Home > Miele > Washer Disinfector > Miele G 7893 Manual

Miele G 7893 Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Miele G 7893 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 171 Miele manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Switching on
    ^Close the door.
    ^Open the stopcocks.
    ^Press theI-0switch.
    When the programme selector is atf,
    a dot lights up in the display.
    Starting a programme
    Choose your programme according to
    the type of instruments or utensils being
    cleaned and the degree of soiling.
    The different programmes and their
    uses are described in theProgramme
    chartat the end of this booklet.
    ^
    Turn the programme selector
    clockwise or anti-clockwise to the
    required programme.
    The temperature of the first stage of the
    programme selected will appear in the
    display, except withDand6.
    The indicator light next to the start
    button6will flash.Selecting the additional Drying0
    function
    The additional Drying function may be
    selected immediately after the
    programme has been selected (except
    with Pre-rinse programmeD).
    To do this:
    ^Press0.
    The programme will be extended by
    Drying block 1 (if available) and 2; see
    Programme chart.
    Drying time 2 is saved and shows in the
    display. This can be altered in 5 minute
    steps by pressing the0button.
    The new drying time is programmed in
    for the programme selected after the
    programme start.
    The total running time for the
    programme increases accordingly.
    ^
    Press the Start button6.
    The current temperature will appear in
    the display.
    The indicator light next to the Start
    button6will light up.
    Operation
    31 
    						
    							Once the programme is running, all the
    other programmes are blocked.
    If the programme selector is then
    turned to another programme, the
    values shown in the display will go out.
    The values can be seen again if the
    programme selector is turned back to
    the position of the programme currently
    in operation.
    During a programme, you can switch
    between the following displays:
    –the current temperature,
    –the elapsed programme duration and
    – the current block:
    1 = Pre-wash 1
    2 = Pre-wash 2
    3 = Main wash 1
    4 = Main wash 2
    5 = Chemical disinfection*
    6 = Rinse 1
    7 = Rinse 2
    8 = Rinse 3
    9 = Rinse 4
    10 = Final rinse 1
    11 = Final rinse 2
    12 = Drying 1
    13 = Drying 2
    ^
    Press
    						
    							At the end of a programme
    The programme has ended whenr
    lights up constantly in the programme
    sequence display and the6indicator
    has gone out. An0will appear in the
    middle of the display. All the other
    indicators come on and go out in
    sequence.
    At the same time a buzzer sounds for a
    maximum of 30 seconds (default
    setting).
    The buzzer can be changed; see
    "Programmable functions".
    Use the
    						
    							To cancel a programme
    Once a programme is running, it should
    only be cancelled in extreme cases,
    e.g. when articles in the machine rattle
    or jostle against each other or the
    wrong programme has been selected.
    Application: Medical equipment
    When testing to EN ISO 15883-1 or
    HTM 2030 standards, a programme
    must not be interrupted or
    cancelled.
    ,With officially registered
    epidemics, any contaminated water
    must be decontaminated using a
    disinfecting agent before it is
    discharged into the sewerage
    system.^Turn the programme selector tof.
    The programme will be cancelled after
    approx. 2 seconds.
    ^Open the doora.
    ,Caution. Water and items in the
    machine may be hot. Danger of
    burning or scalding.
    ^Arrange the load securely.
    (Wear protective gloves where
    necessary, and observe appropriate
    health and safety regulations relating
    to infectious diseases).
    ^Replenish powder cleaning agent if
    necessary.
    ^Close the door.
    ^Start the Drain6programme.
    The water will drain away.
    ^Select and start the programme
    again.
    The following procedure should be
    followed when dealing with official
    cases requiring the containing of
    epidemics using the SPECIAL
    93°C-10’ programme:
    ^
    Start the programme again.
    The display will show the temperature
    holding time required counting
    downwards.
    Operation
    34 
    						
    							For future sevicing work, please document any alterations to the factory default
    settings.
    Enter any alterations in the appropriate box.
    General notes:
    –Switch positions that have not been allocated a function are indicated in the
    middle part of the display by a bar (-).
    –The programming level is called up via thefposition.
    It is shown asE..in the display.
    –When processing medical devices, any changes made to programmes or
    dispensing systems must be documented in a log book kept with the machine
    (in accordance with the Medical Devices Directive). The machine's cleaning
    and disinfecting standards must be re-validated.
    The following applies to all programmable functions:
    To activate programming mode
    ^Turn the programme selector tof.
    ^Switch off the machine.
    ^Press and hold in
    						
    							Dispensing liquid cleaning agent via DOS 1
    Setting the dosage concentration
    The dosage concentration must be set exclusively for liquid cleaning agent.
    Set it according to the manufacturer's recommendations or the official
    standards required.
    The dosage concentration must not be altered even if powder cleaning agent
    is dispensed via the dispenser in the door.
    Step Display / Result
    ^Press0until E02 appears.
    E02
    ^Turn the programme selector toB.
    without DOS Module connected:
    with DOS Module connected:10= 10 sec dosage duration
    1.00= 1.0 % dosage concentration
    ^Press
    						
    							Dispensing neutralising agent via DOS 3
    Setting the dosage concentration
    The dosage concentration for the neutralising agent must be set according to
    the manufacturer's recommenedations or the official standards required.
    Step Display / Result
    ^Press0until E02 appears.
    E02
    ^Turn the programme selector toF. The dosage concentration which
    has been programmed in will
    appear in the display, e.g.0.10=
    0.10 %
    ^Press
    						
    							Buzzer
    A buzzer can be programmed for the following:
    – a constant tone will sound at the end of a programme,
    – a buzzer will sound at one second intervals as a fault warning, and
    – 3 short beeps followed by a 2 second pause, then 3 short beeps etc will sound
    as a reminder to remove a test item.
    The reminder to remove a test item function has to be programmed by an
    approved Miele service technician.
    The buzzer settings are displayed as numbers:
    20Switched off
    21At the end of a programme (factory default setting)
    22Fault indication
    23At end of a programme + fault indication
    24Reminder to remove a test item
    25At the end of a programme + reminder to remove a test item
    26Fault indication + reminder to remove a test item
    27At the end of programme + fault indication + reminder to remove a test item.
    Step Display / Result
    ^Press0until E04 appears.
    E04
    ^Turn the programme selector toB.
    21
    ^
    Press
    						
    							Changing programme parameters
    The settings for individual parameters set at the factory are outlined in the
    Programme chart at the end of these operating instructions.
    Programme parameters can only be changed if they are activated in a
    programme; see Programme chart.
    Any changes carried out must be documented, e.g. handwritten into the
    corresponding column of the Programme chart.
    Changing the wash parameters
    The parameters for the cleaning programmes can be changed to suit particular
    requirements. For this purpose, the temperature and the temperature holding time
    in the Main wash 1 and 2 and Final rinse 1 and 2 programme blocks can be
    changed.
    The programme parameters for the Main wash 1 programme block in the
    SPECIAL 93°C-10' programme cannot be changed.
    Temperatures over 55°C result in blood denaturation and fixing.
    On the varioTD programme, take note of the appropriate requirements for
    infection prevention.
    Values which can be set for the temperature:
    30°C - 93°C in Programmes A to G and varioTD in 1°C increments
    Values which can be set for the holding time:
    1-15 mins.
    Changing the drying parameters
    The parameters for drying can be changed to suit particular requirements. The
    temperature and drying time can both be changed in programme blocks Drying 1
    (if available) and 2.
    Values which can be set for the temperature:
    50°C - 99°C in 1°C increments
    Values which can be set for the drying time:
    Drying 1: 1-99 minutes in 1 minute steps
    Drying 2: 5-95 min in 5 minute steps
    The drying time for programme block Drying 2 can also be altered using the0
    button (see "Operation, Selecting the additional Drying0function").
    Programmable functions
    39 
    						
    							Changing the temperature
    Step Display / Result
    ^Press button0as often as necessary or hold
    pressed in until the level of the programme block
    you wish to change appears in the display.E09for Main wash 1
    E12for Main wash 2
    E22for Final rinse 1
    E25for Final rinse 2
    E28for Drying 1
    E31for Drying 2
    ^Turn the programme selector to the programme you
    wish to change, e.g.A.°C~75
    ^Press
    						
    All Miele manuals Comments (8)