Adam Equipment Coin Counting Scale CCSArevC1 User Manual
Have a look at the manual Adam Equipment Coin Counting Scale CCSArevC1 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 35 Adam Equipment manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
31 | P a g e © Adam Equipment Company 2009 Place the calibration weight on the scale. Press th e [Total / ] key to show the ADC values for this weight. “ LoAd ” “ 10 ” “ 345689 ” Press the [Total / ] key to show the ADC values for this weight. The display will then show: “ SPAN ” “ PASS ” “ ” After few seconds it will show the next function. “ F5 ” “ ACCES ” “ CAL ” Remove the weight from the scale. If [CE/ESp] is pressed during calibration return to normal operation without doing the calibration. 5.4.5 F5- Access Code Number To change the user access code number used to enter the section of the program for changing coin weights (see Section 5.2) follow this procedure. This parameter can only be accessed by skipping the F3 Calibration function by using the [MC/ ] key. When the display is showing: “ F5 ” “ ACCES ” “ SEt ” Press the [Total / ] key. The display will show:
32 | P a g e © Adam Equipment Company 2009 “ Pin 1 ” “ ” “ ” Enter the new access code “ Pin 1 ” “ - - - -” “ ” press the [Total/ ] key to confirm and store the value, The scale will ask you to repeat the number again for securit y. “ Pin 2 ” “ ” “ ” Enter the new access code “ Pin 2 ” “ - - - -” “ ” Press [CE /Esc] to escape without changes before it is stored or press the [Total/ ] key to confirm and store the value. “ F5 ” “ ACCES ” “ Set ” To return to normal operation press the [CE /Esc] key.
33 | P a g e © Adam Equipment Company 2009 6.0 SECURE PARAMETERS 6.1 SETTING UP SECURED PARAMETERS The following parameters can only be accessed when the service access code 9999 is used. Turn the scale on while holding the [Total/ ] key down and then release the key. Key in the Service Access Code as described in sect ion 5.1.1. The first function shown is F1, the Baud Rate Setting. “ F1 ” “ bAud ” “ SEt ” This procedure gives access to the following parame ters: F1 Baud Rate Setting F2 Parity setting F3 Sleep F4 Weight Calibration F5 Access Code Number Setting F6 Symbol for Denomination Setting F7 Denomination Setting F8 Linearity Calibration F9 Factory Calibration Values See Section 5.4 for details of Functions F1 to F5. These parameters are set as in section 5.4. These parameters are only accessible when the service access code is used.
34 | P a g e © Adam Equipment Company 2009 6.1.1 F6 - SETTING ASCII VALUE FOR PRINTED LEGENDS Press the [MC/ ] key to scroll through the parameters. To change the symbol that is printed to represent t he value of the money used, follow this procedure. When the display is showing: “ F6 ” “ SIGN ” “ SEt ” Press the [Total/ ] key and the display will show: “ F6 ” “ 52H ” “ SEt ” Current symbol used in hex, 52H = R To enter the new ASCII value for the symbol, use th e [ ] key to select the digit to change (flashing) and then use [MC/ ] to increment the digit 0-9, A, B, C, D, E, F, 0. “ F6 ” “ 24H ” “ SEt ” New symbol used, 24H = $
35 | P a g e © Adam Equipment Company 2009 Press [CE /Esc] to escape without changes before it is stored or press the [Total / ] key to confirm and store the value. Currenc Symbo Value Rand R 52 Dollar $ 24 Other values for a particular symbol can be found i n a table of ASCII values. A sample table is attached.
36 | P a g e © Adam Equipment Company 2009 6.1.2 F7 - Denomination Values The denomination keys for South Africa have values pre-assigned, from 1c to R5 (0.05 to 5.00). Plus the 2 token keys can be re assigned also. In total 11 keys can be used for denominations or tokens. This procedure will allow you to change the denomination values associated with e ach key. It may be necessary to change keypad designs for other denomi nations. When the display is showing: “ F7 ” “ dEM ” “ SEt ” Press the [Total/ ] key. The display will show the decimal point position: “ Point ” “AddES ” “ 0.00 ” Decimal point current position, Press the [MC/ ] key. To select either 0.00, 0.0 or 0 Press [Total/ ] to set the decimal point. The displays will show: “ F7 ” “ dEM ” “ kEy ” Press the denomination key to be set. Use the key s for the tokens or for the denominations as needed. The displays will show:
37 | P a g e © Adam Equipment Company 2009 “ F7 ” “ K01 ” “ 1.00 ” Where the K01 is the key number for the key being u sed. The display also shows the denomination currently selec ted. “ F7 ” “ K08 ” “ 0.10 ” For example the number 8 key has a value of 0.10. To enter a new value for this key position simple u se the numeric keys and press [Total/ ] to save the new value. To set a key position for a token press the [MC/ ] to set T1, T2 etc. up to T8. The display will show “ SAVE” and then display: “ F7 ” “ dEM ” “ kEy ” You can select another key if necessary. When all changes have been made press the [CE /Esc] key to return to the parameters. Press [MC/ ] to advance to the next parameter. 6.1.3 F8- Linearity Calibration The display will show:
38 | P a g e © Adam Equipment Company 2009 “ F8” “ LinEr” “ SEt ” The linearity of the scale is set by calibrating w ith 2 masses, 1/2capacity (10kg) and full capacity (20kg). Press the [Total/ ] key to enter this function. The display will show the following: “ CAL ” “ 0.000 ” “ 12345 ” The number in the TOTAL window is the number of int ernal counts representing the zero weight, from the anal og to digital converter within the scale. Make sure no other weights are on the scale. Press the [Total / ] key to set the zero point. The display will show: “ LoAd 1 ” “ 10” “ 23456 ” Place 10kg mass on the scale and then press the [Total/ ] key The display will then show: “ LoAd ” “ 20” “ 23456 ” Place 20kg mass on the scale and then press the [Total/ ] key The display will then show: “ SPAN ” “ PASS ” “ ” After few seconds it will show the next function. “ F9 ” “ FACt ” “ SEt ”
39 | P a g e © Adam Equipment Company 2009 6.1.4 F9- Factory Calibration Value This calibration is stored in a separate location f rom the user calibration (F4). If a scale has been disrupted it is possible to rec all this calibration in order to have a scale that has the original calibration rest ored. This calibration can be done at any time without ch ecking for being within 10% of the previous factory calibrations. This val ue will also overwrite the user calibration value as set in F3. When the use r calibration is performed the results must agree within 10% of the factory ca libration. This prevents calibration with no weight on the scale or with gro ssly wrong weights. When the display is showing: “ F9 ” “ FACt” “ Set ” Press the [Total / ] key. The display will show: “ 0.000 ” “ ” “ xxxxxx” Make sure no other weights are on the scale. Press the [Total / ] key. The displays will show the calibration weight requi red, equal to the full scale capacity. “ 20.000 ” “ ” “ xxxxxx”
40 | P a g e © Adam Equipment Company 2009 Place the weight on the scale. When the scale is st able Press the [Total / ] key. The display will then show: “ SPAN ” “ PASS ” “ ” The scale will return to the Function menu. Press [CE /Esc] to return to normal operation or [MC/ ] to select a different function.