    							Hitachi Double Beam 
    Spectrophotometer 
    						
    							1
    Model 124 
    Model U-2000
    Spectrophotometer simpler to use and 
    higher in dependability thanks to mounting
    of a large-size color LCD.  
    High resolution satisfying European Pharmacopoeia (spectral bandpass: 1.5nm).  
    Trace-amount measurement in biotechnological field, etc. can be carried out using optional 50, 25
    and 5mL micro-volume cells. 
    ■An abundance of optional accessories for various applications have been lined up.  
    ■Function for automatic data storage in a USB memory is incorporated.  
    ■Validation function and self-diagnostic function are standard with this instrument.   
    ■The UV Solutions control program available at option enables the user to control 
    operations from a personal computer.   
    (The Model U-2910 is a special
    model for PC control.) 
    *:  Available at option 
    Model 100-50
    Model U-2001Model U-2800 Model 150-20 
    Spectral bandpass 1.5nm satisfying European Pharmacopoeia 
    European Pharmacopoeia requires a ratio of 1.5 or larg-
    er when measuring 0.02% (V/V) solutions of toluene in
    hexane.  In the spectrum at right, a satisfactory value of
    1.6 can be confirmed.  
    (Sample: 0.02% solution of toluene in n-hexane) 
    2
    Stable optics due to double beam  
    Incorporation of stigmatic concave diffraction grating
    The optics of this instrument adopts the Seya-Namioka monochromator widespread as a representative concave
    diffraction grating monochromator.  
    Because a concave diffraction grating has both beam condensing and dispersing functions, an optical system can
    be configured with fewer mirrors.  In a spectrophotometer, use of fewer mirrors signifies a shorter optical path,
    thus giving rise to an aberration-free bright optics.  
    For elimination of the aberrations which were essentially unavoidable in the past, a stigmatic concave diffraction
    grating has been developed by applying Hitachi’s original technology.  As a result, a higher resolution has been
    realized. 
    Detector 1  Mirror D
    2 lamp
    WI lamp Concave diffraction
    grating
    Entrance 
    slit
    Exit slit 
    Filter
    Mirror
    MirrorHalf
    mirror
    Reference 
    side
    Sample 
    side
    MonochromatorSample 
    compartmentLens
    LensDetector 2
    In double beam design, the energy of the light source is divided into two with
    a half mirror so that one passes through the reference side, and the other
    through the sample side, which is unavailable with the single beam design.  
    Since the reference-side energy is also incident on a detector, photometry is
    carried out on the basis of this signal.  Therefore, an energy change in the
    light source can be compensated to ensure stable measurement for a long
    time. 
    A higher resolution has been achieved
    by eliminating coma from the Seya-
    Namioka monochromator which is the
    most popular concave diffraction grat-
    ing monochromator.  Its grooving is
    supported by the only ruling engine in
    Japan.  The diffraction gratings of
    Model U-2900/2910 have also been
    made with this machine.  
    						
    							3
    Mounting of 26.4cm color LCD, the largest 
    in this product class 
    Due to a 640 ×480 color LCD with backlight, working curves, spectra and other minute displays are easier to view, and charac-
    ters are bigger than before, which ensures a high visibility.  A dedicated keyboard (for U-2900 alone) is quite helpful for baseline
    measurement and other operations. 
    *  The size in the photo is 1/2 of the actual LCD size. 
    *  The photo size is 1/2 of the actual keyboard size. 
    4
    Introduction of Measurement Modes 
    Photometry
    The concentration of an unknown sample can be deter-
    mined 
    in comparison with the known concentration of a standard
    sample. 
    The linear/quadratic regression curve or polygonal line
    approximation curve is drawn by measuring plural standard
    samples (max. 20 samples) to calculate the concentration. 
    In addition, judgment of the upper and lower concentration
    limits can be set and quantitative analysis can also be car-
    ried out via factor input. 
    Wavelength scan
    A spectrum resulting from scanning over a desired wave-
    length range within 190 to 1,100nm can be displayed. 
    Because each substance has a unique spectrum, its char-
    acteristics can be examined. 
    After measurement, peak search, smoothing and other
    data processing can be effected.  And, a repeat scan capa-
    bility helps the user trace a chemical reaction process.
    Baseline correction is also available. 
    Time scan
    A spectrum can be drawn by following a change of photo-
    metric value at a single wavelength with the lapse of time.
    This function is used for analysis of enzymatic reactions.
    Enzyme activity is measured according to a change of
    absorbance within the set time period.  Peak detection,
    smoothing, kinetic assay and other data processing can be
    performed. 
    Multiple-wavelength measurement
    Purity of a nucleic acid can be calculated using a ratio of
    absorbance values at 2 wavelengths (A260/A280). 
    Furthermore, measurement can be performed while auto-
    matically shifting measuring wavelength to a maximum of 6
    wavelengths.  This is convenient when only measured data
    at each wavelength is desired.  
    						
    							6
    Easy to Save Measured Data.  
    Measured data can be saved in a commercially available USB flash memory*. 
    The data is savable in the text file format. 
    A commercial spreadsheet program can processing the measured data. 
    * Not all of the commercial USB flash memory products are usable.  Contact us separately. 
    Data saving with USB flash memory
    Data can be transferred easily to PC by means of a USB flash memory. 
    Data loading
    Because the measured data is saved in the text file format,
    it is sent directly to a commercial spreadsheet program. 
    This function can be utilized for detailed data analysis,
    report generation, etc.
    Printout of data from 
    spectrophotometer
    The data measured with the spectrophotometer including
    spectra and working curve graph can be printed with a
    small-sized thermal printer (DPU) separately available.
    * Contact us separately.
    5
    Highly dependable system incorporating auto-
    matic calibration and self-diagnostic function. 
    A highly dependable system was designed by incorporating lots of functions for memory
    check, wavelength drive check, lamp ignition check, automatic wavelength calibration, lamp
    ignition time display, etc. 
    (This function is standard with the Model U-2900 main unit.  The Model U-2910 provides the equivalent functions via PC control.) 
    Validation function
    This is a function for checking the performance of Model
    U-2900/2910 main unit.
    The following items can be checked. 
    ●Wavelength accuracy ●Baseline flatness 
    ●Wavelength setting repeatability ●Baseline stability 
    ●Bandpass ●Noise level 
    Example printout 1 of validation function
    with dot impact printer (for 1 check item) 
    Example printout 2 of validation function with
    dot impact printer (covering all check items) 
    Automatic calibration and 
    self-diagnostic function
    Upon initialization, memory check, wavelength drive mech-
    anism check, lamp ignition check and 656.1nm check are
    automatically diagnosed and adjusted. 
    Also, the automatic wavelength calibration function is
    incorporated for automatic calibration by using the bright
    line of D
    2lamp as a reference wavelength. 
    Besides, each cumulative ignition time of the WI and D
    2
    lamps can be displayed.  This function is useful for instru-
    ment management.  
    						
    							8
    Instrument application further expandable
    by UV Solutions program. 
    Option package program (P/N 2J1-0311)
    Capable of performing color calculation and color difference calculation.
    Nucleic acid measurement program (P/N 2J1-0316)
    Usable for checking extraction and refinement of nucleic acids such as DNA/RNA essential for genetic research.  Data (photo-
    metric values at 230, 260 and 280nm), and calculation results (260/280, nucleic acid concentration, protein concentration,
    molar concentration) can be displayed collectively on the screen. 
    U-2910 is a special model for PC control. 
    GLP/GMP program (P/N 2J1-0317)
    The burden of optical instrument performance check can be alleviated. 
    Item-dependent measurement mode and automatic measurement mode (when no sample is necessary) have also been pre-
    pared.  In addition, a judgment tolerance range is settable so as to meet each analytical purpose. 
    Report generator program (P/N 2J1-0312)
    Developed to enable freely customizing the report output format of measurement results.  Each character size and position of
    report items and comment are settable and besides, mathematical calculations conventionally dependent on manual method
    can be executed by spreadsheet program functions. 
    UV Navigation program (P/N 2J1-0313)
    Provides support for automation of measurement, data processing and printout.
    Measuring procedure can be readily programmed using a graphical editor for imple-
    mentation of spectrophotometer control, measurement result indication and judg-
    ment, etc.  Because analytical procedure can be automated, it is possible to save
    labor and form working patterns.  This is effective for routine work such as quality
    control. 
    7
    Personal computer-based instrument 
    control and data processing.
    The Model U-2900 becomes controllable from PC when the
    UV Solutions program (P/N 2J1-0310) is installed.   Not only
    the U-2900 functions such as photometry and wavelength
    scan, but also data pasting to another application program,
    data transfer, report generation, etc. will be supported. 
    (The Model U-2910 is a special instrument for PC control. 
    A PC compatible with Windows
    ®XP needs to be prepared separately.) 
    Simple to use
    Buttons are laid out so as to follow the operating proce-
    dure.  Hence, operations can be performed smoothly.  
    A rich variety of data processing functions
    Spectrum can be expanded, contracted, smoothed, differ-
    entiated and integrated, and fundamental arithmetic calcu-
    lations are applicable between spectra. 
    Powerful support for report generation
    Measured data can be sent to Microsoft®Excel.  One click
    of the Report button suffices for data transmission. A spectrum can be pasted to another application program. 
    Print preview
    Before printing, the contents can be checked by the Print
    Preview command. 
    Measured data text convertible
    Besides conversion into
    the ASCII text or J-
    CAMP format, spectra
    can be saved in Metafile
    format. 
    Windows®XP and Microsoft®Excel are trademarks of Microsoft Corporation, U.S.A. 
    Microsoft®and Windows®are registered trademarks of U.S. Microsoft Corporation in USA
    and other countries.  
    						
    							10
    Micro cell holder  
    (P/N 122-0060)
    Suitable for measurement of trace samples in
    medical and biochemical fields. 
    Specifications (cell required separately) 
    Wavelength range 220 to 950nm 
    Cell mounting
    /demounting repeatabilityWithin ±0.3%T
    Baseline flatnessWithin ±0.005Abs 
    (50mL micro-volume cell used) 
    The following cells are usable for the above micro
    cell holder (P/N 122-0060). 
    Tandem cell holder 
    (P/N 210-2115)
    Up to three 10mm cells can be mounted on each
    of the sample and reference beam sides.  Sample
    temperature can be maintained at a constant level
    by circulating constant-temperature water through
    the cell holder section. 
    Specifications (circulating thermostatic chamber
    and cell not included in this product)
    Control temperature range 15 to 40°C 
    Temperature stability ±0.3°C 
    Five 10mm rectangular cells can be mounted on
    the sample beam side and a micro cell mask (P/N
    200-1537) can be inserted in each cell holder.
    (Cell and micro cell mask are not included in this
    product.)  It is recommended to prepare a set of
    five cells. 
    Recommended cell sets  
    124-0352 10mm quartz cell set (5 cells in set) 
    124-0378 10mm glass cell set (5 cells in set) 
    Four rectangular long path absorption cells can be
    mounted on the sample beam side and they can be
    changed over externally. 
    Specification  
    Cell length 100mm, 50mm to 10mm cells applicable 
    Film holder 
    (P/N 210-2112)
    Designed for measurement of film-like samples. 
    Specifications  
    Film frame Width 25mm, height 30 to 50mm 
    Beam aperture Width 10mm ×height 20mm 
    Polarizer holder  
    (P/N 210-2130)
    Used for linearly polarizing the sample beam to
    check its polarization characteristic and for meas-
    uring optical rotary power with a sample placed
    between the polarizer and analyzer. 
    Specifications  
    Wavelength range 400 to 750nm Sample thickness 0.5 to 5mm 
    Sample size Min. 12 ×25mm/max. 55 ×100mm 
    Mask for Micro cell
    To be inserted into a standard rectangular cell
    holder for measurement of a trace sample. 
    Optical path lengths measured at 21 points on cell
    using a three-dimensional measuring instrument
    are indicated down to the fourth decimal place
    (mm). 
    Communication cable  
    (P/N 121-1521)
    A computer and the spectrophotometer main unit
    need to be connected with this cable for computer
    control of measurement using the UV Solutions
    program. 
    Specifications  
    200-1537 Mask for Micro cell (1.5mm) 
    200-1538 Mask for Micro cell (1.2mm)
    Cell
    124-0357 Micro quartz cell, 10mm 
    200-0551 Black quartz micro cell, 10mm
    Rectangular long path cell holder 
    (P/N 210-2107)
    Specifications
    Cell capacity 8µL Optical path length 8mm (quartz flow cell used) 
    Baseline flatness±0.001Abs (200 to 350nm) 
    ±0.002Abs (190 to 850nm) 
    Part No. Part name CapacityOptical path length 130-0622 50 micro cell 50µL 10mm130-0623 25 micro cell 25µL 5mm
    130-0621 5 micro cell 5µL 0.5mm
    Conditions of applicable rectangular cell are
    given below.
    Optical path length 10, 20, 30, 40, 50, 100mm 
    Outer width 12.75nm
    4-Position rectangular 
    long path cell holder    
    (P/N 150-0940)
    Used for transmittance/absorbance measurement
    of such a solid sheet sample as glass filter. 
    Glass filter holder     
    (P/N 210-2109)
    Specifications  
    Sample thickness 0.5 to 5mm 
    Sample size 12 ×25mm to 55 ×100mm 
    Used for measurement with a cylindrical cell. 
    Cylindrical long path cell holder
    (P/N 210-2108)
    5-Position turretcell holder
    (P/N 210-2110)
    Examined 10mm 
    rectangular quartz cell
    (P/N 210-1462)
    9
    Diverse optional accessories selectable for specific
    applications such as multiple samples and trace sample. 
    Auto sipper   
    (P/N 2J1-0100)
    This computer-controlled sample sipper is provid-
    ed with a sample recovery function and other ver-
    satile functions.  In combination with an autosam-
    pler, this unit will further advance automation and
    labor saving in the preparatory stage.  This sipper
    cannot control temperature. 
    Specifications 
    Minimum sample volume 0.6mL Carryover 1% or less Cell capacity About 50µL Optical path length 10mm 
    Reference beam side 10mm rectangular cell mountable 
    Water circulated cell holder    
    (P/N 210-2111)
    Water from a thermostatic chamber is circulated
    through this cell holder to maintain a sample cell at
    a constant temperature. 
    Specifications(circulating thermostatic chamber and cell not included in this product)
    Operating temperature rangeFrom normal temperature to 40°C 
    Temperature stability Within ±0.3°C 
    Micro flow cell unit 
    (P/N 210-2113)
    Suitable for continuous measurement of a trace
    sample. 
    Specifications 
    Difference in capacity due to cell changeoverWithin ±0.5% (at 100%T) Applicable cell10mm cell (not included in this product) Settable temperature 20 to 40°C 
    Designed to have the same structure as the above
    auto sipper.  Because the flow cell section can be
    maintained at a constant temperature, exact con-
    trol is ensured. 
    Specifications(without temperature control on reference side) 
    Minimum sample volume 0.6mL Carryover 1% or less Cell capacity About 50µL Optical path length 10mm Settable temperature 20 to 40°C 
    Reference beam side 10mm rectangular cell mountable 
    Specifications
    Cell capacity 70µL Optical path length 10mm (quartz flow cell used) Withstand pressure Max. 0.1MPa 
    Connecting tubeTeflon tube. Outer diameter 2mm, 
    inner diameter 1mm 
    Flow cell unit  
    (P/N 210-2173)
    The inside of this cell is structured to minimize
    stagnation of liquid and adhesion of air bubbles. 
    Specifications
    Cell capacity 600µL Optical path length 5mm (quartz flow cell used) Withstand pressure Max. 0.1MPa 
    Connecting tube Teflon tube.  Outer diameter 4mm, 
    inner diameter 3mm 
    Reference beam side5mm rectangular cell (standard accessory) 
    LC flow cell unit   
    (P/N 210-2131)
    Exclusively used for a liquid chromatograph.  In
    wavelength scan, the baseline is always kept flat. 
    Electronic thermostatted
    cell holder 
    (P/N 131-0306/0307)
    Standard-equipped with a magnet stirrer to main-
    tain a constant sample temperature in each cell.
    Temperature can be indicated down to a minimum
    0.01°C scale.  Because of an electronic thermostat-
    ted type using forced air cooling, this cell holder is
    capable of quick heating and cooling.  A thermo-
    static chamber is unnecessary.  
    (Temperature control: S only) 
    Specifications
    Control temperature range10° to 60°C (settable in steps of 0.01°C) (normal temperature 25°C)Temperature setting accuracy Within ±2°C*(difference between set temperature and sample temperature) Temperature stability Within ±0.5°C*
    * Room temperature:  25°C, sample:  distilled water 
    In the measurement of protein or nucleic acid
    fusion, a sample temperature can be changed con-
    tinuously to determine variation in absorbance.
    Being of an electronic thermostatted type, this cell
    holder is capable of quick heating and cooling.
    Sample temperature can be increased and
    decreased isothermally.  In addition, the set tem-
    perature can be maintained evenly inside a cell,
    because a stirrer is provided. 
    (Temperature control: S and R) 
    Specifications
    Control temperature range0° to 100°C (settable in steps of 0.1°C)(normal temperature 25°C) Temperature setting accuracyWithin ±2°C* (difference between set temperature and sample temperature)Temperature stability Within ±0.5°C* 
    Equipped with isothermal temperature regulation function 
    * Room temperature:  25°C, sample:  distilled water, circulatory
    water temperature: 22°C, set temperature: 10 to 60°C  
    A circulating thermostatic chamber needs to be prepared separately.
    AS-1010 Auto sampler
    (P/N 2J1-0121/0122)
    In combination with the auto sipper or in flow
    injection analysis, this unit is used for multiple-
    sample measurement.  A suction needle can be
    moved in three directions X, Y and Z. 
    Six 10mm cells can be mounted on the sample
    beam side, and they can be changed over automat-
    ically at certain intervals.
    Specifications(sample tube not included)
    Outer diameter 15mm, height 105mm 
    Sample tube size (option required) 
    Outer diameter 12 mm, height 105 mm 
    6-cell positioner with
    temperature control
    (P/N 2J1-0103/0104)
    Auto sipper with 
    temperature  control 
    (P/N 2J1-0101)
    Electronic thermostatted
    cell holder 
    (Thermostatted water bath is need-
    ed separately)(P/N 131-0301/0302) 
    						
    							NOTICE: For proper operation, follow the instruction manual when using the instrument.
    Specifications in this catalog are subject to change with/or without notice, as Hitachi High-Technologies Corporation continues to develop the latest 
    technologies and products for our customers.
    Tokyo, Japan
    http://www.hitachi-hitec.com/science/
    24-14 Nishi-Shimbashi 1-chome, Minato-ku, Tokyo, 105-8717,  Japan
    Tel: +81-3-3504-7211   Fax: +81-3-3504-7302
    Printed in Japan (H) HTB-E036P2007.2
    
    						