      FXi-08  HRA-08 FXi-08 
     HRA-08  
    Ethernet Interface
     
     
    When installing this Ethernet board (
    FXi-08  HRA-08
    ) into the A&D balances, such as 
    FX-i, FZ-i, 
    HR-A  HR-AZ
    , the balance can communicate with a computer through the LAN port and the 
    weighing data can be managed with the software "Wi n CT- Pl us".  
      
    Unpacking the Ethernet Board 
     
     
     
     
     
     
     
     
     
     
     
     
     
     
    Installing the Ethernet Board 
    Caution:
     Remove the AC adapter from the balance while 
    doing the following work. 
     
    Step 1
      Record the MAC address
    * (Hardware address) 
    printed on the Ethernet board (Shown in the 
    illustration). The MAC address is required when 
    specifying the IP address. 
     
    * Example: 00-20-4A-84-23-6A 
     
    Step 2
      Unfasten the two screws from the rear of the 
    balance and remove the panel. Peel the connector 
    off that is attached on the panel with double-side 
    adhesive tape. 
     
    Step 3
      Connect the connector to the Ethernet board. 
     
    Step 4
      Install the Ethernet board in the balance and 
    secure it using the two screws in step 2. 
     
     
     
     
     
    Connecting the Ethernet Board to a Computer 
    Caution 
      Contact the network manager before connecting the Ethernet board to a network.   
    A network error may occur due to the Ethernet board. A&D and dealers assume no 
    responsibility for errors.  Preset the 
    IP
     address and subnet mask into the Ethernet board one by one. The 
    IP
     address, 
    172.16.100.2, is set for all Ethernet boards at the factory. Do not duplicate an 
    IP
     address.   
     The 
    IP
     address can not be returned to factory setting automatically.   
    We recommend that you write the 
    IP
     address on the supplied label. 
    Step 1
      Connect the Ethernet board and the computer using a hub 
    and straight cables. (When connecting directly without 
    using hub, use a cross cable)   
    Step 2
      Reconnect the AC adapter to the balance.
     
    Step 3
     Preset the IP addresses and subnet mask of the Ethernet 
    board and the computer. Operate in accordance with the 
    procedure of [A&D WinCT-Plus]
     
     [Manual.PDF
    ] in the 
    CD-ROM. A communication error may occur with the 
    wrong operation. 
    Step 4
     Write the IP address on "
    IP address label
     and affix it 
    where you can see the address easily. 
    Step 5
      Install the data acquisition software RsMulti" into the 
    computer. Refer to [A&D WinCT-Plus]
      [Manual.PDF] of the 
    CD-ROM for this installation. 
     
    The Outline for Windows Communication Software 
      Run the data acquisition software "RsMulti". Menu: [Start]
      [Program]
      [A&D WinCT-Plus]
      [RsMulti]. 
    Refer to the [A&D WinCT-Plus]
      [Manual] for the operation. 
      The software can acquire data from multiple devices connected with a LAN or RS-232C.   The software can control these devices with commands.   The software can acquire data transmitted from devices.   Example
    : When pressing the 
     PRINT 
      key of the balance, data is output and is acquired with the computer. 
      The stored data can be used with the Microsoft EXCEL, when installing it before use.  8SWRZHLJKLQJLQVWUXPHQWVFDQEHFRQQHFWHG+RZHYHUWKHQXPEHURIFRQQHFWLRQVPD\EH
    XHQF\RIGDWDDFTXLVLWLRQ
    
    Example
        A sample window of the "RsMulti"
     
    Two balances are connected. 
     IP address  Subnet mask
    Computer 172.16.100.1FX-300i 172.16.100.2FX-3000i172.16.100.3255.255.0.0 
     
     
    FX
    i-08 / HRA-08
     
    Ethernet Interface board 
    IP address label (5 labels)
     
    Write the IP address on a label and use it for identification. 
    CD-ROMWindows communication software WinCT-Plus 
    IP address protective cover (5 covers)
     
    Affix a protective cover over the IP address label. 
    MAC address 
    Step 2
     
    Step 3
    Cable
    Ethernet board
    Step 4
     MAC address
    Panel 
    Connecto
    r
    Screw
    s
    Screw
    s
    172.16.100.1 172.16.100.2 172.16.100.3
     FX-300i FX-3000i  Compute
    r 
    Hub
     
    Straight cable×3
    AC adapte
    r
    St e p 2
    Hub
    St e p 1
    Compute
    r
    St e p 1
    St r a ig h t  c a b leCross cableRear of the 
    balance  
    						