    PMB TEMPERATURE CALIBRATION KIT 
     
     
    Part Number 3.07.6.6.10650,  Revision E, Feb 2015 
     
     
     
     
     
     
       
     
    Adam Equipment   
    C  O  P Y  
    						
    							© 2016 Adam Equipment 
        
    PMB TEMPERATURE CALIBRATION KIT 
     
    The temperature probe Serial No. AE xxxxx included in this kit has been calibrated to 
    an accuracy of +/- 1ºC 
     
    The Temperature measuring range is from -50ºC to 300ºC 
     
    Display resolution is 0.1ºC 
     
    Power is 1.5v DC LR44 button-cell type battery 
     
    Operating Instructions 
     
    ON/OFF: Press this button to power the probe up and down 
     
    ºC/ºF: Press this button to alternate between ºC and ºF 
     
    Temperature Calibration 
     
    1.  Place a sample pan on the pan support during this test.    
    2.  Place the temperature probe to be used to monitor the temperature within the weighing chamber on or near the sample pan.  
    3.  Close the chamber. 
    4.  Enter the menu for calibration section. Select the Temperature Calibration  option,  CAL TEMP . 
    5.  Press the  [Setup/Enter]  key. 
     
    The display will guide you through the steps necessary to calibrate the temperature 
    of the heating chamber. 
     
    When the analyzer chamber is closed the display will show  
     
     
     
     
     
    with the timer running. The timer will count up to 10:00 to allow time for the chamber 
    to come to a steady (room) temperature. During this time observe the temperature in 
    the chamber. 
     
    During this time the beeper will sound every minute to remind you that a calibration is 
    in progress.   
     
    Note the temperature inside the chamber. 
     
    When the temperature is stable and constant for some time press  [Setup/Enter] to 
    enter the lower temperature value. 
     
    The display will change to allow the value to be entered. 
     
     
     
     
     
    Using the numeric entry method, enter the temperature inside the chamber, for 
    example 23ºC you would enter:     FIRST 
      00:00 
       FIRST  
      000 
    C  O  P Y  
    						
    							© 2016 Adam Equipment 
        
     
     
     
     
     
    Press [Setup/Enter]  to continue the test.    
     
    The heaters will turn on and begin heating the test chamber. The timer will show the 
    elapsed time as before. The chamber will be heated to approximately 150ºC.   
     
    At the end of 30 minutes the beeper will sound, Press  [Setup/Enter] to enter the 
    second temperature value as displayed. 
     
    Enter this value as before pressing  [Setup/Enter] to confirm and store the value. 
     
     
     
     
     
    The PMB will now return to normal weighing. 
     
    External Temperature Probe Troubleshooting 
    If the probe displays  LL.L or H.HH , or does not turn on, try replacing the battery 
    inside the probe handle.If this still does not work then the probe may be defective. 
    Contact your distributor to discuss a replacement. 
     
      ################################################################### 
    EU Declaration of Conformity / Déclaration de conformité / Konformitätserklärung  Declaración de Conformidad / Dichiarazione di conformità  
     
     
     Adam Equipment Co. Ltd. 
    Maidstone Road, Kingston  Tel: +44 (0)  1908 274545 
    Milton Keynes, MK10 0BD  Fax: +44 (0)  1908 641339 
    United Kingdom  Email: [email protected] 
    www.adamequipment.com 
    We declare under our sole responsibility that the balance models listed below marked with 
    “CE” are in conformity with the directives and standards mentioned. 
    Nous déclarons sous notre seule responsabilité, que les types de balance cités ci-dessous 
    munis de la mention CE sont conformes aux directives et aux normes mentionnées ci-après. 
    Wir erklären in alleiniger Verantwortung, dass die untenstehenden Waagentypen 
    gekennzeichnet mit CE mit den genannten Richtlinien und Normen übereinstimmen. 
    Nosotros declaramos bajo responsabilidad exclusiva que los modelos de balanzas indicados a 
    continuación con el distintivo CE son conformes con las directivas y normas citadas. 
    Noi dichiariamo sotto nostra unica responsabilità che i tipi di bilance specificati di seguito  
    contrassegnati con la marcatura “CE” sono conformi alle direttive e norme citate. 
    Adam Equipment Digital Temperature Probe.  
    2014/30/EU EN61326-1:2013 – Part 1 
    2011/65/EC, RoHS 2 EN50581: 2012, 
    Signed for and on behalf of: Adam Equipment Company Ltd. 
    United Kingdom, 20 Feb 2016 
     
    Clive Jones, Quality & Product Manager. 
    ########################################################################################## 
       
       FIRST TEMP 
      023 
       SEC TEMP 
      XXX     
    C  O  P Y  
    						
    							© 2016 Adam Equipment 
        
     
    WEEE 2012/19/EU 
      
    This device may not be disposed of in domestic waste. This also applies to countries outside the EU, per 
    their specific requirements. Disposal of batteries (if fitted) must conform to local laws and restrictions. 
    Cet appareil ne peut être éliminé avec les déchets ménagers. L’élimination de la batterie doit être 
    effectuée conformément aux lois et restrictions locales. 
    Dieses Gerät nicht mit dem Hausmüll entsorgt. 
    Dispositivo no puede ser desechado junto con los residuos domésticos 
    Dispositivo non può essere smaltito nei rifiuti domestici. 
     
    FCC / IC CLASS A DIGITAL DEVICE EMC VERIFICATION STATEMENT  
    NOTE: This equipment has been tested and found to comply with the limits for a Class A 
    digital device, pursuant to Part 15 of the FCC rules and Canadian ICES-003/NMB-003 
    regulation. These limits are designed to provide reasonable protection against harmful 
    interference when the equipment is operated in a commercial environment.  This equipment 
    generates, uses and can radiate radio frequency energy and, if not installed and used in 
    accordance with the instruction manual, may cause harmful interference to radio 
    communications.  Operation of this equipment in a residential area is likely to cause harmful 
    interference in which case the user will be required to correct the interference at his own 
    expense. 
     
     
    ADAM EQUIPMENT  is an ISO 9001:2008 certified global company with more than 40 
    years’ experience in the production and sale of electronic weighing equipment. Adam products 
    are predominantly designed for the Laboratory, Educational, Health and Fitness, Retail and 
    Industrial Segments. 
     
    Our products include: Analytical and Precision Laboratory Balances, Compact and Portable 
    Balances, High Capacity Balances, Moisture analysers / balances, Mechanical Scales, Counting 
    Scales, Digital Weighing/Check-weighing Scales, High performance Platform Scales, Crane scales, 
    Mechanical and Digital Electronic Health and Fitness Scales, Retail Scales for Price computing 
     
     
    Adam Equipment Co. Ltd . 
    Maidstone Road, Kingston  Milton Keynes 
    MK10 0BD  UK 
    Phone:+44 (0)1908 274545 
    Fax: +44 (0)1908 641339 
    e-mail:  [email protected]  
       
    Adam Equipment Inc.  1, Fox Hollow Rd.  Oxford CT 
    06478 USA 
    Phone: +1 203 790 4774 
    Fax:  +1 203 792 3406 
    e-mail:  [email protected]     
    AE Adam GmbH.  Instenkamp 4 
    D-24242 Felde 
     
    Germany 
    Phone +49 (0)4340 40300 0 
    Fax:  +49 (0)4340 40300 20 
    e-mail:  [email protected]    
    Adam Equipment S.A. (Pty) Ltd. 
    7 Megawatt Road,  
    Spartan EXT 22  Kempton Park,  Johannesburg,  
    Republic of South Africa   
     
    Phone +27 (0)11 974 9745  Fax:  +27 (0)11 392 2587 
    e-mail:  [email protected]  
     
     
    Adam Equipment (S.E. ASIA) PTY Ltd  2/71 Tacoma Circuit 
    CANNING VALE 6155  Perth 
    Western Australia 
     
     
     
    Phone: +61 (0) 8 6461 6236  Fax +61 (0) 8 9456 4462 
    e-mail:  [email protected]  
       
    Adam Equipment (Wuhan) Co. Ltd.  
    A Building East Jianhua  
    Private Industrial Park   Zhuanyang Avenue  
    Wuhan Economic & Technological 
    Development Zone  430056 Wuhan   P.R.China  
    Phone: + 86 (27) 59420391  Fax + 86 (27) 59420388 
    e-mail:  [email protected]   
     
    © Copyright by Adam Equipment Co.  All rights reserved.  No part of this publication may be reprinted or translated in 
    any form or by any means without the prior permission of Adam Equipment. 
     
    Adam Equipment reserves the right to make changes to the technology, features, specifications and design of the equipment  without notice. 
     
    All information contained within this publication is to the best of our knowledge timely, complete and accurate when issued.  
    However, we are not responsible for misinterpretations which may result from the reading of this material. 
     
    For a complete listing of all Adam products and the latest version of this publication visit our website at: 
    www.adamequipment.com 
    C  O  P Y  
    						
    All Adam Equipment manuals Comments (0)

    Related Manuals for Adam Equipment PMB Moisture Analyser PMBTEMP User Manual