectraV iew? Answer: No. SpectraView calculates the Delta-E values using both the luminance and color differences for more realistic and accurate values. Other applications may calculate Delta-E differently. Question 7: After calibration I manually adjusted some of the monitor controls to get a better color match for my particular application. What happens the next time I calibrate the monitor? Answer: The manual adjustments will be reset by the calibration process. In general it is not a good idea to manually adjust the monitor after calibration because it not only invalidates the calibration and ICC/ColorSync profile, but also means that the adjustments will be lost the next time the monitor is calibrated. If you still feel it is necessary to tweak the display after calibration, a much better approach is to create a new Target file based from the measurements of the display after it has been manually adjusted. That way the monitor will be automatically calibrated to that particular set of characteristics the next time it is calibrated. Use the Custom White Point tool to measure and set the white point and Intensity values.

46 | SPECTRAVIEW II - USER’S GUIDE Installati\fn and Setup Problem: Calibrate button is disabled. Solution: The calibrate button is disabled for either of the following reasons: • No supported calibration sensor was detected. • No supported display monitor was detected. See the following sections for further explanations. Problem: No supported monitor was detected. Solution: • Verify that the display monitor connected is a model supported by this version of SpectraView II. Check if a newer version of the software is available from NEC. • If more than one video output is present on the computer, try connecting to the other output connector. • Always use a DVI or DisplayPort video output if possible. • Connect a USB cable between display to the host PC (MultiSync PA Series only) and restart SpectraView. Mac OS: • See the README file included with the software for further information about any specific incompatibilities or issues. Note that some Macs require at least OS 10.5.2 in order support the functionality necessary with SpectraView. ÿ Windows: • Some video graphics adapter cards may not support the necessary functions for SpectraV iew to communicate with the display monitor. For Windows systems, verify that the video graphics adapter is supported by SpectraView by checking the Windows Graphics Chipset Compatibility Information. See the README file included with the software for further information. • Install the latest video drivers for the video graphics adapter card. • Communications with the display monitor requires a special driver to be installed. This is installed and configured when SpectraView is installed, however any of the following changes may necessitate that SpectraView be re-installed to re-configure the driver: • Updating the video graphics adapter card driver. • Changing the monitor model • Changing the video port that the monitor is connected to. • Check the display configuration by clicking the Display Configuration button. In the Display Configuration dialog, confirm that your model of display monitor is listed. If it is not listed, then there is a problem communicating with and detecting the display monitor. If it is listed, confirm that the monitor is associated with the correct Windows desktop display. 7 Troubleshoo\fing

47 | SPECTRAVIEW II - USER’S GUIDE Problem: No supported calibration sensor was detected. Solution: • Try reconnecting the calibration sensor and click the Auto Detect button in the Preferences dialog to re-detect the sensor. • It may be necessary to close and re-launch the SpectraView application. • (ÿ Windows) It may be necessary to install the drivers for the color sensor device. Note that only the drivers for the X-Rite iOne devices are installed by the SpectraV iew setup application. Drivers for other devices must be installed manually. Problem: Severe color banding appears after calibrating on images and when the test pattern is shown. Solution: • Always use a DVI video output if possible. • If an analog video signal must be used perform the following steps once prior to calibration: 1. Display something on the screen that has both full black and full white, such as the Test Pattern in SpectraView. 2. On the monitor’s On Screen Display, select and operate the AUTO CONTRAST control. 3. Proceed with the calibration. Problem (ÿ Windows): The ICC profile generated by SpectraView does not appear to be correctly installed and recognized in Windows. Solution: • This problem may occur especially when using a multi-monitor configuration. In some cases updating the video driver may fix this issue. • The profile can be manually assigned if necessary as follows: 1. In the Windows Display Properties control panel, select the Settings tab and click Advanced. 2. Select the Color Management tab on the Display Properties dialog. 3. Find the color profile generated by SpectraView and use Set As Default. • A more advanced control panel is available from Microsoft for Windows XP which is much more reliable for checking the assigned monitor profile. On the www.microsoft.com website, search for “Color Control Panel Applet”. Problem (ÿ Windows): SpectraView is unable to save the generated ICC profile. Solution: If SpectraView is unable to save the ICC profile for this reason it will prompt you with the name and location of the file that needs to be manually deleted in order to replace it with a new profile. • This issue may occur if there are multiple users assigned to the machine. An ICC profile created by a user can normally only be updated and replaced by that user, or an administrator. Starting with Windows Vista, even if you are currently an Administrator, applications do not normally allow full administrative privileges by default. You can either delete the current ICC profile manually (in which case Windows will ask for administrative level credentials or permissions), or you can re-run SpectraView “as Administrator” by right-clicking the

48 | SPECTRAVIEW II - USER’S GUIDE application icon and selecting “Run as administrator”. This process needs to be done once each time the user who created the last ICC profile for the display is changed. • This issue may also occur if another application is currently using the ICC profile. For example if the Windows Advanced Display Settings control panel is open, the ICC profile may be locked. Be sure to close any applications that may be using the profile.

49 | SPECTRAVIEW II - USER’S GUIDE Using SpectraVie\b \bith NaViSet Administrat\fr ( ÿ Wind\f\bs versi\fn \fnly) The SpectraView II software integrates with the NEC DISPLAY SOLUTIONS NaViSet Administrator software to provide remote administration of the display monitors on networked machines. NaViSet Administrator will report all of the current SpectraView settings and the current calibration status within the NaViSet Administrator Console application. When using the Single Connection feature of NaViSet Administrator to administer a remote machine, the SpectraView information will be shown in the monitor metadata report. When using the Monitor Asset Report feature to poll and gather information about the display monitors on multiple remote machines, the SpectraView information can be included in the Output Monitor Asset Report file by selecting the items to include in the report in the Configuration Wizard. 8 NaViSe\f A\bminis\fra\for Example metadata report showing the SpectraView settings using the Single Connection feature.

50 | SPECTRAVIEW II - USER’S GUIDE 9 Tes\f Pa\f\fern Gui\be The SpectraView II software includes several test patterns that are useful in evaluating the monitor calibration. It is also good practice to use the Test Patterns to quickly visually verify the integrity of the calibration. A pattern based on the DICOM TG18-GC test pattern has a number of useful features that allow very detailed analysis of the monitor grayscale performance in one pattern. The main features of this pattern are described below. “QUALITY CONTROL” characters with decreasing contrasts at minimum, mid-point, and maximum pixel values for simple low-contrast evaluation at three luminance levels. Sixteen luminance patches with pixel values varying from 8 to 248 for luminance response evaluation. Each patch contains four small 10x10 pixel corner patches at ±4 of pixel value difference from the background, +4 in upper-left and lower-right, -4 in lower-left and upper-right. The small patches are used for visual assessment of luminance response. Two patches of minimum and maximum pixel value with 13, and 242 pixel value internal patches, similar to 5% and 95% areas in the SMPTE test pattern. 256 level grayscale with 2 lines per level. Line-pair patterns for resolution evaluation, having pixel values of 0-255 and 128-130. Grey background of pixel value 128. Several other patterns are available in SpectraView and can be selected using the buttons at the top left of the screen, or using the [ and ] keys.