    							Description of the machine
    This washer-disinfector can be set up
    or retro-fitted to use the OXIVARIO
    programme by adding an additional
    dispenser pump together with a buffer
    tank for hydrogen peroxide H
    2O2solution. The DOS 2 dispenser is used
    for the H
    2O2solution.
    Two special programmes are available
    for this cleaning process, OXIVARIO
    PLUS and OXIVARIO. The machine is
    supplied with these two programmes
    allocated to programme places 51 and
    52.
    The cleaning process releases active
    oxygen under alkaline conditions. The
    cleaning agent used must be tenside
    free and have a pH value of between
    11 and 11.5.Areas of application
    The OXIVARIO process uses an
    alkaline cleaning agent and is
    recommended for processing surgical
    instruments where particularly stringent
    levels of cleanliness are required.
    It is particulary suitable for surgical
    instruments such as those used in HF
    (high frequency) surgery, orthopaedic
    surgery and for instruments which have
    dried out because of the length of time
    that has elapsed between using them
    and reprocessing them. It is also
    suitable for antiseptic circumstances.
    The cleaning process is suitable for the
    careful handling of minimally invasive
    instruments, including fibre optics,
    where the manufacturer has declared
    that they are suitable for processing
    with an alkaline cleaning agent.
    It is not suitable for anodised
    aluminium.
    Be aware that items made from alloys
    containing titanium such as some
    implants, do not always state the
    material's compatibility. Processing
    such items can result in colour
    changes, rendering any coding on
    them ineffective. If in doubt please
    consult the manufacturer of the item.
    Due to the high efficacy of the
    Oxivario process it is important to
    lubricate joints and contact parts on
    instruments pre sterilisation. Please
    refer to the instrument manufacturer
    for guidance.
    Special features - OXIVARIO
    61 
    						
    							Warning and Safety
    instructions
    These warning and safety
    instructions are in addition to those
    given at the beginning of this
    booklet.
    H2O2solution must only be used in
    the special containers and adapters
    supplied by the Ecolab and Dr. Weigert
    companies.
    Important: observe the
    manufacturer's safety instructions as
    given on their safety data sheet.
    Be especially careful when handling
    H
    2O2solution. It is an irritant chemical.
    Observe all current safety instructions
    pertaining to its use.
    Wear protective gloves and goggles.
    Empty containers must be disposed
    of in accordance with the
    manufacturer's instructions.
    H2O2solution must not be mixed
    with other chemicals. This could cause
    a serious chemical reaction, e. g. the
    release of harmful vapours or gases.
    Connecting the H2O2solution
    container
    The connection hose for the H2O2solution container is identified by a
    black label. It is supplied without an
    adapter because the system for
    drawing the solution out of the
    container varies depending on supplier.
    Connect the supplier's adapter to the
    connection hose (black).
    Then connect this to the H
    2O2solution container.
    Then start the DOS2-FILL
    programme.
    Unlike with other chemical agents, the
    H
    2O2solution container must be fully
    emptied before it is exchanged for a
    new one.
    Do not replace the H
    2O2solution
    container with a new one until the FILL
    DOS 2 CONTAINER message is
    displayed. After connecting the new
    container you must run the DOS2-FILL
    programme.
    If the message CHECK DISPENSING
    SYSTEM 2 appears in the display you
    must check the container and all
    dispensing tubes and hoses. The
    programme will stop automatically.
    Special features - OXIVARIO
    62 
    						
    							Electrical and electronic machines
    often contain materials which, if
    handled or disposed of incorrectly,
    could be potentially hazardous to
    human health and to the environment.
    They are, however, essential for the
    correct functioning of your or machine.
    Please do not therefore dispose of it
    with your general waste.
    Please dispose of it at your local
    community waste collection / recycling
    centre or contact your dealer for
    advice. Ensure that it presents no
    danger to children while being stored
    for disposal.