Page 561
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 10 Periodic Maintenance 10-27 10-7-9-5 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-15 and Table 10- 16. Record all data on the PM Inspection Certificate. . CAUTIONEquipment damage possibility. Never switch the Polarity and the status of Neutral when the unit is powered ON. Be sure to turn the unit power OFF...
Page 562
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 10-28 Section 10-8 - When Theres Too Much Leakage Current... Section 10-8 When Theres Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent. Replace any defective part. Tighten all grounds. Ensure star washers are under all ground studs. Inspect wiring for bad crimps, poor connections, or damage....
Page 563
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 10 Periodic Maintenance 10-29 PM INSPECTION CERTIFICATE * Scan Format: Phased Array, Linear Array, Curved Array, Mechanical Array or Other FUNCTIONAL CHECKS PHYSICAL INSPECTION AND CLEANING COMMENTS: ______________________________________________________________________________ ______________________________________________________________________________...
Page 564
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 10-30 Section 10-8 - When Theres Too Much Leakage Current... ELECTRICAL SAFETY Final Check. All system covers are in place. System scans with all probes as expected. Accepted by: ______________________________________________________________________ Electrical Test PerformedMax Value AllowedValue Measured OK? Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source...
Page 565
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Index I A,B AC Power Replacement Procedure, 8-89 Acoustic Noise Output, 3-9 Acquisition Diagnostics, 7-14 Archiving Images Move, 4-18 Automatic Troubleshooting, 7-17 B/M Mode Processor (BMP) Board, 5-17 B/M-Mode Checks Functional Checks, 4-8 Back End Processor, 5-23 See BEP, 8-13 Theory, 5-23 Backup Patient Database, 4-15 Preset Configurations, 4-15 Base Load Software Check Devices, 3-31, , 8-76...
Page 566
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL II - EMI, 1-10 EMI Protection, 3-5 EQ Board Basic Functions, 5-15 General Description, 5-14 EQ Diagnostics Board Level, 7-25 Equalization (EQ) Board, 5-14 ESD, 1-10 Ethernet Signal, 5-49 Exam study display location, 4-7 External I/O, 5-46 Bi-directional Signals, 5-47 Input Signals, 5-47 Location in the Unit, 5-46 Output Signals, 5-49 Replacement Procedure, 8-126 F Fan Replacement Procedure, 8-122 Focal...
Page 567
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL III Operator I/O, 4-22 Mechanical Safety, 1-7 Media Formatting, 4-15 Memory Preferred Test Strategy, 7-37 Minimum Cable Configuration, 7-59 for System to Scan, 7-59 MOD Functonal Checks, 4-14 Models Covered, 1-2, , 9-2 Modem, 5-50 Inputs/Outputs, 5-50 LEDs, 5-50 Location in the Unit, 5-50 Monitor, 5-46 Adjustment, 6-2 Counter-balance Adjustment, 6-4 Tilt, 6-4 Tilt Adjustment, 6-4 Monitor Display...
Page 568
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL IV - Theory, 5-8 S Safety Considerations, 1-7 Scan Control Board High Level Features, 5-12 Scan Control Board (SCB), 5-11 SCB Diagnostics Board Level, 7-22 SCB LED, 5-12 Screen Captures, 7-4 Service Adjustments, 6-1 Service Manual Purpose of, 1-1 Shut Down Procedure, 4-3 Shutdown, 3-7 Side Covers Replacement Procedure, 8-5 Stand-by Mode, 7-1 System Does Not Boot, 7-51 System Maintenance, 10-7 System...
Page 570
GE Medical Systems GE Medical Systems: Telex 3797371 P.O. Box 414; Milwaukee, Wisconsin 53201, U.S.A. (Asia, Pacific, Latin America, North America) GE Ultraschall:Tel: +49 (0) 212 28 02 208 Deutschland GmbH & Co KG Beethovenstrabe 239, Postfach 11 05 60 D-42655 Solingen, German