3M CablePipeFault Locator 25502573 Series User Manual
Have a look at the manual 3M CablePipeFault Locator 25502573 Series User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 307 3M manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
78-9000-0192-2 Rev D 31 E. Directional Null (DirNull) TRACE [2] + Mode [SK] + DirNull [SK Toggle] or Menu/OK [5] + Cable/Pipe [SK] + Mode [SK] + DirNull [SK Toggle] In Directional Null mode, as the operator approaches the cable or pipe, the numerical Signal Strength [9] ([8] for 2550 locator) will increase then fall sharply as the receiver crosses the target cable or pipe. The Bar Graph [11] ([10] for 2550 locator) fills from both sides toward the middle and the receiver speaker volume decreases. As the antenna moves off the target path, the Bar Graph [11] ([10] for 2550 locator) opens, the Signal Strength [9] ([8] for 2550 locator) increases, and the speaker volume increases. Gain adjust is automatic in Directional Null mode. The center of the DirNull display provides a ‘compass view’ of the target path. An arrow will point toward the location of the cable/pipe in 45-degree steps. A solid line will appear over the cable/pipe, indicating its orientation to the receiver handle. For example: Figure 1: If the target path is to the right of the receiver, and running parallel to the operator, the right arrow will display. Figure 2: If the target path is to the left and in front of the operator (not running parallel to the receiver) the arrow will point toward the top left side of the display. Figure 3: When the receiver crosses the target path, a solid line will appear, instead of arrows, indicating the target path and its orientation to the receiver handle. Before marking target path, always use Directional Peak (Dir Pk), Special Peak (Spl Pk) or Trace View (T-View) mode to verify location.
32 78-9000-0192-2 Rev D F. Expanded Mode VOL [13] (press up arrow) When the third ring of the speaker icon is dotted or broken and ‘xpnd’ appears below the speaker icon as shown, the receiver is in “Expanded” mode. This mode is used for pinpointing a target cable or pipe. The area of response of the receiver narrows, allowing the locator to detect very small signal changes. Press the Speaker Volume Control [13] ([12] for 2550 locator) key to activate the “Expanded” mode. 8. Depth and Current Estimate Verifying the target path, depth and current can be helpful tools. • Pinpoint the cable or pipe being located. • Lower the tip of the receiver to the ground and press DEPTH [4]. − The depth to the target cable or pipe is displayed in the units specified in the receiver set up menu. The bold current reading is the relative current measurement. This reading can be compared to the relative current reading that alternately flashes with the frequency on the transmitter to help identify the target cable/pipe. The highest relative current value will be the target cable/pipe and will compare closest to the transmitter relative current value. The milliamp reading is an actual current measurement. − There are two options for measuring depth. Live depth is a continuous measurement. 1-Shot Depth is an averaging of the depth reading over a short period of time. When in 1-Shot mode, the unit will average the depth reading for three seconds, and then display the result on the display. Press DEPTH [4] to alternate between the two modes. − Five depth readings can be saved with the time, date and relative current measurements. • Pressing Mem Select [SK] and then Save [SK] will place each entry in sequential order in memory (M1–M5) until five readings have been stored. The unit will overwrite saved entries in excess of five, beginning with M1.
78-9000-0192-2 Rev D 33 • Optional step: Press Clear All [SK] to delete all stored depth information. − The operator may select a specific memory location to store the depth readings by pressing Mem Select [SK]. When the preferred location appears on the display, press Save [SK]. The display and memory location will populate with the current information. − Each memory location can be reviewed by pressing Mem Select [SK]. • Press TRACE [2] or DEPTH [4] to return to Locate mode, or wait until receiver automatically returns to the Locate mode. 9. Locating Frequencies A. Active Frequencies Active frequencies are trace signals supplied by a 3M™ Dynatel™ Transmitter 2500 Series (577 Hz, 1 kHz, 8 kHz, 33 kHz, 82 kHz or 200 kHz). • Select the same frequency that the transmitter is generating. • Press Menu/OK [5] (you may have to press Menu/OK [5] twice to toggle to the LOCATE MENU display). • Press Cable/Pipe [SK] • Press Freq [SK] • Press Active [SK Toggle] until the desired frequency is displayed: (“Active” 577, 1k, 8k, 33k, 82k or 200k) • Press Menu/OK [5] to return to Locate mode. B. Power Frequencies Power frequencies refer to 50 or 60 Hz signals, and their harmonics, that can be traced without the use of a transmitter. 60: Best for general locating of passive power. 60H (high harmonic): If the 60 choice appears to be responding slowly, or poorly, then 60H is the second choice for locating passive power. 60L (low harmonic): Third choice for passive power locating. May be used when 60 or 60H is weak or erratic. 60L may be the best choice when locating primary power cables. 120 Hz: Used for locating rectified AC power signals, often found on pipelines using impressed current cathodic protection. • Press Menu/OK [5] (you may have to press Menu/OK [5] twice to toggle to the LOCATE MENU display).
34 78-9000-0192-2 Rev D • Press Cable/Pipe [SK] • Press Freq [SK] • Press Power [SK Toggle] until the desired frequency is displayed: (“Power” 60, 60L, 60H, or 120). • Press Menu/OK [5] to return to Locate mode. All U.S. receivers default to 60 Hz. To set the receiver to detect 50 Hz signals refer to 5.D. Enabling/Disabling Locating Frequencies section of this manual. C. Passive Frequencies The receiver (without a 3M™ Dynatel™ Transmitter 2500 Series) can be used to detect some CATV cables (31.5 kHz). (A horizontal-scan television NTSC must be turned on to generate this frequency.) • Press Menu/OK [5] (you may have to press Menu/OK [5] twice to toggle to the LOCATE MENU display). • Press Cable/Pipe [SK] • Press Freq [SK] • Press Pasv [SK Toggle] • Press Menu/OK [5] to return to Locate mode. D. Auxiliary Frequencies Auxiliary frequencies are signals generated from remote location transmitters, or frequency generating transmitters. • Press Menu/OK [5] (you may have to press Menu/OK [5]twice to toggle to the LOCATE MENU display). • Press Cable/Pipe [SK] • Press Freq [SK] • Press Aux [SK Toggle] until the desired frequency is displayed; (“Aux” 512, 560, 333 Hz, or user defined frequencies). • Press Menu/OK [5] to return to Locate mode.
78-9000-0192-2 Rev D 35 10. Locating in Directional Peak Mode The following are instructions for locating a buried pipe or cable using the direct connect method and the Directional Peak (Dir Pk) mode. Other methods of connection and tracing modes/features are explained in previous sections. Step 1. Insert the ground rod into the ground, perpendicular to the suspected target path. Step 2. Remove the grounding from the near-end of the target cable/pipe. Note: Never connect or disconnect the transmitter when the unit is on. Step 3. Connect the red lead of the transmitter to the shield, neutral, or deenergized target conductor. Step 4. Connect the black lead of the transmitter to the ground rod. Step 5. Perform a battery check by pressing and holding off [T-1]. Step 6. Power on the transmitter by pressing on [T-2] once. This will place the transmitter in Ohm-meter mode. An Indicator Flag will be displayed above the ohm symbol, Ω, in the Digital Display [T-4]. a. A solid tone from the transmitter indicates a complete circuit with a good ground. b. A beeping tone from the transmitter indicates a usable ground. An attempt should be made to improve the ground. c. No tone from the transmitter indicates a poor, or no ground. The transmitter will display ‘OL’ in this instance. An attempt should be made to improve the ground. Verify that the far end is grounded. Step 7. Press on Trace (Frequency on 2550 locator) [T-3] to set the transmitter to Trace mode. Step 8. Select a frequency on the transmitter by pressing Trace (Frequency on 2550 locator) [T-3]. The unit will cycle through the available frequencies (577, 1K, 8K, 33K, 82K, 200K and ALL). Step 9. Press On/Off [1] to power on the receiver. Step 10. Press Menu/OK [5] (you may have to press Menu/OK [5] twice to toggle to the LOCATE MENU display). Step 11. Press Cable/Pipe [SK].
36 78-9000-0192-2 Rev D Step 12. Set the Frequency and Mode of the receiver. a. Press Freq [SK] b. Select the same frequency on the receiver that was selected on the transmitter by pressing Active [SK Toggle]. c. Press Menu/OK [5] to save the setting and return to Locate mode. d. Press Mode [SK Toggle] until Dir Pk is displayed. Step 13. Stand away from the suspected target path and adjust the Gain [3] down until the Bar Graph [11] ([10] for 2550 locator) opens completely. Step 14. Walk in a wide circle with your back toward the transmitter (about 10 to 15 feet (3 to 4.5 m) away). − Watch the receiver display and listen to the signal. Take note of where the receiver detects the strongest Signal Strength [9] ([8] for 2550 locator). − The Bar Graph [11] ([10] for 2550 locator) will close when the unit detects a signal, and the arrows will reverse. − Adjust the Gain [3] down if the Bar Graph [11] ([10] for 2550 locator) closes completely. − The Signal Strength [9] ([8] for 2550 locator) numbers on the display will change with the signal strength (smaller, as you walk away from the target path; larger, as you approach the target path). − Make a complete circle around the transmitter. Return to each point in the circle that the receiver detected. Step 15. Measure the Depth and Current of each point to identify the target path. The depth of the target path should be as expected and the relative current should compare to the relative current that is alternately flashing with the frequency value on the transmitter. Note: The Relative Current reading will decrease steadily as the locator moves away from the transmitter. When using higher frequencies, this decline is more evident. Step 16. Adjust the gain so that the Bar Graph [11] ([10] for 2550 locator) responds to the target path (open when off path, almost completely closed when directly over target). Step 17. Trace the cable/pipe at a slow walk while moving the receiver in a side-to-side motion, keeping the receiver perpendicular to the ground.
78-9000-0192-2 Rev D 37 Step 18. Measure the Depth and Current occasionally to verify target path. Note: In order to measure the Depth and Relative Current accurately, the operator must pinpoint the target pipe or cable, and the receiver handle should be in-line with the target path. − While in Dir Pk mode, find the highest Signal Strength [9] ([8] for 2550 locator). − Lower the tip of the receiver to the ground. Twist the receiver left and right while watching the Signal Strength [9] ([8] for 2550 locator). − When the highest reading is displayed, the handle of the unit is in line with the target pipe or cable. Occasionally a signal will appear on adjacent cables or pipes. Compare t\ he relative and actual current readings (value that is under the relative \ current value and in mA units) over each path to help determine the target path. Current readings will be significantly less on the adjacent cable/pipe compared \ to the target path. Step 19. As tracing proceeds, remember that the most powerful signal is near the transmitter. As the receiver gets farther away from the transmitter the Signal Strength [9] ([8] for 2550 locator) decreases. It may be necessary to readjust the gain as needed to be sure there is adequate signal for the receiver to operate. Press the Gain [3] up or down when the Bar Graph [11] ([10] for 2550 locator) is no longer visible (too little signal) or when the Bar Graph [11] ([10] for 2550 locator) is closed (too much signal). Step 20. Trace the path until you reach a logical termination point (i.e.: terminal, meter, cabinet, etc). 11. Locating Active Duct Probes (Sondes) Step 1. Press On/Off [1] on the receiver. Step 2. Press Menu/OK [5] (you may have to press Menu/OK [5] twice to toggle to the LOCATE MENU display). Step 3. Press Cable/Pipe [SK]. Step 4. Press Mode [SK Toggle] to select Special Peak (Spl Pk). Step 5. Press Freq [SK]. Step 6. Press Active [SK Toggle] to select the 33kHz frequency (for a 33 kHz Sonde or ADP (Active Duct Probe)). Step 7. Press Menu/OK [5].
38 78-9000-0192-2 Rev D Step 8. With the receiver handle perpendicular to the conduit path, locate the ADP position by moving along the path until the strongest Signal Strength [9] ([8] for 2550 locator) is found. Adjust Gain [3] up or down when the Bar Graph [11] ([10] for 2550 locator) remains either fully open or fully closed. Step 9. Refer to the ADP operating instructions for further information. A. Determining Active Duct Probe Depth Step 1. Place the tip of the receiver on the ground directly above the located ADP position. Step 2. Maintain the handle orientation perpendicular to the target path. Step 3. Press DEPTH [4]. Step 4. Press Sonde Depth [SK] to read ADP Depth. − The depth to the ADP is displayed in units as specified in the receiver set up menu. − Five Sonde depth readings can be saved with the time and date measured. Press Mem Select [SK] after the Sonde depth displays. − Press Mem Select [SK] to select a specific memory location (M1–M5) or select Save [SK]. Save [SK] will place each entry in sequential order in memory (M1–M5) until five readings have been stored. The unit will overwrite saved entries in excess of five, beginning with M1. Step 5. Optional step: Press Clear All [SK] to delete all saved depth readings. Step 6. Press Mem Select [SK] to select a specific memory location (M1 - M5) to store the depth readings. Step 7. When the preferred location appears on the display, press Save [SK]. The display and memory location will populate with the current information. Step 8. Each memory location can be reviewed by pressing Mem Select [SK]. Step 9. Press Cable Depth [SK] to switch to the cable depth display, or Menu/OK [5] to return to Locate Mode. Note: During a depth measurement, the display will exhibit ‘- -’ when the received signal is too low, too high or erratic.
78-9000-0192-2 Rev D 39 12. Locating Buried Sheath Faults And Earth Return Faults (3M™ Dynatel™ Models 2573 and 2573-iD only) Note: Remove (disconnect) both the near-end and far-end grounding from the test section. A. Transmitter Setup Note: Do not make any connections while the transmitter is on. Fault loc\ ation will not work unless the near-end and far-end grounds are disconnected. Note: This method only works on direct buried cables/conductors. It will\ not work if the cables/conductors are faulted inside of a conduit, such as a PVC con\ duit. Step 1. Attach the red clip to the earth-faulted conductor of the cable or conductor under test. Step 2. Place the ground rod behind the transmitter and in parallel with the target path. Step 3. Connect the black clip to the ground rod behind the transmitter. Step 4. Press and hold On/Off [1] to perform a battery test. Step 5. Press on [T-2] once to power on the transmitter. This will place the transmitter in Ohm-meter mode. An Indicator Flag will be displayed above the ohm symbol, Ω, in the Digital Display [T-4]. − The resistance of the fault will be displayed in ohms on the transmitter Digital Display [T-4]. − Resistance less than 50KΩ = Significant fault exists − Resistance 50KΩ to 1.0MΩ = High resistance fault exists (May or may not cause problems yet, but gets worse with time.) − Resistance greater than 1.0MΩ = No significant fault exists Note: The fault locating limit is 2.0MΩ, although the transmitter's ohm-meter mode can display a fault resistance up to 10MΩ. Step 6. Press on [T-2] again, to select Fault mode. − An Indicator Flag will be displayed above the fault locating symbol in the Digital Display [T-4]. (as shown in the illustration below) off + - o Trace onon Output Level DynatelTM257 3 Flag Note: Two active line locating frequencies, 577 Hz and 33 kHz, are simultaneously transmitted with the fault finding signal. This allows convenient switching back and forth between fault finding mode and cable/pipe locate mode on the receiver during fault finding without returning to the transmitter. The transmitter display [T-4] will flash between 577 Hz and 33 kHz. It does not display the fault finding frequencies.
40 78-9000-0192-2 Rev D B. Pinpointing the Buried Fault Step 1. Connect the 3M™ Earth Contact Frame to the External Jack [15] ([14] for 2550 locator) of the receiver using the Earth Contact Frame cable (4 ft. (1.2 m) cable). Step 2. Press On/Off [1] to power on the receiver. The receiver display screen will display “Fauly Calibrating” for about 5 seconds. Step 3. Press Menu/OK [5]. Step 4. Press Fault [SK] to select Fault mode. Step 5. Hold the receiver in one hand and the Earth Contact Frame in the other with the solid green-banded leg of the frame toward the test section. Near the location of the ground rod (about one Earth Contact Frame width away from the ground rod, slightly to the side and towards the faulted cable.), insert the Earth Contact Frame probes fully into the ground in line with the target path. Step 6. Press REF [SK] to record the fault Signal Strength [9] ([8] for 2550 locator) level reference. The signal level will be recorded in the box above REF [SK] on the display. This reference indicates the Signal Strength [9] ([8]) for 2550 locator) level at the ground rod. When the operator reaches the major fault location, the Signal Strength [9] ([8]) for 2550 locator) indicated on the receiver will be very close (within 12dB) to this reference signal strength level. Step 7. Continue along the cable path, re-inserting the Earth Contact Frame probes every few steps while watching the receiver bar graph. The bar graph on the receiver will fill toward the right side of the display (green Fault Finding Direction Indicator [8] (See illustration below)), indicating that the fault is ahead of the operator (in the direction of the green-banded leg of the Earth Contact Frame). Move in direction of red-banded leg of 3M™ Earth Contact Frame. Move in direction of green-banded leg of 3M™ Earth Contact Frame. Transmitter Signal Green Band Red-and-White-Striped Band Ground Fault location