Home > 3M > Locator > 3M CablePipeFault Locator 25502573 Series User Manual

3M CablePipeFault Locator 25502573 Series User Manual

    Download as PDF Print this page Share this page

    Have a look at the manual 3M CablePipeFault Locator 25502573 Series User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 307 3M manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							78-9000-0192-2 Rev D 31
    E. Directional Null (DirNull)
    TRACE [2] + Mode [SK] + DirNull [SK Toggle]  
    or Menu/OK [5] + Cable/Pipe [SK] + Mode [SK] + DirNull [SK Toggle]
    In Directional Null mode, as the operator approaches the cable or pipe, the numerical 
    Signal Strength [9] ([8] for 2550 locator) will increase then fall sharply as the receiver 
    crosses the target cable or pipe. The Bar Graph [11] ([10] for 2550 locator) fills from 
    both sides toward the middle and the receiver speaker volume decreases. As the antenna 
    moves off the target path, the Bar Graph [11] ([10] for 2550 locator) opens, the Signal 
    Strength [9] ([8] for 2550 locator) increases, and the speaker volume increases. Gain 
    adjust is automatic in Directional Null mode.
    The center of the DirNull display provides a ‘compass view’ of the target path. An arrow 
    will point toward the location of the cable/pipe in 45-degree steps. A solid line will 
    appear over the cable/pipe, indicating its orientation to the receiver  handle.
    For example:
    Figure 1: If the target path is to the right of the 
    receiver, and running parallel to the operator, the 
    right arrow will display.
    Figure 2: If the target path is to the left and in front 
    of the operator (not running parallel to the receiver) 
    the arrow will point toward the top left side of 
    the display.
    Figure 3: When the receiver crosses the target path, 
    a solid line will appear, instead of arrows, indicating 
    the target path and its orientation to the receiver 
    handle.
    Before marking target path, always use Directional Peak (Dir Pk), Special Peak (Spl Pk) 
    or Trace View (T-View) mode to verify location. 
    						
    							 32 78-9000-0192-2 Rev D
    F.  Expanded Mode
    VOL [13] (press up arrow)
    When the third ring of the speaker icon is dotted 
    or broken and ‘xpnd’ appears below the speaker 
    icon as shown, the receiver is in “Expanded” 
    mode. This mode is used for pinpointing a 
    target cable or pipe. The area of response of the 
    receiver narrows, allowing the locator to detect 
    very small signal changes. Press the Speaker 
    Volume Control [13] ([12] for 2550 locator) key 
    to activate the “Expanded” mode.
    8.  Depth and Current Estimate
    Verifying the target path, depth and current can be helpful tools.
    •  Pinpoint the cable or pipe being located.
    •  Lower the tip of the receiver to the ground and 
    press  DEPTH [4].
     
    − The depth to the target cable or pipe is displayed 
    in the units specified in the receiver set up 
    menu. The bold current reading is the relative 
    current measurement. This reading can be 
    compared to the relative current reading that 
    alternately flashes with the frequency on the 
    transmitter to help identify the target cable/pipe. 
    The highest relative current value will be the 
    target cable/pipe and will compare closest to the 
    transmitter relative current value. The milliamp 
    reading is an actual current measurement.
     
    − There are two options for measuring depth. 
    Live depth is a continuous measurement. 1-Shot 
    Depth is an averaging of the depth reading over 
    a short period of time. When in 1-Shot mode, 
    the unit will average the depth reading for three 
    seconds, and then display the result on the 
    display. Press DEPTH [4] to alternate between 
    the two modes.
     
    − Five depth readings can be saved with the time, 
    date and relative current measurements.
    • Pressing  Mem Select  [SK] and then Save  [SK] will place each entry in sequential 
    order in memory (M1–M5) until five readings have been stored. The unit will 
    overwrite saved entries in excess of five, beginning with M1. 
    						
    							78-9000-0192-2 Rev D 33
    • Optional step: Press Clear All  [SK] to delete all 
    stored depth information.
     
    − The operator may select a specific memory 
    location to store the depth readings by pressing 
    Mem Select [SK]. When the preferred location 
    appears on the display, press Save [SK]. The 
    display and memory location will populate with 
    the current information.
     
    − Each memory location can be reviewed by 
    pressing Mem Select [SK].
    •  Press  TRACE [2] or DEPTH [4] to return to Locate 
    mode, or wait until receiver automatically returns 
    to the Locate  mode.
    9. Locating Frequencies
    A.  Active Frequencies
    Active frequencies are trace signals supplied by a 3M™ Dynatel™ Transmitter 2500 Series 
    (577 Hz, 1 kHz, 8 kHz, 33 kHz, 82 kHz or 200 kHz).
    •  Select the same frequency that the transmitter is generating.
    •  Press Menu/OK [5] (you may have to press Menu/OK  [5] twice to toggle to the 
    LOCATE MENU display).
    •  Press  Cable/Pipe [SK]
    •  Press  Freq [SK]
    •  Press  Active  [SK Toggle] until the desired frequency is displayed: (“Active” 577, 1k, 
    8k, 33k, 82k or 200k)
    •  Press Menu/OK [5] to return to Locate mode.
    B.  Power Frequencies
    Power frequencies refer to 50 or 60 Hz signals, and their harmonics, that can be traced 
    without the use of a transmitter.
    60:  Best for general locating of passive power. 
    60H (high harmonic): If the 60 choice appears to be responding slowly, or poorly, then 
    60H is the second choice for locating passive power.
    60L (low harmonic): Third choice for passive power locating. May be used when 60 or 
    60H is weak or erratic. 60L may be the best choice when locating primary power cables.
    120 Hz: Used for locating rectified AC power signals, often found on pipelines using 
    impressed current cathodic protection.
    •  Press Menu/OK [5] (you may have to press Menu/OK  [5] twice to toggle to the 
    LOCATE MENU display). 
    						
    							 34 78-9000-0192-2 Rev D
    • Press  Cable/Pipe [SK]
    •  Press  Freq [SK]
    •  Press  Power  [SK Toggle] until the desired frequency is displayed: (“Power” 60, 60L, 
    60H, or 120).
    •  Press Menu/OK [5] to return to Locate mode.
    All U.S. receivers default to 60 Hz.
    To set the receiver to detect 50 Hz signals refer to 5.D.  Enabling/Disabling Locating 
    Frequencies  section of this manual.
    C. Passive Frequencies
    The receiver (without a 3M™ Dynatel™ Transmitter 2500 Series) can be used to detect 
    some CATV cables (31.5 kHz). (A horizontal-scan television NTSC must be turned on to 
    generate this frequency.)
    •  Press Menu/OK [5] (you may have to press Menu/OK  [5] twice to toggle to the 
    LOCATE MENU display).
    •  Press  Cable/Pipe  [SK]
    •  Press  Freq [SK]
    •  Press  Pasv [SK Toggle]
    •  Press Menu/OK [5] to return to Locate mode.
    D.  Auxiliary Frequencies
    Auxiliary frequencies are signals generated from remote location transmitters, or 
    frequency generating transmitters.
    • Press Menu/OK [5] (you may have to press Menu/OK  [5]twice to toggle to the 
    LOCATE MENU display).
    •  Press  Cable/Pipe  [SK]
    •  Press  Freq [SK]
    •  Press  Aux [SK Toggle] until the desired frequency is displayed; (“Aux” 512, 560, 
    333 Hz, or user defined frequencies).
    •  Press Menu/OK [5] to return to Locate mode. 
    						
    							78-9000-0192-2 Rev D 35
    10. Locating in Directional Peak Mode
    The following are instructions for locating a buried pipe or cable using the direct connect 
    method and the Directional Peak (Dir Pk) mode. Other methods of connection and 
    tracing modes/features are explained in previous sections.
    Step 1.  Insert the ground rod into the ground, perpendicular to the suspected  
    target path.
    Step 2. Remove the grounding from the near-end of the target cable/pipe.
    Note: Never connect or disconnect the transmitter when the unit is on.
    Step 3.  Connect the red lead of the transmitter to the shield, neutral, or deenergized 
    target conductor.
    Step 4.  Connect the black lead of the transmitter to the ground rod.
    Step 5.  Perform a battery check by pressing and holding off  [T-1].
    Step 6.  Power on the transmitter by pressing on [T-2] once. This will place the 
    transmitter in Ohm-meter mode. An Indicator Flag will be displayed above the 
    ohm symbol, Ω, in the Digital Display [T-4]. 
    a. A solid tone from the transmitter indicates a complete circuit with a good 
    ground.
    b.  A beeping tone from the transmitter indicates a usable ground. An attempt 
    should be made to improve the ground.
    c.  No tone from the transmitter indicates a poor, or no ground. The transmitter 
    will display ‘OL’ in this instance. An attempt should be made to improve the 
    ground. Verify that the far end is grounded.
    Step 7.  Press on Trace  (Frequency on 2550 locator) [T-3] to set the transmitter to  
    Trace mode.
    Step 8.  Select a frequency on the transmitter by pressing Trace  (Frequency on 2550 
    locator) [T-3]. The unit will cycle through the available frequencies (577, 1K, 
    8K, 33K, 82K, 200K and ALL).
    Step 9.  Press On/Off [1] to power on the receiver.
    Step 10.  Press Menu/OK [5] (you may have to press Menu/OK  [5] twice to toggle to the 
    LOCATE MENU display).
    Step 11.  Press Cable/Pipe  [SK]. 
    						
    							 36 78-9000-0192-2 Rev D
    Step 12. Set the Frequency and Mode of the receiver.
    a. Press  Freq [SK]
    b.  Select the same frequency on the 
    receiver that was selected on the 
    transmitter by pressing Active  [SK 
    Toggle].
    c.  Press Menu/OK [5] to save the setting 
    and return to Locate mode.
    d.  Press  Mode  [SK Toggle] until Dir Pk  
    is displayed.
    Step 13.  Stand away from the suspected target path and adjust the Gain [3] down until 
    the Bar Graph [11] ([10] for 2550 locator) opens completely.
    Step 14.  Walk in a wide circle with your back toward the transmitter (about 10 to 15 feet 
    (3 to 4.5 m) away).
     
    − Watch the receiver display and listen to the signal. Take note of where the 
    receiver detects the strongest Signal Strength [9] ([8] for 2550 locator).
     
    − The Bar Graph [11] ([10] for 2550 locator) will close when the unit detects 
    a signal, and the arrows will reverse.
     
    − Adjust the Gain [3] down if the Bar Graph [11] ([10] for 2550 locator) 
    closes completely.
     
    − The Signal Strength [9] ([8] for 2550 locator) numbers on the display will 
    change with the signal strength (smaller, as you walk away from the target 
    path; larger, as you approach the target path).
     
    − Make a complete circle around the transmitter. Return to each point in the 
    circle that the receiver detected.
    Step 15.  Measure the Depth and Current of each point to identify the target path. The 
    depth of the target path should be as expected and the relative current should 
    compare to the relative current that is alternately flashing with the frequency 
    value on the transmitter.
    Note: The Relative Current reading will decrease steadily as the locator moves away 
    from the transmitter. When using higher frequencies, this decline is more evident.
    Step 16.  Adjust the gain so that the Bar Graph [11] ([10] for 2550 locator) responds to 
    the target path (open when off path, almost completely closed when directly 
    over target).
    Step 17.  Trace the cable/pipe at a slow walk while moving the receiver in a side-to-side 
    motion, keeping the receiver perpendicular to the ground. 
    						
    							78-9000-0192-2 Rev D 37
    Step 18. Measure the Depth and Current occasionally to verify target path.
    Note: In order to measure the Depth and Relative Current accurately, the operator 
    must pinpoint the target pipe or cable, and the receiver handle should be in-line with 
    the target path.  
    − While in Dir Pk mode, find the highest Signal Strength [9] ([8] for 2550 
    locator).
     
    − Lower the tip of the receiver to the ground. Twist the receiver left and right 
    while watching the Signal Strength [9] ([8] for 2550 locator).
     
    − When the highest reading is displayed, the handle of the unit is in line with 
    the target pipe or cable.
    Occasionally a signal will appear on adjacent cables or pipes. Compare t\
    he 
    relative and actual current readings (value that is under the relative \
    current 
    value and in mA units) over each path to help determine the target path. Current 
    readings will be significantly less on the adjacent cable/pipe compared \
    to the 
    target path.
    Step 19.  As tracing proceeds, remember that the most powerful signal is near the 
    transmitter. As the receiver gets farther away from the transmitter the Signal 
    Strength [9] ([8] for 2550 locator) decreases. It may be necessary to readjust the 
    gain as needed to be sure there is adequate signal for the receiver to operate. 
    Press the Gain [3] up or down when the Bar Graph [11] ([10] for 2550 locator) 
    is no longer visible (too little signal) or when the Bar Graph [11] ([10] for 2550 
    locator) is closed (too much signal).
    Step 20.  Trace the path until you reach a logical termination point (i.e.: terminal, meter, 
    cabinet, etc).
    11. Locating Active Duct Probes (Sondes)
    Step 1.  Press On/Off [1] on the receiver.
    Step 2.  Press Menu/OK [5] (you may have to press Menu/OK  [5] twice to toggle to the 
    LOCATE MENU display).
    Step 3.  Press Cable/Pipe  [SK].
    Step 4.  Press Mode  [SK Toggle] to select Special Peak (Spl Pk).
    Step 5.  Press Freq [SK].
    Step 6.  Press Active  [SK Toggle] to select the 33kHz frequency (for a 33 kHz Sonde 
    or  ADP (Active Duct Probe)).
    Step 7.  Press Menu/OK [5]. 
    						
    							 38 78-9000-0192-2 Rev D
    Step 8. With the receiver handle perpendicular to the conduit path, locate the ADP 
    position by moving along the path until the strongest Signal Strength [9] ([8] 
    for 2550 locator) is found. Adjust Gain [3] up or down when the Bar Graph 
    [11] ([10] for 2550 locator) remains either fully open or fully closed.
    Step 9.  Refer to the ADP operating instructions for further information.
    A. Determining Active Duct Probe Depth
    Step 1.  Place the tip of the receiver on the ground directly above the located ADP 
    position.
    Step 2.  Maintain the handle orientation perpendicular to the target path.
    Step 3.  Press DEPTH [4].
    Step 4.  Press Sonde Depth [SK] to read ADP Depth.
     
    − The depth to the ADP is displayed in 
    units as specified in the receiver set up 
    menu.
     
    − Five Sonde depth readings can be saved 
    with the time and date measured. Press 
    Mem Select [SK] after the Sonde depth 
    displays.
     
    − Press Mem Select [SK] to select a 
    specific memory location (M1–M5) or 
    select Save [SK]. Save [SK] will place 
    each entry in sequential order in memory 
    (M1–M5) until five readings have been 
    stored. The unit will overwrite saved 
    entries in excess of five, beginning 
    with M1.
    Step 5.  Optional step: Press Clear All  [SK] to delete all saved depth readings.
    Step 6.  Press Mem Select  [SK] to select a specific memory location (M1 - M5) to store 
    the depth readings.
    Step 7.  When the preferred location appears on the display, press Save  [SK]. The 
    display and memory location will populate with the current information.
    Step 8.  Each memory location can be reviewed by pressing Mem Select  [SK].
    Step 9.  Press Cable Depth [SK] to switch to the cable depth display, or Menu/OK [5] to 
    return to Locate Mode.
    Note: During a depth measurement, the display will exhibit ‘- -’ when the received 
    signal is too low, too high or erratic. 
    						
    							78-9000-0192-2 Rev D 39
    12. Locating Buried Sheath Faults And Earth 
    Return Faults  
    (3M™ Dynatel™ Models 2573 and 2573-iD only)
    Note: Remove (disconnect) both the near-end and far-end grounding from the  
    test section.
    A.  Transmitter Setup
    Note: Do not make any connections while the transmitter is on. Fault loc\
    ation will not 
    work unless the near-end and far-end grounds are disconnected.
    Note: This method only works on direct buried cables/conductors. It will\
     not work if 
    the cables/conductors are faulted inside of a conduit, such as a PVC con\
    duit.
    Step 1.  Attach the red clip to the earth-faulted conductor of the cable or conductor 
    under test.
    Step 2.  Place the ground rod behind the transmitter and in parallel with the target path.
    Step 3.  Connect the black clip to the ground rod behind the transmitter.
    Step 4.  Press and hold On/Off [1] to perform a battery test.
    Step 5.  Press on [T-2] once to power on the transmitter. This will place the transmitter 
    in Ohm-meter mode. An Indicator Flag will be displayed above the ohm 
    symbol,  Ω, in the Digital Display [T-4].
     
    − The resistance of the fault will be displayed in ohms on the transmitter 
    Digital Display [T-4].
     
    − Resistance less than 50KΩ = Significant fault exists
     
    − Resistance 50KΩ to 1.0MΩ = High resistance fault exists (May or may not 
    cause problems yet, but gets worse with time.)
     
    − Resistance greater than 1.0MΩ = No significant fault exists
    Note: The fault locating limit is 2.0MΩ, although the transmitter's ohm-meter mode 
    can display a fault resistance up to 10MΩ.
    Step 6.  Press on [T-2] again, to select Fault mode.
     
    − An Indicator Flag will be displayed above the fault locating symbol in the 
    Digital Display [T-4]. (as shown in the illustration below)
      off
    +
    -
    o
     
    Trace
    onon Output Level
    DynatelTM257 3
    Flag
    Note: Two active line locating frequencies, 577 Hz and 33 kHz, are simultaneously 
    transmitted with the fault finding signal. This allows convenient switching back and 
    forth between fault finding mode and cable/pipe locate mode on the receiver during 
    fault finding without returning to the transmitter. The transmitter display [T-4] will 
    flash between 577 Hz and 33 kHz. It does not display the fault finding frequencies. 
    						
    							 40 78-9000-0192-2 Rev D
    B. Pinpointing the Buried Fault
    Step 1.  Connect the 3M™ Earth Contact Frame to the External Jack [15] ([14] for 2550 
    locator) of the receiver using the Earth Contact Frame cable (4 ft. (1.2 m) cable).
    Step 2.  Press On/Off [1] to power on the receiver. The receiver display screen will 
    display  “Fauly Calibrating” for about 5 seconds.
    Step 3.  Press Menu/OK  [5].
    Step 4.  Press Fault  [SK] to select Fault mode.
    Step 5.  Hold the receiver in one hand and the Earth Contact Frame in the other with the 
    solid green-banded leg of the frame toward the test section. Near the location 
    of the ground rod (about one Earth Contact Frame width away from the ground 
    rod, slightly to the side and towards the faulted cable.), insert the Earth Contact 
    Frame probes fully into the ground in line with the target path.
    Step 6.  Press REF [SK] to record the fault  Signal Strength [9] ([8] for 2550 locator) 
    level reference. The signal level will be recorded in the box above REF  [SK] 
    on the display. This reference indicates the Signal Strength [9] ([8]) for 2550 
    locator) level at the ground rod. When the operator reaches the major fault 
    location, the Signal Strength [9] ([8]) for 2550 locator) indicated on the receiver 
    will be very close (within 12dB) to this reference signal strength level.
    Step 7.  Continue along the cable path, re-inserting the Earth Contact Frame probes 
    every few steps while watching the receiver bar graph. The bar graph on the 
    receiver will fill toward the right side of the display (green Fault Finding 
    Direction Indicator [8] (See illustration below)), indicating that the fault is 
    ahead of the operator (in the direction of the green-banded leg of the Earth 
    Contact Frame).
     
    Move in direction of red-banded leg  
    of 3M™ Earth Contact Frame.
    Move in direction of green-banded leg   of 3M™ Earth Contact Frame.
    Transmitter Signal Green Band
    Red-and-White-Striped Band
    Ground Fault location 
    						
    All 3M manuals Comments (0)

    Related Manuals for 3M CablePipeFault Locator 25502573 Series User Manual