Home > Honeywell > Detector > Honeywell Lr953291 Manual

Honeywell Lr953291 Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Honeywell Lr953291 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 178 Honeywell manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							 11 EN1C-0015SZ20 R0314
    Fig. 10_3. Wiring diagram for C7061A1020-79/C7061F1003 detec-tor with 7800 SERIES Flame Safeguard controls 
    with shutter drive circuitry.Fig.10_4. Wiring diagram for C7061A1020-79/C7061F1003 detec-tor with R4348B Flame Safeguard controls with 
    shutter drive circuitry.     
    						
    							EN1C-0015SZ20 R031412
    ADJUSTMENTS AND CHECKOUT
    Adjust Detector Sighting
    With  the flame detector  installed  and  the  burner  running, 
    adjust the sighting position of the detector for optimum flame 
    signal. It is suggested that a volt-Ohm meter with a minimum 
    sensitivity of one megohm/volt and a zero to five or ten Vdc 
    scale  be  used  for  R7861 amplifier  flame signal  measure-
    ments.
    Measure the flame signal as illustrated in Fig. 11. Be care-
    ful  to  make  the  proper  connections  of  positive  (red)  meter 
    lead  to  positive  (+)  control  jack  and  negative  (black)  meter 
    lead  to  negative  (-)  or  (-Com)  jack  with  7800  SERIES  con-
    trols. When the 7800 SERIES control has the keyboard dis-
    play  module,  a  zero  to five Vdc  voltage  is  displayed  on  the 
    module.
    NOTE:  The flame Signal m ust be steady.
    Use a W136 test meter (or similar) when measur-
    ing the current on an R7061 amplifier . 
    Move the detector and sight pipe around to sight the  flame 
    from  various  positions  and  angles. Try  to  get  a  maximum 
    steady (or stable) reading on the meter. The signal must be 
    above the minimum acceptable voltage listed in Table 1.
    Measure the  flame signal for the pilot alone, the main burner 
    flame  alone, and both together (unless monitoring only the 
    pilot  flame  when using an intermittent pilot, or only the main 
    burner  flame when  using  direct  spark  ignition).  Also  mea-
    sure the  flame signal at low and high  firing rates and while 
    modulating in between (as applicable). With the detector in 
    its  final  position,  all  required  flame signals  must  be  steady 
    (or stable) and as  specified in Table 1. If you cannot obtain 
    the proper signal, refer to the Troubleshooting section.
    Pilot Turndown Test
    If the detector is used to prove a pilot  flame before the main 
    fuel  valve(s)  can  be  opened,  perform  a  pilot  turndown  test 
    before welding the sight pipe into position. Follow the proce-
    dures in the  flame safeguard control instructions and in the 
    burner manufacturer instructions.
    lgnition Spark Response Test
    Test to make certain that ignition spark is not actuating the 
    flame relay in the flame safeguard control.
    A  Close the pilot and main burner manual shutoff valves.
    B  Start the burner and run through the lgnition period. lgni-
    tion spark should occur, but the flame LED must not light. 
    The flame signal should not be greater than 0.25 Vdc.
    C  If the flame relay does pull in, reposition the detector far-
    ther  from  the  spark,  or  relocate/resight  the  detector  to 
    eliminate/reduce  the  detector  response  to reflected UV 
    radiation.  It  may  be  necessary  to  construct  a  barrier  to 
    block the ignition spark from the detector view. Continue 
    adjustments until the flame signal due to ignition spark is 
    less than the flame signal values indicated in step B.
    Response to other Ultraviolet Radiation Sources
    Some  sources  of  ar tificial light  produce  small  amounts  of 
    ultraviolet radiation. Under certain conditions, an ultraviolet 
    detector  responds  to  them  as  if  it  is  sensing  a  flame.  Do 
    not use an ar tificial light source to check the response of an 
    ultraviolet  flame detector. To check for proper detector oper-
    ation, conduct  flame failure response tests under all operat-
    ing conditions.
    Weld the Sight Pipe
    When  the  flame signal  is  acceptable  after  all  adjustments 
    are made, remove the detector and weld the sight pipe in its 
    final  position. (lf you are using a swivel mount, the pipe may 
    already be welded). Then reinstall the detector.
    Final Checkout
    Before putting the burner into service, check out the instal-
    lation  using  the  checkout  procedures  in  the  instructions  for 
    the  appropriate flame safeguard  control.  After  completing 
    the checkout, run the burner through at least one complete 
    cycle to verify correct operation.
    CAUTION
    Do  not  put  the  system  into  operation  until  all 
    checkout  tests  in  the  instructions  for  the  appro-
    priate flame safeguard  control  and  any  others 
    specified in  the  burner  installation  instructions 
    are satisfactorily completed.
    Fig. 11.  Measuring  voltage flame Signal  with  7800  SERIES 
    controls
     Flame  Amplifier min. acceptable  max. acceptable
      detector   steady signal  steady signal
    C7061A1079 R7061 3,0 µA  ——
    C7061A1020
    C7061F1003  R7861 
    1,25 VdC  5,0 VdC
    Table 1. Flame signal    
    						
    							 13 EN1C-0015SZ20 R0314
    VARIOUS
    TROUBLESHOOTING
    CAUTION
    1. Be  extremely  careful  while  troubleshooting 
    the detector; line voltage is present on some 
    of the terminals when power is on.
    2.  Open the master switch to disconnect power 
    before  removing  or  installing  the  detector  or 
    its  cover. More  than  one  disconnect  may  be 
    involved.
    Equipment Required
    A volt-Ohm meter with a minimum sensitivity of one megohm 
    / Volt and a zero to five or ten Vdc scale is suggested. When 
    the keyboard display module is available, a flame signal dis-
    plays on the module. For revision of a flame safeguard con-
    trol using the R7061 amplifier, use the W136 or a similar test 
    meter. For replacement parts, see specifications section.
    UV Sensor Tube Test
    see UV sensor tube test section.
    Unsatisfactory Flame Signal
    If  a  satisfactory  flame signal  (see  Table  1)  cannot  be 
    obtained  while  adjusting  the  sighting  position  of  the  detec-
    tor,  follow  these  procedures.  If  you  encounter  other  prob-
    lems  in  the  system,  refer  to  the Troubleshooting  section  in 
    the instructions for the appropriate  flame safeguard control.
    NOTE:  For  instructions  for  replacing  the  viewing  window,  sensing  tube,  and  coil  and  shutter  assembly,  see 
    the service section.
    Troubleshooting Procedures
    First  perform  the  preliminary  inspection.  Then  follow  the 
    applicable  procedures  for  either  a  low  reading  or  a  zero 
    reading  on  the  meter.  After  reinstalling  the  detector  or 
    replacing  its  cover,  recheck  the  meter  reading.  To  try  to 
    obtain  the  proper flame signal,  adjust  the  position  of  the 
    detector.  If  you  complete  all  of  the  procedures  and  still 
    cannot obtain a proper flame signal, replace the detector.
    Preliminary lnspection
    A Check for the proper line voltage. Make sure the master 
    switch  is  closed,  connections  are  correct,  and  power 
    supply is of the correct voltage and frequency.
    B  Check the detector wiring for defects:
    a. lncorrect connections.
    b.  Wrong type or size of wire.
    c.  Deteriorated wire.
    d.  Open circuits.
    e.  Short circuits.
    f.  Leakage paths caused by moisture, soot, or dirt.
    C  With  the  burner  running,  check  the  temperature  at  the 
    detector. If it exceeds 79°C (175°F):
    a. Add additional insulation between the wall of the com-
    bustion chamber and the detector.
    b.  Add  a  shield  or  screen  to reflect radiated  heat  away 
    from the detector, or
    c.  Add cooling (refer to sight pipe ventilation and acces-
    sories sections).
    Removing the Detector from the Sight Pipe
    C7061A:
    Loosen  the  three  screws  holding  the  mounting flange 
    together; rotate the detector slightly so the screws clear the 
    slots in the back section of the flange; separate the flange; 
    and pull off the back section (with the UV Sensor).
    NOTE:  The  detector  will  be  free  as  soon  as  the  collar  is 
    unscrewed; do not drop it.
    C7061F:
    Carefull unscrew from sight pipe
    Procedure for a Zero Meter Reading
    A  Replace  the  plug-in amplifier.  Then  recheck  the flame 
    signal.
    B  Replace  the  ultraviolet  sensing  tube  (see  Service  sec-
    tion). Then recheck the flame signal.
    C  Replace  the  coil  and  shutter  assembly  (see  Service  below). Then recheck the flame signal.
    D  If  you  still  cannot  obtain  a  meter  reading,  replace  the 
    detector.
    IMPORTANT
    At the completion of troubleshooting, be sure to 
    perform  the  adjustments  and  checkout  proce-
    dures.   
    						
    							EN1C-0015SZ20 R031414
    SERVICE
    CAUTION
    Open  the  master  switch  to  disconnect  power 
    before  removing  or  installing  the  detector  or 
    its  cover.  More  than  one  disconnect  can  be 
    involved.
    Periodic Maintenance
    A Clean  the  viewing  window  (or  focusing  lens)  when  nec-
    essary. Remove  the  detector  (see  Troubleshooting  sec-
    tion)  and  use  a  clean  cloth  over  the  eraser  end  of  a 
    pencil. Do not remove the window (or lens) to clean it. If 
    it is broken or damaged or it is coated with a substance 
    that cannot be cleaned off, replace it (see Fig. 14).
    B  Keep the flame detection system adjusted for the smooth-
    est,  most  reliable  operation  as  recommended  by  the 
    burner manufacturer.
    C  Replace  the  sensing  tube,  coil  and  shutter  assembly,  or 
    viewing  window  only  when  necessary  to  obtain  proper 
    operation.
    Removing the Detector Cover
    Open the Master Switch
    C7061A:  Unscrew  the  four  captive  cover  screws  (Fig.  8 
    and 9) and carefully slide off the cover.
    NOTE: These bolts are removable. Put them in a 
    safe place to avoid losing them.
    C7061F  Carefully  unscrew  the  cover  from  the  detector 
    cap.
    Replacing the Ultraviolet Sensing Tube
    A Open  the  master  switch  and  remove  the  cover  from  the 
    detector (see instructions above).
    B  Locate the UV sensing tube.
    IMPORTANT
    Be very careful not to kink or otherwise damage 
    the flexible shutter.
    C   Gently bend the alignment guid just enough to free the tip 
    of the tube.
    D  Insert  a  screwdriver  between  the  tube  base  and  socket, 
    and gently pry the tube out of its socket.
    E  Pull the tube out of its socket.
    F  Insert  the  new  tube  through  the  openings  in  the  shutter 
    assembly.
    G  Align the three pins on the new tube with the holes in the 
    socket.
    H  Carefully  push  the  new  tube firmly  into  the  socket;  the 
    alignment guide will snap into place around the tip of the 
    tube.
    I  Make sure the new UV sensor tube is seated securely.
    J  Replace the detector cover.
    Replacing the Coil and Shutter Assembly
    NOTE:  Use only a 190971 B coil and shutter assembly.
    A  Open  the  master  switch  and  remove  the  detector  cover 
    (see Removing the Detector Cover section).
    B  Remove  the  ultraviolet  sensing  tube  (steps  A  through  E 
    of Replacing the Ultraviolet Sensing Tube section).
    C  Cut  the  white  wires  as  close  as  possible  to  the  crimped 
    connectors, and remove the crimped connectors.
    D  Remove  the  three  mounting  screws  from  the  base  of 
    the  coil  and  shutter  assembly. Put  the  screws  in  a  safe 
    place.
    E  Remove the coil and shutter assembly.
    F  Put the new coil and shutter assembly into place.
    IMPORTANT
    Be very careful not to kink or otherwise damage 
    the flexible shutter.
    G  Insert the three mounting screws into the base of the coil 
    and shutter assembly and tighten securely.
    H  Remove sufficient insulation  from  each  of  the  two  white 
    lead wires remaining on the detector, and also from each 
    of the two white leadwires on the new coil.
    I  Using  solderless  connectors,  connect  one  of  the  coil 
    wires  to  one  of  the  remaining  white  leadwires. Connect 
    the other coil wire to the other remaining white leadwire.
    J  Reinstall  the  sensing  tube  (steps  F  through  I  of  Replac-
    ing the Ultraviolet Sensing Tube section).
    K  Replace the detector cover.
    Fig. 12. Replacing ultraviolet radiation sensing tube.      
    						
    							 15 EN1C-0015SZ20 R0314
    Replacing the Quartz Viewing Window (or Focus-
    ing Lens) C7061A Only
    IMPORTANT
    A quartz window or lens must be used. Ordinary 
    glass absorbs or filters out ultraviolet radiation.
    A  Open  the  master  switch;  remove  the  detector  from  the 
    sight pipe and remove the detector cover. (see appropri-
    ate sections.)
    B  Remove  the  ultraviolet  sensing  tube  (steps  A  through  E 
    of Replacing the Ultraviolet Sensing Tube section).
    C  Loosen the three screws holding the back section of the 
    mounting flange to  the  faceplate. Carefully  remove  and 
    keep together the three screws, the gray asbestos- neo-
    prene gasket, the red rubber washer, and the back sec-
    tion of the mounting flange.
    NOTE:  if the viewing window (with its rubber mounting gas-
    kets) is stuck to the mounting flange. skip step D.
    IMPORTANT 
    Be very careful not to kink or otherwise damage 
    the flexible shutter.
    D  Using  the  eraser  end  of  a  pencil,  push  out  the  viewing 
    window (with its rubber mounting gaskets) from the inside 
    of the faceplate.
    E   Insert one rubber mounting gasket into the window aper-
    ture in the faceplate.
    F  Insert the new quartz viewing window (or focusing lens).
    G  Insert  two  rubber  mounting  gaskets  (only  one  gasket 
    when replacing a focusing lens) into the aperture.
    H  Put  the  back  section  of  the  mounting flange,  rubber 
    washer,  and  asbestos-neoprene  gasket  in  place  on 
    the  faceplate,  and  securely  tighten  the  three  mounting 
    screws.
    NOTE:  Make sure the red rubber washer between the gray 
    fiber-neoprene gasket  and  the  back  section  of  the 
    mounting flange does not protrude over the window 
    aperture  or  otherwise  obscure  the  detector  line-of-
    sight. I 
    Clean  the  viewing  window  (or  focusing  lens)  on  both 
    sides using a clean cloth placed over the eraser end of a 
    pencil.
    J  Reinstall  the  sensing  tube  (steps  F  through  I  of  Replac-
    ing the Ultraviolet Sensing Tube section).
    K  Replace the detector cover and reinstall the detector on 
    the sight pipe.
    Replacing Focusing Lens C7061F Only
    IMPORTANT
    A quartz window or lens must be used. Ordinary 
    glass absorbs or filters out ultraviolet radiation.
    A  Remove the detector from the sight pipe (see Fig. 15).
    B  Remove  the  3  screws  from  the  threaded flange (pipe 
    connection) which houses the 124204 Focusing Lens.
    C  Replace the lens and tighten the 3 screws. 
    Fig. 13. Replacing coil and shutter assembly.
    Fig. 14. Replacing quartz viewing window or focusing lens
    Fig. 15. Replacing focusing lens on C7061F1003          
    						
    							(2)
    (1)Equipment andprotectivesystemsintended foruse inpotentially explosiveatmospheres
    Directive 94/9/EC
    EC-TYPE EXAMINATION CERTIFICATE
    (3)NumberoftheEC type examination certificate:INERIS03ATEX0158X
    (4)Protective systemorequipment :
    BOX TYPE AD4320001
    orAD432002
    (5)
    (6)
    Manufacturer:
    Address:IDRM
    Z.I duSauvoy
    25, avenue deMontboulon
    F - 77165 SAINT -SOUPPLETS
    (7) Thisprotective system
    orequipment and anyotherac;ceptablealternativeofthisone aredescribedintheannexof
    this certificate andthedescriptive documentsquotedinthis annex.
    (8) TheINERIS, notifiedbodyandidentified undernumber 0080,inaccordance witharticle 9
    ofCouncil Directive
    94/9/ECofthe23rdMarch1994,certifies thatthisprotective systemorequipment fulfilstheEssentialofHealth
    and Safety Requirements relatingtothe design. andconstruction
    ofequipment andprotective systemsintended
    for use inpotentially explosiveatmospheres, describedinappendix II
    oftheDirective.
    The examinations andthetests areconsigned inofficial reportN°p42448/03.
    (9) Therespect
    oftheEssential HealthandSafety Requirementsisensured by:
    - conformity with:
    EN50014ofJune1997+AlandA2
    EN50018ofNovember2000+Al
    EN50281-1-1ofSeptember1998+Al
    -specific solutions adoptedbythe manufacturer tomeet theEssential HealthandSafety Requirements
    described inthe descriptive documents.
    (10) Sign
    X,whenitisplaced following theNumberofthe EC type examination certificate,indicatesthatthis
    equipment andprotective system
    issubjectedtothe special conditions forsafe use,mentionedintheannexof
    this certificate.
    This document mustnotbereproduced otherthaninitsentirety
    PareTeehnologique AlataBP2F-60550Verneuil-en-Halatte
    H~I+33(0)3445566 77fax+33(0)344556699internetwww.ineris.fr
    InstitutnationaldeIenvironnementindustrieletdesrisques
    Etablissement publicacaractereindustrieletcommercial -ReSSenlisB381984921-Siret38198492100019-APE7438
    Folio 115
    EN1C-0015SZ20 R0314 
    						
    							EC-Type Examination CertificateN°INERlS03ATEXO158X
    (11) This
    ECtypeexamination certificaterelatesonlytothe design, examination andtestsofthespecified equipment
    or protective systeminaccordance tothe directive 94/9/EC. Furtherrequirements
    oftheDirective applytothe
    manufacturing processandsupply
    ofthisequipment orprotective system,thesearenot covered bythis
    certificate.
    (12) Themarking
    oftheequipment orthe protective systemwillhavetocontain:
    \Q1I2GD
    X.LEFEBVRE
    Engineer atthe Laboratory
    ofCertificationofATEX
    Equipment
    EEx dlICT6orEEx dlICT5
    TIOOoCorT135°C IP65
    Vemeuil-en-Halatte, 20031028
    DirectoroftheCertifying Body,
    By delegation
    C.MICHOT
    Certification Manager
    Thisdocumentmustnotbereproduced otherthaninitsentirety, includedannexesFolio2/5
    EN1C-0015SZ20 R0314 
    						
    							EC-TypeExaminationCertificateN°INERIS 03ATEXO158X
    (13)ANNEX
    (14)ECTYPEEXAMINATIONCERTIFICATEN°INERIS03ATEX0158X
    (15)DESCRIPTIONOFTHEEQUIPMENTORTHEPROTECTIVESYSTEM
    Thisboxconsistsinabodyclosedbyascrewedcover.
    Theenclosurecontainsawholeofapparatusdefinedinthedescriptive
    note.
    ThisboxcanbefittedwiththefollowingExcomponents:
    -PneumaticbushingtypeXW8-TL1certificateINERIS03ATEX9014U
    -AuxiliarycontroltypeTCDcertificateLCIE02ATEX0036U
    ThisapparatuspresentsthedegreesofprotectionIP65accordingto
    EuropeanstandardEN60529.
    PARAMETERSRELATINGTOTHESAFETY
    Supplyvoltage1000V
    Maximum
    powerdissipated
    Boxtypeandmaximumpowerdisspa'!:~~TemperatureclassAD432001AD432002
    T6ouT85°C7watts12watts
    T5ouT100°C10watts18watts
    Thisdocumentmustnotbereproducedotherthaninitsentirety,includedannexesFolio 3I 5
    EN1C-0015SZ20 R0314 
    						
    							EC-Type Examination CertificateN°INERIS 03ATEXO158X
    MARKING
    Markingmustbereadableandindelible
    indications:
    IDRM
    F-77165SAINT-SOUPPLETS
    AD432001orAD432002
    INERIS03ATEX0158X
    (numerodeserie)
    (Anneedeconstruction)
    ~II2GD
    EExdIICT(*)
    IP65T(*)
    DONOTOPENWHENENERGIZED
    (*)seetablebelow
    itmustcomprisethefollowing
    Incaseoftheboxisfittedwithadisconnector,theindicationnearto
    thedisconnector:
    DONOTOPERATEUNDERLOAD
    Thewholemarkingcanbecarriedoutinthelanguageofthecountryof
    use.
    Theprotectiveapparatusorsystemmustalsocarrythemarkingnormally
    envisagedbythestandardsofconstructionwhichrelatetoit.
    ROUTINEEXAMINATIONSANDTESTS
    Accordingto16.2ofstandardEN50018,theapparatusisexemptedof
    routinetestinviewofthefactthathasundergoneastatictypetest
    at4timesthereferencepressureunder32bar.
    (16)DESCRIPTIVEDOCUMENTS
    Thetechnicalreportiscomposed
    constitutingthedescriptivefile
    certificate.
    ofthe
    ofthe
    documentquotedhereafter,
    apparatus,objectofthis
    Certificationfile10-210rev.Cof2003.10.06
    Thisdocument,signedon2003.10.06included4parts(20pages).
    Thisdocument mustnotbereproduced otherthaninitsentirety, includedannexesFolio4/5
    EN1C-0015SZ20 R0314 
    						
    							EC-TypeExamination CertificateN°INERlS 03ATEX0158X
    (17)SPECIAL CONDITIONSFORSAFEUSE
    ThecableentrieshavetogetadegreesofprotectionatleastIP65,
    Foruseinpotentiallyexplosiveatmospheresduetocombustibledust:
    Thesurfaceofthedifferentsgapshavetobecoveredwithgrease,
    Userhastodoaregularcleaningofmaterialtolimitdustlayerson
    thematerial.sides.
    (18)ESSENTIALREQUIREMENTSOFSAFETYANDHEALTH
    TherespectoftheEssential.HealthandSafetyRequirementsisensuredby:
    conformitytotheEuropeanstandardsEN50014,50018andEN50281-1-1.
    thewholeoftheprovisionsadoptedbythemanufactureranddescribed
    inthedescriptivedocuments.
    Thisdocumentmustnotbereproduced otherthaninitsentirety, included annexesFolio5/5
    EN1C-0015SZ20 R0314 
    						
    All Honeywell manuals Comments (0)